Skip to content

CPS DIRECT™ DS2C002Recalls & reported problems

St. Jude Medical, Inc. · Slittable Outer Guide Catheter · Class 2

All models of CPS DIRECT (16)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 05414734218207), received between Mar 7, 2019 and Mar 7, 2019 — classified as 1 injury.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • Vascular Dissection1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 7, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING A PROCEDURE FOR A NEW IMPLANT, WHEN ATTEMPTING TO CANNULATE THE CORONARY SINUS FOR LEAD PLACEMENT, THE PHYSICIAN INSERTED THE CATHETER INTO THE VESSEL, WHEN GLIDING THE SHEATH OVER, MOVEMENT (DEEP BREATHING) AND FORCE FROM THE SHEATH CONTACT CAUSED A CORONARY SINUS DISSECTION. THE PHYSICIAN BELIEVED HE MAY HAVE SELECTED THE WRONG CURVE OF SHEATH. THE LEADS WERE SUCCESSFULLY IMPLANTED AND THE PATIENT WAS MONITORED THROUGHOUT. A SUBSEQUENT ECHOCARDIOGRAM POST PROCEDURE CONFIRMED NO EFFUSION WAS PRESENT. THE PATIENT WAS STABLE WITH NO OTHER COMPLICATIONS.

    Patient problems: Vascular Dissection

    MAUDE report 8400688 ↗

Other CATHETER, PERCUTANEOUS devices

All CATHETER, PERCUTANEOUS devices →