Safire™ 402807Recalls & reported problems
St. Jude Medical Cardiovascular Division · Bi-Directional Ablation Catheter · Class 3
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 05414734205986), received between Jul 28, 2022 and Mar 19, 2025 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Activation, Positioning or Separation Problem” (2 reports).
1 other recall cover devices cleared under the same FDA submission (P960016); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Activation, Positioning or Separation Problem2
- High impedance2
- Signal Artifact/Noise2
- Failure to Deliver Energy1
- Use of Device Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 19, 2025Activation, Positioning or Separation Problem; Activation, Positioning or Separation Problem
RELATED MANUFACTURING REF: 2182269-2025-00005. DURING A PAROXYSMAL SUPRAVENTRICULAR TACHYCARDIA PROCEDURE, THE CATHETERS COULD NOT BE STRAIGHTENED CAUSING A PROCEDURE DELAY. WHEN BENDING THE CATHETER, IT COULD NOT BE STRAIGHTENED. THE CATHETER WAS REPLACED AND THE SAME ISSUE OCCURRED. A THIRD CATHETER WAS USED TO COMPLETE THE PROCEDURE WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21641758 ↗MalfunctionMar 19, 2025Activation, Positioning or Separation Problem; Activation, Positioning or Separation Problem
RELATED MANUFACTURING REF: 2182269-2025-00004. DURING A PAROXYSMAL SUPRAVENTRICULAR TACHYCARDIA PROCEDURE, THE CATHETERS COULD NOT BE STRAIGHTENED CAUSING A PROCEDURE DELAY. WHEN BENDING THE CATHETER, IT COULD NOT BE STRAIGHTENED. THE CATHETER WAS REPLACED AND THE SAME ISSUE OCCURRED. A THIRD CATHETER WAS USED TO COMPLETE THE PROCEDURE WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21641739 ↗MalfunctionMay 1, 2024Signal Artifact/Noise; Signal Artifact/Noise
DURING THE ATRIOVENTRICULAR NODAL REENTRANT TACHYCARDIA (AVNRT) PROCEDURE, SIGNAL ISSUES RESULTED IN A PROCEDURAL DELAY. WHEN THE CATHETER WAS INSERTED INTO THE PATIENT, NOISE WAS NOTED ON THE DISTAL ELECTRODE AND WAS PRESENT ON BOTH THE ENSITE SYSTEM AND THE NON-ABBOTT SYSTEM (GE RECORDING SYSTEM). THE CABLE WAS REPLACED BUT THE ISSUE WAS NOT RESOLVED. THE CATHETER WAS REPLACED WITH ANOTHER SAFIRE BI-DIRECTIONAL ABLATION CATHETER, BUT THE ISSUE PERSISTED. THE SECOND CATHETER WAS REPLACED WITH A TACTICATH SE ABLATION CATHETER WHICH RESOLVED THE ISSUE. THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19226047 ↗MalfunctionMay 1, 2024Signal Artifact/Noise; Signal Artifact/Noise
DURING THE ATRIOVENTRICULAR NODAL REENTRANT TACHYCARDIA (AVNRT) PROCEDURE, SIGNAL ISSUES RESULTED IN A PROCEDURAL DELAY. WHEN THE CATHETER WAS INSERTED INTO THE PATIENT, NOISE WAS NOTED ON THE DISTAL ELECTRODE AND WAS PRESENT ON BOTH THE ENSITE SYSTEM AND THE NON-ABBOTT SYSTEM (GE RECORDING SYSTEM). THE CABLE WAS REPLACED BUT THE ISSUE WAS NOT RESOLVED. THE CATHETER WAS REPLACED WITH ANOTHER SAFIRE BI-DIRECTIONAL ABLATION CATHETER, BUT THE ISSUE PERSISTED. THE SECOND CATHETER WAS REPLACED WITH A TACTICATH SE ABLATION CATHETER WHICH RESOLVED THE ISSUE. THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19225969 ↗MalfunctionNov 23, 2022Failure to Deliver Energy; High impedance; Failure to Deliver Energy; High impedance
RELATED MANUFACTURING REFERENCE: 2182269-2022-00063. DURING THE SVT PROCEDURE, IT WAS NOTED THAT THE CATHETER DISPLAYED A HIGH IMPEDANCE. THE DEVICE WAS REPLACED, AND LOW NO POWER AND HIGH IMPEDANCE WERE NOTED. THE DEVICE WAS REPLACED A SECOND TIME, AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15852690 ↗MalfunctionNov 23, 2022High impedance; High impedance
RELATED MANUFACTURING REFERENCE: 2182269-2022-00064. DURING THE SVT PROCEDURE, IT WAS NOTED THAT THE CATHETER DISPLAYED A HIGH IMPEDANCE. THE DEVICE WAS REPLACED, AND LOW NO POWER AND HIGH IMPEDANCE WERE NOTED. THE DEVICE WAS REPLACED A SECOND TIME, AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15852673 ↗MalfunctionJul 28, 2022Use of Device Problem; Use of Device Problem
DURING THE ATRIAL TACHYCARDIA PROCEDURE, AN EXPIRED DEVICE WAS USED. THERE WAS NO PERFORMANCE ISSUES NOTED AND THERE WAS NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15118647 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P960016. One submission often covers many models, so check the product and lot details in each record.
- Z-2313-2008 · Jun 5, 2008 · TerminatedThe St. Jude Medical Livewire TC" Ablation Catheter, 8mm Tip w/Thermocouple & Thermistor REF 402196, Length 115 cm, Lot 2056960, Manufacturer St. Jude Medical 14901 DeVeau Place, M
Other Cardiac ablation percutaneous catheter devices
- THERMOCOOL SMARTTOUCH SF D134805Biosense Webster Inc · Cardiac ablation percutaneous catheter1 recall176 deaths4,106 reports
- THERMOCOOL SMARTTOUCH SF D134804Biosense Webster Inc · Cardiac ablation percutaneous catheter44 deaths784 reports
- THERMOCOOL SMARTTOUCH SF D134801Biosense Webster Inc · Cardiac ablation percutaneous catheter15 deaths608 reports
- ThermoCool SmartTouch D132705Biosense Webster Inc · Cardiac ablation percutaneous catheter25 deaths574 reports
- THERMOCOOL SMARTTOUCH SF D134702Biosense Webster Inc · Cardiac ablation percutaneous catheter19 deaths366 reports
- THERMOCOOL SMARTTOUCH SF D134701Biosense Webster Inc · Cardiac ablation percutaneous catheter11 deaths278 reports
- ThermoCool SmartTouch D133602Biosense Webster Inc · Cardiac ablation percutaneous catheter8 deaths225 reports
- SMARTABLATE SAT001Biosense Webster Inc · Cardiac ablation percutaneous catheter185 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.