Skip to content

SJM™ Masters Series 23MECJ-502Recalls & reported problems

St. Jude Medical, Inc. · Mitral Expanded Rotatable Mechanical Heart Valve · Class 3 · Implant

All models of SJM Masters Series (27)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 05414734006170), received between Aug 31, 2020 and Jul 19, 2024 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Biocompatibility” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Biocompatibility1
  • Fracture1
  • Gradient Increase1
  • Incomplete Coaptation1
  • Obstruction of Flow1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Dyspnea1
  • Thrombus1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 19, 2024Incomplete Coaptation; Incomplete Coaptation

    IT WAS REPORTED THAT ON (B)(6) 2024, A 23MM MITRAL MASTERS VALVE WAS SELECTED FOR IMPLANT. THE VALVE WAS SIZED WITH A 905 ABBOTT SIZER SET. DURING DEVICE PREPARATION, THE LEAFLETS WERE TESTED AND MOVING NORMALLY. AFTER IMPLANT, THE VALVE LEAFLETS WERE TESTED WITH A TESTER AND THE VALVE DID NOT CLOSE COMPLETELY; THE LEAFLETS DID NOT HAVE DIFFICULTY OPENING COMPLETELY. THEREFORE, THE VALVE WAS EXPLANTED; THE PATIENT HAD NOT BEEN CLOSED PRIOR TO EXPLANTING THE VALVE. UPON EXPLANT, THE LEAFLET WAS NOT TESTED AGAIN; A NEW 25MM MITRAL MASTERS VALVE WAS SUCCESSFULLY IMPLANTED. THE PATIENT'S ANTITHROMBOTIC REGIMEN CONSISTED OF HEPARIN; THE PATIENT'S LATEST INTERNATIONAL NORMALIZED RATIO (INR) AND INR HISTORY ARE UNKNOWN. THE PATIENT WAS REPORTED TO BE IN STABLE CONDITION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19784680 ↗
  • InjuryOct 21, 2021Fracture; Fracture

    IT WAS REPORTED THAT ON (B)(6) 2021, A 23 MM SJM MASTERS SERIES VALVE EXPANDED CUFF WAS SELECTED FOR IMPLANT. THE VALVE WAS IMPLANTED, HOWEVER, AS THE PHYSICIAN WENT TO ROTATE THE VALVE, A LEAFLET BROKE INTO THREE PIECES. ALL PIECES OF THE VALVE WERE CONFIRMED TO HAVE BEEN REMOVED. A NEW 21 MM EPIC AORTIC VALVE WAS SUCCESSFULLY IMPLANTED. THE USER ALLEGE A CLINICALLY SIGNIFICANT PROLONGED PROCEDURE TIME. THE PATIENT WAS REPORTED TO BE IN STABLE CONDITION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12674432 ↗
  • InjuryAug 31, 2020Gradient Increase; Obstruction of Flow; Biocompatibility; Gradient Increase; Obstruction of Flow; Biocompatibility

    ON AN UNKNOWN DATE, A (B)(6) YEAR OLD FEMALE PRESENTED TO THE HOSPITAL WITH A DIAGNOSIS OF RHEUMATIC HEART DISEASE AND SEVERE MITRAL STENOSIS. THE PATIENT UNDERWENT BALLOON MITRAL VALVOTOMY(BMV), BUT IT WAS UNSUCCESSFUL. ON (B)(6) 2019, THE PATIENT UNDERWENT EMERGENCY MITRAL VALVE REPLACEMENT AND THE MITRAL VALVE WAS REPLACED WITH A 23MM SJM MASTERS SERIES VALVE EXPANDED CUFF (SN: (B)(4)). ON (B)(6) 2020, THE PATIENT PRESENTED TO THE HOSPITAL WITH BREATHLESSNESS. HER PROTHROMBIN TIME AND INR WERE IN THERAPEUTIC RANGE(INR WAS 2.8). HOWEVER, HER ECHOCARDIOGRAPHY SHOWED A STUCK VALVE DUE TO ACUTE THROMBOSIS AND THE PATIENT UNDERWENT SUCCESSFUL THROMBOLYSIS AND THE PATIENTS ANTICOAGULATION DOSE WAS INCREASED. ON (B)(6) 2020, THE PATIENT PRESENTED WITH BREATHLESSNESS AGAIN. THE PATIENT'S PROTHROMBIN TIME WAS PROLONGED WITH PROTHROMBIN TIME TEST MORE THAN 100 SECONDS. THE ECHOCARDIOGRAPHY SHOWED AN ELEVATED MITRAL VALVE GRADIENT. FLUOROSCOPY WAS DONE AND IT SHOWED IMMOBILITY OF THE PROSTHETIC MITRAL VALVE LEAFLET. THE PATIENT UNDERWENT THROMBOLYSIS, BUT THE THROMBOLYSIS FAILED TO RELEASE THE STUCK VALVE. THE PHYSICIAN DECIDED TO PERFORM A RE-DO MITRAL VALVE REPLACEMENT. ON (B)(6) 2020, T

    Patient problems: Dyspnea; Thrombus

    MAUDE report 10473149 ↗

Other HEART-VALVE, MECHANICAL devices

All HEART-VALVE, MECHANICAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page