Supreme™ 402004Recalls & reported problems
St. Jude Medical Cardiovascular Division · Electrophysiology Catheter · Class 2
All models of Supreme (16)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 05414734201995), received between Jun 25, 2018 and Oct 31, 2025 — classified as 3 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Fracture” (3 reports).
1 other recall cover devices cleared under the same FDA submission (K894500); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem3
- Fracture3
- Difficult to Insert1
- Material Twisted/Bent1
Patient problems
- Cardiac Perforation3
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 31, 2025Fracture; Fracture
DURING A SUPRAVENTRICULAR TACHYCARDIA PROCEDURE, THE CATHETER FRACTURED WHILE INSERTED IN THE PATIENT. THE CATHETER WAS PLACED IN THE RIGHT ATRIUM THROUGH THE SUPERIOR VENA CAVA WHEN IT WAS NOTICED ON X-RAY THAT THE EXTERNAL DISTAL PART OF THE CATHETER WAS FRACTURED WITH JUST THE INNER WIRE CONNECTING THE TWO PARTS. THE CATHETER WAS REPLACED AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23433495 ↗MalfunctionJun 24, 2024Fracture; Fracture
THIS REPORT IS TO ADVISE OF A FRACTURE FOUND IN THE CATHETER SHAFT DURING ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19601663 ↗MalfunctionFeb 7, 2024Fracture; Difficult to Insert; Material Twisted/Bent; Fracture; Difficult to Insert; Material Twisted/Bent
DURING THE ATRIOVENTRICULAR NODAL REENTRANT TACHYCARDIA (AVNRT) ABLATION PROCEDURE, WHEN INSERTING THE CATHETER THROUGH THE FEMORAL VEIN ACCESS, RESISTANCE WAS MET. MORE FORCE WAS USED WITH NO RESOLUTION. THE CATHETER WAS REMOVED AND IT WAS NOTED TO BE FRACTURED AND BENT. THE CATHETER AND INTRODUCER WERE BOTH REPLACED AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18659508 ↗InjuryMay 6, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
RELATED MANUFACTURING REF: 2182269-2022-00026, 3008452825-2022-00408. DURING A SUPRAVENTRICULAR ABLATION PROCEDURE WITH WOLFE PARKINSON WHITE SYNDROME, A PERICARDIAL EFFUSION WAS NOTED IN THE CORONARY SINUS. DURING THE PROCEDURE, ELECTRODE 7-8 OF THE STEERABLE CATHETER SHOWED NOISE AND STRETCHING ON THE SYSTEM SCREEN AND THE PHYSICIANS INVOLVED STATED HIGH IMPEDANCE WAS NOTED. THE SIGNALS OF CS 7-8 WERE HIDDEN FROM BOTH THE ENSITE PRECISION AND THE EP RECORDING SYSTEM TO CONTINUE THE PROCEDURE. ACCESS THE CORONARY SINUS (CS) WAS ATTEMPTED WITH THE STEERABLE CATHETER, AND SUBSEQUENTLY WITH THE HELP OF THE SL3 SHEATH. A NON ABBOTT GUIDEWIRE WAS THEN USED TO GUIDE THE CS CATHETER. AFTER THE CATHETERS WERE IN PLACE, THE STUDY WAS PERFORMED. THE PATIENT THEN BECAME RESTLESS AND STATS DROPPED TO APPROXIMATELY 70+. AN ECHO WAS THEN PERFORMED WHICH REVEALED AN EFFUSION. HOWEVER, THE PROCEDURE WAS ABANDONED MIDWAY THROUGH THE STUDY AS THE PATIENT EXPERIENCED A PERICARDIAL EFFUSION WHICH WAS CONFIRMED VIA ECHOCARDIOGRAM. A PERICARDIOCENTESIS WAS PERFORMED WHICH STABILIZED THE PATIENT.
