Response™ 401353Recalls & reported problems
St. Jude Medical Cardiovascular Division · Electrophysiology Catheter · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 05414734200899), received between Jan 2, 2020 and May 7, 2022 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
1 other recall cover devices cleared under the same FDA submission (K894500); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Break1
- Material Frayed1
- Signal Artifact/Noise1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Cardiac Perforation1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 7, 2022Signal Artifact/Noise; Signal Artifact/Noise
RELATED MANUFACTURING REF: 2030404-2022-00037. DURING THE SUPRAVENTRICULAR TACHYCARDIA PROCEDURE, INTERFERENCE WAS NOTED WHICH CAUSED A DELAY. SIGNAL INTERFERENCE WAS NOTED DURING THE PROCEDURE AND THE FOURTH ELECTRODE APPEARED DISTORTED IN THE SYSTEM. AFTER MULTIPLE ATTEMPTS OF CATHETER INTRODUCTION AND WITHDRAWAL, THE CATHETER STILL APPEARED DISTORTED IN THE SYSTEM. ANOTHER CATHETER WAS USED WHICH HAD THE SAME ISSUE. A THIRD CATHETER WAS USED TO CONTINUE THE PROCEDURE, HOWEVER THIS CAUSED A DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14319759 ↗MalfunctionMar 17, 2021Material Frayed; Material Frayed
DURING A PREMATURE VENTRICULAR CONTRACTION ABLATION PROCEDURE, THE CATHETER WAS NOTED TO HAVE EXPOSED WIRING. THE DEVICE WAS REPLACED AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11502785 ↗InjuryMay 14, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
DURING A LEFT VENTRICULAR CRT-D IMPLANT PROCEDURE, A POSSIBLE CARDIAC PERFORATION OCCURRED. IT WAS NOT POSSIBLE TO CANNULATE THE CORONARY SINUS DUE TO THE PATIENT'S ANATOMY. A POSSIBLE PERFORATION WAS OBSERVED USING CONTRAST NEAR THE ORIFICE OF THE CORONARY SINUS AND THE PROCEDURE WAS STOPPED. THE PATIENT WAS IMPLANTED WITH A RIGHT ATRIAL AND RIGHT VENTRICULAR LEAD AND A LEFT VENTRICULAR LEAD WILL BE IMPLANTED AT A LATER DATE. THE PATIENT IS IN STABLE CONDITION. THERE WERE NO PERFORMANCE ISSUES WITH ANY ABBOTT DEVICE.
Patient problems: Cardiac Perforation
MAUDE report 10055647 ↗MalfunctionJan 2, 2020Break; Break
THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS CONFIRMING A FRACTURED CATHETER SHAFT WITH EXPOSED INTERNAL WIRING.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9540672 ↗
FDA clearance and classification
Cleared through 510(k) K894500 on Sep 18, 1989 (decision: Substantially Equivalent). Submitted by Daig Corp. as “Daig Electrophysiology Catheter”. The FDA review took 61 days. FDA record ↗
The FDA regulates product code DRF (catheter, electrode recording, or probe, electrode recording) under 21 CFR 870.1220 as a Class II device. 21 CFR 870.1220 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K894500. One submission often covers many models, so check the product and lot details in each record.
- Z-0177-2015 · Oct 14, 2014 · TerminatedSt. Jude Medical, Fixed Curve Response 6F Electrophysiology Catheter, CSL, Model 401353
Common questions about the Response™ 401353
Has the Response™ 401353 been recalled?
No FDA recall record in our data names this exact device identifier (DI 05414734200899), based on the FDA recall file dated Nov 27, 2026. However, 1 recall involves devices cleared under the same FDA submission. It may or may not include this model, so check the product, model and lot details in that record.
What problems have been reported with the Response™ 401353?
The FDA's MAUDE database lists 4 adverse event reports for this device, received between Jan 2, 2020 and May 7, 2022. In those reports, the most common specific device problems are “Break” (1), “Material Frayed” (1) and “Signal Artifact/Noise” (1); the most common patient problem is “Cardiac Perforation” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Response™ 401353?
None of the reports are classified as deaths, 1 as an injury and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Response™ 401353 FDA approved?
It was cleared through 510(k) K894500 on Sep 18, 1989. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the Response™ 401353, and is it still sold?
The FDA's device database lists St. Jude Medical Cardiovascular Division as the company responsible for it (brand Response™, model 401353). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING devices
- IntellaMap Orion™ M004RC64S0Boston Scientific Corporation · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING6 deaths356 reports
- Advisor™ HD Grid, Sensor Enabled™ D-AVHD-DF16St. Jude Medical, Inc. · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING6 deaths356 reports
- ACHIEVE™ 990063-020Medtronic, Inc. · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING7 deaths129 reports
- Inquiry™ IBI-81104Irvine Biomedical, Inc. · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING1 recall2 deaths64 reports
- DECANAV R7F282CTBiosense Webster Inc · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING4 deaths108 reports
- LASSO NAV D134301Biosense Webster Inc · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING2 deaths82 reports
- DECANAV R7D282CTBiosense Webster Inc · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING1 death76 reports
- ACHIEVE ADVANCE™ 2ACH20Medtronic, Inc. · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING3 deaths69 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page