Skip to content

Supreme™ 401442Recalls & reported problems

St. Jude Medical Cardiovascular Division · Electrophysiology Catheter · Class 2

All models of Supreme (16)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
1
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 05414734201360), received between Jun 22, 2018 and Feb 18, 2020 — classified as 1 death, 2 injuries.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

1 other recall cover devices cleared under the same FDA submission (K894500); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Detachment of Device or Device Component1

Patient problems

  • Cardiac Perforation2
  • Death1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • DeathFeb 18, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DURING A CRT-D IMPLANTATION PROCEDURE A CARDIAC TAMPONADE AND SUBSEQUENT PATIENT EXPIRATION OCCURRED. THE DIAGNOSTIC CATHETER WAS USED FOR CORONARY SINUS (CS) CANNULATION. WHILE IMPLANTING THE LEADS THERE WAS DIFFICULTY ADVANCING PAST THE CS BUT EVENTUALLY THE LEADS WERE ABLE TO OVERPASS THE VALVE. ONCE PAST THE VALVE THE PATIENT BECAME HYPOTENSIVE AND AN ECHOCARDIOGRAM REVEALED BLOOD IN THE PERICARDIUM. A PERICARDIOCENTESIS WAS PERFORMED TO STABILIZE THE PATIENT HOWEVER CARDIOPULMONARY RESUSCITATION WAS NEEDED. THE PATIENT EXPIRED FOLLOWING INTERVENTION. THERE WERE NO PERFORMANCE ISSUES WITH ANY ABBOTT DEVICE.

    Patient problems: Cardiac Perforation; Death

    MAUDE report 9720001 ↗
  • InjurySep 11, 2019Detachment of Device or Device Component; Detachment of Device or Device Component

    MANUFACTURING RELATED REF: 3005334138-2019-00502. DURING THE PROCEDURE, WHEN ATTEMPTING TO REMOVE THE HD GRID CATHETER, IT BECAME ENTANGLED IN ANOTHER MAPPING CATHETER THAT WAS WITHIN THE RIGHT ATRIUM. THE HD GRID CATHETER WAS ABLE TO BE REMOVED BUT THE DISTAL PORTION OF THE OTHER MAPPING CATHETER BECAME DETACHED. ALL PORTIONS OF THE CATHETER WERE ABLE TO BE REMOVED FROM THE PATIENT WITH THE HD GRID CATHETER. THE HD GRID CATHETER WAS THEN ABLE TO BE USED TO CONTINUE AND COMPLETE THE PROCEDURE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8995930 ↗
  • InjuryJun 22, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    RELATED MANUFACTURER REPORT NUMBER: 9680001-2018-00084, 3005334138-2018-00174, 3008452825-2018-00195, 2182269-2018-00080, 3008452825-2018-00196, 3008452825-2018-00198. DURING A VENTRICULAR TACHYCARDIA PROCEDURE, A PERICARDIAL EFFUSION OCCURRED. AN ELDERLY PATIENT WHO WAS VERY PRO-ARRHYTHMIC DURING THE VT PROCEDURE, WITH A SYSTOLIC BLOOD PRESSURE (SBP) OF 100 FROM THE INITIATION OF THE PROCEDURE. WHILE THE TACTICATH WAS BEING MANIPULATED IN THE VENTRICLE, THE PATIENT'S SBP DECREASED TO 85-90, THE PROCEDURE WAS PAUSED AND AFTER 30 MINUTES THE BP REMAINED LOW AN ECHOCARDIOGRAM REVEALED A SMALL PERICARDIAL EFFUSION. WHILE PREPPING TO PERFORM A PERICARDIOCENTESIS THE PATIENT'S BP DROPPED AND THE PATIENT DECOMPENSATED. A CATHETER WAS INTRODUCED INTO THE CORONARY SINUS FOR PACING AND A PERICARDIOCENTESIS AND BLOOD TRANSFUSION WERE PERFORMED TO STABILIZE THE PATIENT. THERE WERE NO PERFORMANCE ISSUES WITH ANY ABBOTT DEVICE.

    Patient problems: Cardiac Perforation

    MAUDE report 7630752 ↗

FDA clearance and classification

Cleared through 510(k) K894500 on Sep 18, 1989 (decision: Substantially Equivalent). Submitted by Daig Corp. as “Daig Electrophysiology Catheter”. The FDA review took 61 days. FDA record ↗

The FDA regulates product code DRF (catheter, electrode recording, or probe, electrode recording) under 21 CFR 870.1220 as a Class II device. 21 CFR 870.1220 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K894500. One submission often covers many models, so check the product and lot details in each record.

  • Z-0177-2015 · Oct 14, 2014 · Terminated
    St. Jude Medical, Fixed Curve Response 6F Electrophysiology Catheter, CSL, Model 401353

Common questions about the Supreme™ 401442

Has the Supreme™ 401442 been recalled?

No FDA recall record in our data names this exact device identifier (DI 05414734201360), based on the FDA recall file dated Nov 27, 2026. However, 1 recall involves devices cleared under the same FDA submission. It may or may not include this model, so check the product, model and lot details in that record.

What problems have been reported with the Supreme™ 401442?

The FDA's MAUDE database lists 3 adverse event reports for this device, received between Jun 22, 2018 and Feb 18, 2020. In those reports, the most common specific device problem is “Detachment of Device or Device Component” (1); the most common patient problems are “Cardiac Perforation” (2) and “Death” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Supreme™ 401442?

1 report is classified as a death, 2 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Supreme™ 401442 FDA approved?

It was cleared through 510(k) K894500 on Sep 18, 1989. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the Supreme™ 401442, and is it still sold?

The FDA's device database lists St. Jude Medical Cardiovascular Division as the company responsible for it (brand Supreme™, model 401442). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING devices

All CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page