Patient problems: Cardiac Perforation
MAUDE report 14306196 ↗InjuryAug 8, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
RELATED MANUFACTURING REF: 3005334138-2019-00421, 3005334138-2019-00422, 3008452825-2019-00385, 3008452825-2019-00387. DURING A PULMONARY VEIN ISOLATION PROCEDURE, A PERICARDIAL EFFUSION OCCURRED. FOLLOWING GEOMETRY COLLECTION AND ABLATION OF THE RIGHT-SIDED UPPER PULMONARY VEIN, THE PATIENT BECAME HYPOTENSIVE. AN ECHOCARDIOGRAM REVEALED AN EFFUSION. A TAMPONADE WAS DIAGNOSED AND A PERICARDIOCENTESIS WAS PERFORMED WHICH STABILIZED THE PATIENT. THERE WERE NO PERFORMANCE ISSUES WITH ANY ABBOTT DEVICES.
Patient problems: Cardiac Perforation
MAUDE report 8876582 ↗InjuryJun 25, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
RELATED MANUFACTURING REFERENCE: 3008452825-2018-00189, 3008452825-2018-00190, 3005334138-2018-00171, 3008452825-2018-00191, 3005334138-2018-00172, 2030404-2018-00046, 3005334138-2018-00173. DURING AN ATRIAL FIBRILLATION ABLATION PROCEDURE A CARDIAC PERFORATION OCCURRED. DURING ROUTINE ECHOCARDIOGRAPHY A SMALL EFFUSION WAS DIAGNOSED. THE PATIENT'S BLOOD PRESSURE WAS STABLE. OCTAPLEX WAS ADMINISTERED TO REVERSE ANTICOAGULANT THERAPY AND THE PATIENT WAS TRANSFERRED TO THE CORONARY CARE UNIT. THE PATIENT RECOVERED AND WAS STABLE. THERE WERE NO PERFORMANCE ISSUES WITH ANY ABBOTT DEVICES.
Patient problems: Cardiac Perforation
MAUDE report 7633426 ↗
FDA clearance and classification
Cleared through 510(k) K894500 on Sep 18, 1989 (decision: Substantially Equivalent). Submitted by Daig Corp. as “Daig Electrophysiology Catheter”. The FDA review took 61 days. FDA record ↗
The FDA regulates product code DRF (catheter, electrode recording, or probe, electrode recording) under 21 CFR 870.1220 as a Class II device. 21 CFR 870.1220 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K894500. One submission often covers many models, so check the product and lot details in each record.
- Z-0177-2015 · Oct 14, 2014 · TerminatedSt. Jude Medical, Fixed Curve Response 6F Electrophysiology Catheter, CSL, Model 401353
Common questions about the Supreme™ 402004
Has the Supreme™ 402004 been recalled?
No FDA recall record in our data names this exact device identifier (DI 05414734201995), based on the FDA recall file dated Nov 27, 2026. However, 1 recall involves devices cleared under the same FDA submission. It may or may not include this model, so check the product, model and lot details in that record.
What problems have been reported with the Supreme™ 402004?
The FDA's MAUDE database lists 6 adverse event reports for this device, received between Jun 25, 2018 and Oct 31, 2025. In those reports, the most common specific device problems are “Fracture” (3), “Difficult to Insert” (1) and “Material Twisted/Bent” (1); the most common patient problem is “Cardiac Perforation” (3). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Supreme™ 402004?
None of the reports are classified as deaths, 3 as injuries and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Supreme™ 402004 FDA approved?
It was cleared through 510(k) K894500 on Sep 18, 1989. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the Supreme™ 402004, and is it still sold?
The FDA's device database lists St. Jude Medical Cardiovascular Division as the company responsible for it (brand Supreme™, model 402004). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING devices
- IntellaMap Orion™ M004RC64S0Boston Scientific Corporation · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING6 deaths356 reports
- Advisor™ HD Grid, Sensor Enabled™ D-AVHD-DF16St. Jude Medical, Inc. · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING6 deaths356 reports
- ACHIEVE™ 990063-020Medtronic, Inc. · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING7 deaths129 reports
- Inquiry™ IBI-81104Irvine Biomedical, Inc. · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING1 recall2 deaths64 reports
- DECANAV R7F282CTBiosense Webster Inc · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING4 deaths108 reports
- LASSO NAV D134301Biosense Webster Inc · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING2 deaths82 reports
- DECANAV R7D282CTBiosense Webster Inc · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING1 death76 reports
- ACHIEVE ADVANCE™ 2ACH20Medtronic, Inc. · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING3 deaths69 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page