SJM™ Masters Series 25MTJ-503Recalls & reported problems
St. Jude Medical, Inc. · Mitral PTFE Rotatable Mechanical Heart Valve · Class 3 · Implant
All models of SJM Masters Series (27)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 05414734053617), received between Sep 11, 2018 and Apr 7, 2023 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Backflow” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Backflow1
- Biocompatibility1
- Obstruction of Flow1
- Perivalvular Leak1
Patient problems
- Calcium Deposits/Calcification1
- Endocarditis1
- Fever1
- Mitral Regurgitation1
- Thrombus1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryApr 7, 2023Perivalvular Leak; Perivalvular Leak
IT WAS REPORTED THAT ON (B)(6) 2018, A 25MM SJM MASTERS SERIES HEART VALVE WITH TEFLON CUFF WAS SUCCESSFUL IMPLANTED. ON (B)(6) 2023, THE PATIENT PRESENTED WITH INFECTIVE ENDOCARDITIS AND A RE-DO PROCEDURE WAS PERFORMED. THERE WAS NO PARTICULAR PROBLEM DUE TO IT BEING A MECHANICAL VALVE, BUT REOPERATION WAS PERFORMED BECAUSE OF PARA-VALVULAR LEAK. THE DEVICE WAS EXPLANTED AND REPLACED WITH ANOTHER 25MM SJM MASTERS SERIES HEART VALVE WITH TEFLON CUFF. SUBSEQUENT TO THE PREVIOUSLY FILED REPORT, ADDITIONAL INFORMATION WAS RECEIVED THAT THE PATIENT HAD A FEVER DUE TO THE INFECTIVE ENDOCARDITIS. THE PATIENT DOES NOT HAVE ANY HISTORY OF ENDOCARDITIS. THE PATIENT DOES NOT HAVE A HISTORY AN INDWELLING VASCULAR CATHETER, INTRAVENOUS DRUG USE, RECENT DENTAL WORK, OR INJURY FROM A NON-STERILE OBJECT. THE PARAVALVULAR LEAK IS BELIEVED TO HAVE BEEN A RESULT OF THE INFECTIVE ENDOCARDITIS. THE IMPLANTING PHYSICIAN BELIEVES THAT THE CAUSE OF THE ENDOCARDITIS IS DUE TO A PATIENT CONDITION. THE PATIENT WAS STABLE AT THE TIME OF REPORT.
Patient problems: Endocarditis; Fever
MAUDE report 16704681 ↗MalfunctionJun 26, 2020Backflow; Biocompatibility; Backflow; Biocompatibility
ON (B)(6) 2019 A SJM MASTERS SERIES HEART VALVE W/TEFLON CUFF WAS IMPLANTED IN THE DIABETIC AND DIALYSIS PATIENT, A SUPRA-ANNULAR POSITION USING NON-EVERTING MATTRESS SUTURE TECHNIQUE WITH THE PLEDGETS ON THE OUTFLOW SIDE. AN ABBOTT SIZER (MODEL#: 905) WAS USED IN THE SURGERY. IT WAS CONFIRMED BY THE X-RAY FLUOROSCOPY DURING A FOLLOW-UP THAT ONE OF THE LEAFLETS ON THE VALVE WAS NOT MOVING, AND ON (B)(6) 2020, A RE-DO AVR WAS PERFORMED. THE 25MM SJM MASTER SERIES HEART VALVE W/ TEFLON CUFF WAS EXPLANTED AND REPLACED WITH A 23MM SJM MASTER SERIES HEART VALVE W/TEFLON CUFF (SERIAL # (B)(4)) . UPON EXPLANT OF THE VALVE, THROMBUS WAS CONFIRMED TO BE ADHERED AT THE PIVOT GUARD HINGE, IMPEDING THE LEAFLET MOBILITY. IT WAS REPORTED THAT INTERNATIONAL NORMALIZED RATIO OF PROTHROMBIN TIME (INR) WAS MAINTAINED AT 2.1 BEFORE THE RE-DO AVR, AND THE PATIENT HAS HAD BEEN ADMINISTERED ANTICOAGULANTS. THE PATHOLOGICAL EXAMINATION CONCLUDED THAT THE ISSUE WAS DUE TO THROMBUS, BUT THE SURGEON REQUESTED PANNUS FORMATION ON THE VALVE, ALSO.
Patient problems: Thrombus
MAUDE report 10201349 ↗InjurySep 11, 2018Obstruction of Flow; Obstruction of Flow
ON (B)(6) 2016, A 25MM MASTERS HEART VALVE WAS IMPLANTED IN THE MITRAL POSITION. ON (B)(6) 2018, A RE-DO MVR WAS PERFORMED DUE TO PERIVALVULAR LEAKAGE (PVL) CAUSED BY MITRAL ANNULAR CALCIFICATION (MAC) BETWEEN THE PATIENT'S NATIVE ANNULUS AND THE MIDDLE OF THE POSTERIOR WALL. THE DEVICE WAS EXPLANTED AND REPLACED WITH A 27MM MASTERS HEART VALVE (SERIAL NUMBER UNKNOWN). THE PHYSICIAN ATTRIBUTES THE PATIENT'S DISEASE TO HAVE CAUSED THIS EVENT, AND NOT RELATED TO THE VALVE.
Patient problems: Calcium Deposits/Calcification; Mitral Regurgitation
MAUDE report 7864465 ↗
Other HEART-VALVE, MECHANICAL devices
- On-X Mitral Heart Valve with Conform-X Sewing Ring ONXMC-25/33On-X Life Technologies, Inc. · Heart-Valve, Mechanical1 recall1 death34 reports
- SJM Regent™ AGFN-756St. Jude Medical, Inc. · HEART-VALVE, MECHANICAL3 deaths70 reports
- SJM Regent™ AGFN-756St. Jude Medical, Inc. · HEART-VALVE, MECHANICAL5 deaths66 reports
- SJM Regent™ 23AGFN-756St. Jude Medical, Inc. · HEART-VALVE, MECHANICAL3 deaths44 reports
- CARBOMEDICS SUPRA-ANNULAR (TOP HAT) S5-021Corcym Srl · HEART-VALVE, MECHANICAL7 deaths37 reports
- Regent™ 21AGFN-756St. Jude Medical, Inc. · HEART-VALVE, MECHANICAL3 deaths35 reports
- Regent™ 23AGFN-756St. Jude Medical, Inc. · HEART-VALVE, MECHANICAL2 deaths33 reports
- Regent™ 19AGFN-756St. Jude Medical, Inc. · HEART-VALVE, MECHANICAL31 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page