SJM™ Masters Series 21AECJ-502Recalls & reported problems
St. Jude Medical, Inc. · Aortic Expanded Rotatable Mechanical Heart Valve · Class 3 · Implant
All models of SJM Masters Series (27)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 1
- Injuries reported
- 7
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 05414734053884), received between Oct 16, 2018 and Mar 29, 2023 — classified as 1 death, 7 injuries, 1 malfunction.
The most frequently reported specific device problem is “Fracture” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Fracture4
- Device Dislodged or Dislocated2
- Material Separation2
- Incomplete Coaptation1
- Intravalvular regurgitation1
- Material Split, Cut or Torn1
- Obstruction of Flow1
Patient problems
- No Known Impact Or Consequence To Patient3
- No Clinical Signs, Symptoms or Conditions2
- Aortic Valve Insufficiency/ Regurgitation1
- Death1
- Low Blood Pressure/ Hypotension1
- No Consequences Or Impact To Patient1
- Thrombus1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
InjuryMar 29, 2023Intravalvular regurgitation; Intravalvular regurgitation
IT WAS REPORTED THAT ON (B)(6) 2023, A 21MM AORTIC EXPD CUFF MASTERS WAS SUCCESSFULLY IMPLANTED. ON (B)(6) 2023, IT WAS REPORTED THAT THE PATIENT WAS ADMITTED WITH SEVERE AORTIC REGURGITATION. IT IS UNKNOWN IF AN INTERVENTION WAS PERFORMED. PATIENT STATUS IS UNKNOWN, NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation
MAUDE report 16640524 ↗InjuryFeb 23, 2022Material Separation; Material Separation
IT WAS REPORTED, A 21MM SJM MASTERS SERIES VALVE EXPANDED CUFF WAS CHOSEN FOR PROCEDURE. DURING THE PROCEDURE WHILE THE USER WAS POSITIONING THE DEVICE AND ROTATED THE VALVE, IT WAS REPORTED THAT THE USER APPLIED INCREASED TENSION TO THE VALVE AS THE IT WAS NOT ROTATING PROPERLY AND THE LEAFLET BROKE. THE DEVICE WAS REMOVED AND EXCHANGED WITH A NEW 21MM SJM MASTERS SERIES VALVE EXPANDED CUFF. NO ADDITIONAL INFORMATION HAS BEEN PROVIDED. THE EVENT DID CAUSE A REPORTED EXTRA 45 MINUTES ON CARDIOPULMONARY BYPASS AND WAS STABLE THROUGHOUT THE PROCEDURE. NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13587924 ↗MalfunctionFeb 17, 2021Material Separation; Device Dislodged or Dislocated
ON (B)(6) 2021, A 21 MM SJM MASTERS SERIES VALVE EXPANDED CUFF WAS SELECTED FOR IMPLANT. WHILE PLACING THE VALVE ONE OF THE LEAFLETS DISLODGED. THE VALVE WAS SUCCESSFULLY REMOVED AND REPLACED WITH A 19 MM SJM MASTERS SERIES VALVE EXPANDED CUFF. THERE WAS A CLINICALLY SIGNIFICANT DELAY IN PROCEDURE. THE PATIENT WAS REPORTED TO BE IN STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11337251 ↗InjuryJan 2, 2020Fracture; Fracture
ON AN UNKNOWN DATE, A 21MM MASTERS MECHANICAL HEART VALVE WAS SELECTED FOR IMPLANT. DURING IMPLANT, THE VALVE BROKE. ADDITIONAL INFORMATION WAS REQUESTED, BUT NO FURTHER INFORMATION WAS AVAILABLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9540565 ↗InjuryNov 4, 2019Fracture; Device Dislodged or Dislocated; Fracture; Device Dislodged or Dislocated
ON (B)(6) 2019, A 21 MM MASTERS VALVE WAS SELECTED FOR IMPLANT. WHILE ROTATING THE VALVE, THE LEAFLET FRACTURED. THE DEVICE WAS EXPLANTED AND REPLACED WITH A 19 MM MASTERS VALVE. THE PATIENT REMAINED HEMODYNAMICALLY STABLE BUT THERE WAS A CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9271281 ↗InjurySep 17, 2019Obstruction of Flow; Obstruction of Flow
ON (B)(6) 2018, A 21MM MASTERS VALVE WAS IMPLANTED. ON (B)(6) 2019, THE PATIENT PRESENTED WITH LEAFLET OBSTRUCTION AND THE PHYSICIAN ELECTED TO EXCHANGE THE VALVE FOR A 21MM MASTERS VALVE. THE PHYSICIAN NOTED THERE WAS THROMBUS PRESENT ON THE EXPLANTED VALVE. THE PATIENT IS REPORTED TO BE DISCHARGED.
