SJM™ Masters SeriesHemashield™ 27CAVGJ-514 00Recalls & reported problems
St. Jude Medical, Inc. · St. Jude Medical Valved Graft Rotatable Mechanical Heart Valve · Class 3 · Implant
All models of SJM Masters SeriesHemashield (5)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 1
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 05414734006422), received between Nov 18, 2020 and Feb 13, 2026 — classified as 1 death, 2 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem3
- Detachment of Device or Device Component2
- Fracture2
- Material Separation1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Abdominal Pain1
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Atrial Fibrillation1
- Brain Injury1
- Cardiac Tamponade1
- Dizziness1
- Dyspnea1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 13, 2026Fracture
IT WAS REPORTED THAT ON (B)(6) 2026, A 27MM SJM MASTERS SERIES COATED AORTIC VALVED GRAFT WAS CHOSEN FOR A PROCEDURE. DURING THE IMPLANTATION, IT WAS FOUND THAT THE VALVE RING WAS DAMAGED, SO IT WAS REPLACED WITH A NON-ABBOTT VALVE FOR THE PROCEDURE WHILE PATIENT ON BYPASS. THERE WAS NO DELAY IN THE PROCEDURE. THE PATIENT WAS REPORTED STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24332305 ↗MalfunctionMay 10, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT ON (B)(6) 2023, A 27MM SJM MASTERS SERIES COATED AORTIC VALVED GRAFT WAS CHOSEN FOR IMPLANT. AFTER THE DEVICE WAS PLACED AND SECURED, IT WAS NOTICED THAT OF THE LEAFLETS WAS NOT IN ITS HOUSING. NO INSTRUMENTS WERE IN CONTACT WITH THE VALVE AT THE TIME OF DISLODGEMENT. THE LEAFLET WAS RETRIEVED FROM THE LEFT VENTRICLE AND THE DEVICE WAS EXPLANTED. AN UNKNOWN SIZE NON-ABBOTT VALVE WAS SUCCESSFULLY IMPLANTED AS A REPLACEMENT DEVICE. THE PATIENT REMAINED HEMODYNAMICALLY STABLE THROUGHOUT THE PROCEDURE AND NO CLINICALLY SIGNIFICANT DELAY WAS REPORTED. THE PATIENT WAS REPORTED AS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16900343 ↗MalfunctionMay 3, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT ON (B)(6) 2023, A 27MM SJM MASTERS SERIES COATED AORTIC VALVED GRAFT WAS CHOSEN FOR IMPLANT. DURING PROCEDURE, IT WAS NOTICED ONE OF THE LEAFLETS HAD BECOME COMPLETELY DETACHED IN ONE PIECE. IT WAS REPORTED THE HOLDER HANDLE WAS FULLY INSERTED INTO THE ORIFICE AT A 90 DEGREE ANGLE. THE DETACHED LEAFLET WAS REMOVED FROM THE PATIENT'S BODY. A DECISION WAS MADE TO REPLACE THE DEVICE. THERE WAS NO ATTEMPT MADE TO REINSERT THE DETACHED LEAFLET. THE PATIENT STATUS WAS REPORTED AS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16865527 ↗DeathApr 27, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON (B)(6) 2022, A 27MM MASTERS AORTIC VALVED GRAFT WAS SUCCESSFULLY IMPLANTED. THERE WERE NO INTRAPROCEDURAL ADVERSE EVENTS. ON (B)(6) 2022, PATIENT'S C-REACTIVE PROTEIN WAS 237.3 MG/L. ON (B)(6) 2022, THE PATIENT HAD A SEVERE ISCHEMIC STROKE WITH MRS SCORE OF 5. THE PATIENT WAS UNCONSCIOUS WITH NO NOTICEABLE REACTION TO STIMULI AND NO REFLEXES. THE PATIENT HAD LEFT-SIDED PARESIS. A COMPUTED TOMOGRAPHY (CT) SCAN REVEALED ACUTE ISCHEMIC CHANGES AND A SWELLING OF ALMOST THE ENTIRE RIGHT HEMISPHERE OF THE BRAIN. THE PATIENT WAS INTUBATED AND PLACED ON MECHANICAL VENTILATION. MEDICATIONS INCLUDING ANTIBIOTICS, ANTIFUNGAL, NARCOTICS, DIURETICS, SEDATIVES, BETA BLOCKERS, AND ANTICOAGULANTS WERE ADMINISTERED. ON (B)(6) 2022, THE PATIENT WAS DISCHARGED FROM THE IMPLANT SITE HOSPITAL AND TRANSFERRED TO ANOTHER HOSPITAL. THE PATIENT'S SYMPTOMS WERE ONGOING AT TIME OF TRANSFER. THERE WERE NO DEVICE DEFICIENCIES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Brain Injury; Respiratory Insufficiency; Stroke/CVA
MAUDE report 14217124 ↗InjuryDec 9, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED ON (B)(6) 2021, A 27MM SJM MASTERS SERIES COATED AORTIC VALVED GRAFT WAS IMPLANTED. ON (B)(6) 2021, VIA ECHOCARDIOGRAM A PERICARDIAL EFFUSION SPANNING THE ENTIRE HEART WAS NOTED. A PERICARDIOCENTESIS WAS PERFORMED AND 480ML OF BLOODY SECRETION WAS REMOVED WITHOUT ANY COMPLICATIONS. THE PATIENT WAS TRANSFERRED BACK TO THE NORMAL WARD AFTER SPENDING SOME TIME IN THE INTENSIVE CARE UNIT (ICU). IT IS THOUGH THAT THE PERICARDIAL EFFUSION WAS DUE TO A INFLAMMATORY REACTION TO THE SURGERY. THE PATIENT WAS DISCHARGED ON (B)(6) 2021. (CRD_992 - VALVED GRAFTS PAS; EU0123 -303; R643214001).