Patient problems: Thrombus
MAUDE report 9076241 ↗InjuryJun 11, 2019Fracture; Fracture
ON (B)(6) 2019, A 21MM MASTERS VALVE WAS SELECTED FOR IMPLANT IN THE AORTIC POSITION. DURING ROTATION, A LEAFLET OF THE VALVE FRACTURED. ALL FRAGMENTS WERE REMOVED FROM THE PATIENT, AND THERE WAS NO EXTENT TO PROCEDURE TIME NOR ADVERSE CONSEQUENCES TO THE PATIENT. ANOTHER 21MM MASTERS VALVE (SN: (B)(4)) WAS IMPLANTED. THE PATIENT IS REPORTED TO BE STABLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8688146 ↗DeathApr 29, 2019Incomplete Coaptation; Incomplete Coaptation
ON (B)(6) 2019, A 21MM MASTERS SERIES VALVE WAS IMPLANTED IN A (B)(6) FEMALE. ON (B)(6) 2019, THE PATIENT COLLAPSED IN THE ICU. AN ECHO PERFORMED REVEALED ONE OF THE LEAFLETS WAS NOT MOVING AND THE OTHER WAS PARTIALLY COAPTING, RESULTING IN A HIGH GRADIENT AND HYPOTENSION. THE PATIENT'S NATIVE ANNULUS WAS REPORTED TO BE SEVERELY CALCIFIED. A RE-DO PROCEDURE WAS PERFORMED AND THE PATIENT WAS UNABLE TO SUSTAIN THE OPERATION AND EXPIRED ON (B)(6) 2019, DUE TO PERSISTENT HYPOTENSION AND RENAL FAILURE.
Patient problems: Death; Low Blood Pressure/ Hypotension
MAUDE report 8562210 ↗InjuryOct 16, 2018Fracture; Material Split, Cut or Torn; Fracture; Material Split, Cut or Torn
ON (B)(6) 2018, A 21MM MASTERS VALVE WAS SELECTED. WHEN THE VALVE WAS IMPLANTED AND ROTATED WITH THE RECOMMENDED VALVE ROTATOR, THE LEAFLET FRACTURED INTO TWO PIECES DURING THE ROTATION OF THE VALVE. THE VALVE AND PIECES WERE REMOVED AND A SECOND 21MM MASTERS VALVE WAS USED TO COMPLETE THE PROCEDURE. THE PROCEDURE WAS EXTENDED BY HALF AN HOUR, BUT THE PATIENT DID NOT EXPERIENCE ANY ADVERSE CONSEQUENCES AND IS STABLE AND DISCHARGED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7970642 ↗
Other HEART-VALVE, MECHANICAL devices
- On-X Mitral Heart Valve with Conform-X Sewing Ring ONXMC-25/33On-X Life Technologies, Inc. · Heart-Valve, Mechanical1 recall1 death34 reports
- SJM Regent™ AGFN-756St. Jude Medical, Inc. · HEART-VALVE, MECHANICAL3 deaths70 reports
- SJM Regent™ AGFN-756St. Jude Medical, Inc. · HEART-VALVE, MECHANICAL5 deaths66 reports
- SJM Regent™ 23AGFN-756St. Jude Medical, Inc. · HEART-VALVE, MECHANICAL3 deaths44 reports
- CARBOMEDICS SUPRA-ANNULAR (TOP HAT) S5-021Corcym Srl · HEART-VALVE, MECHANICAL7 deaths37 reports
- Regent™ 21AGFN-756St. Jude Medical, Inc. · HEART-VALVE, MECHANICAL3 deaths35 reports
- Regent™ 23AGFN-756St. Jude Medical, Inc. · HEART-VALVE, MECHANICAL2 deaths33 reports
- Regent™ 19AGFN-756St. Jude Medical, Inc. · HEART-VALVE, MECHANICAL31 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page