Patient problems: Pericardial Effusion
MAUDE report 12967111 ↗InjuryOct 13, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED ON (B)(6) 2021, A 27MM SJM MASTERS SERIES COATED AORTIC VALVED GRAFT WAS IMPLANTED. ON (B)(6) 2021, THE PATIENT PRESENTED TO THE CLINICAL EXPERIENCING DYSPNEA, WEAKNESS, ABDOMINAL DISCOMFORT, ATRIAL FIBRILLATION (AF), DIZZINESS AND ANXIETY. MEDICATION WAS GIVEN TO TREAT THE AF. AN ECHOCARDIOGRAM WAS PERFORMED A PERICARDIAL TAMPONADE WAS NOTED. A PROCEDURE WAS PERFORMED AND LIQUID WAS REMOVED. IT IS THOUGHT THAT POSTOPERATIVE BLEEDING OR POSTPERICARDIOTOMY SYNDROME CAUSED THE TAMPONADE. THE PATIENT WAS REPORTED TO BE IN STABLE CONDITION AND WAS DISCHARGED ON (B)(6) 2021. (CRD_992 - VALVED GRAFTS PAS; PL4195-251; R628368101).
Patient problems: Abdominal Pain; Atrial Fibrillation; Cardiac Tamponade; Dizziness; Dyspnea; Muscle Weakness/Atrophy
MAUDE report 12625008 ↗MalfunctionNov 18, 2020Fracture; Material Separation; Fracture; Material Separation
ON (B)(6) 2020, A 27MM MASTERS SERIES CAVG WAS SELECTED FOR IMPLANT IN THE PATIENT. DURING DEVICE IMPLANTATION, THE VALVE LEAFLETS SNAPPED. THE PHYSICIAN USED THE VALVE ROTATOR RECOMMENDED IN THE IFU, BUT RESISTANCE WAS FELT AND ONE LEAFLET WAS INTACT AND THE OTHER SHATTERED AS IT FELL INTO THE LEFT VENTRICLE. ALL PIECES OF THE DISLODGED LEAFLET AND FRACTURED PIECES WERE REMOVED FROM THE PATIENT AND THE PATIENT DID NOT EXPERIENCE ANY SERIOUS INJURIES. THE VALVE WAS REPLACED WITH A 25MM MASTERS SERIES CAVG AND THE PATIENT REMAINED HEMODYNAMICALLY STABLE THROUGHOUT THE PROCEDURE. AN INCREASED BYPASS TIME WAS REPORTED. THE PATIENT IS CURRENTLY STABLE AND DISCHARGED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10864525 ↗
Other HEART-VALVE, MECHANICAL devices
- On-X Mitral Heart Valve with Conform-X Sewing Ring ONXMC-25/33On-X Life Technologies, Inc. · Heart-Valve, Mechanical1 recall1 death34 reports
- SJM Regent™ AGFN-756St. Jude Medical, Inc. · HEART-VALVE, MECHANICAL3 deaths70 reports
- SJM Regent™ AGFN-756St. Jude Medical, Inc. · HEART-VALVE, MECHANICAL5 deaths66 reports
- SJM Regent™ 23AGFN-756St. Jude Medical, Inc. · HEART-VALVE, MECHANICAL3 deaths44 reports
- CARBOMEDICS SUPRA-ANNULAR (TOP HAT) S5-021Corcym Srl · HEART-VALVE, MECHANICAL7 deaths37 reports
- Regent™ 21AGFN-756St. Jude Medical, Inc. · HEART-VALVE, MECHANICAL3 deaths35 reports
- Regent™ 23AGFN-756St. Jude Medical, Inc. · HEART-VALVE, MECHANICAL2 deaths33 reports
- Regent™ 19AGFN-756St. Jude Medical, Inc. · HEART-VALVE, MECHANICAL31 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page