EP-4™ H21911Recalls & reported problems
St. Jude Medical, Inc. · Touch Screen Computer · Class 2
All models of EP-4 (3)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 05414734219112), received between Jun 26, 2018 and May 17, 2023 — classified as 6 injuries, 6 malfunctions.
The most frequently reported specific device problem is “Smoking” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Smoking4
- No Tactile Prompts/Feedback2
- Communication or Transmission Problem1
- Improper or Incorrect Procedure or Method1
- Output Problem1
- Pacing Problem1
- Power Problem1
- Therapeutic or Diagnostic Output Failure1
Patient problems
- No Consequences Or Impact To Patient5
- No Clinical Signs, Symptoms or Conditions3
- Arrhythmia1
- Atrial Fibrillation1
- No Patient Involvement1
- Ventricular Fibrillation1
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
InjuryMay 17, 2023No Tactile Prompts/Feedback; No Tactile Prompts/Feedback
DURING AN ATRIAL FLUTTER PROCEDURE, THE STIMULATOR WAS UNRESPONSIVE TO A COMMAND TO STOP PACING RESULTING IN THE PATIENT EXPERIENCING ATRIAL FIBRILLATION WHICH REQUIRED CARDIOVERSION TO RESOLVE THE ATRIAL FIBRILLATION. THE STIMULATOR WAS TURNED OFF AND REBOOTED AND THE ISSUE WAS RESOLVED WITH NO FURTHER PATIENT CONSEQUENCES. THE PHYSICIAN WAS PACING DOWN TO A RATE IN THE 200S, THEN THE STIMULATOR SWITCHED TO PACING ON CHANNEL 3 AT 340 ON ITS OWN. WHEN THE COMMAND TO STOP PACING WAS GIVEN VIA THE KEYBOARD, IT DID NOT STOP PACING. IT WAS ATTEMPTED TO STOP PACING USING THE TOUCHSCREEN BUT THIS WAS ALSO UNSUCCESSFUL. WHILE THIS ISSUE WAS OCCURRING, THE SYSTEM AND TOUCHSCREEN DISPLAYS SHOWED A GREEN STIM BAR, INDICATING THE SYSTEM DID NOT THINK IT WAS PACING, EVEN THOUGH IT WAS. THE CONTINUED PACING CAUSED THE PATIENT TO BE PUT INTO ATRIAL FIBRILLATION. THE STIMULATOR WAS TURNED OFF WHICH STOPPED THE PACING. THE SYSTEM WAS REBOOTED AND THE ISSUE WAS RESOLVED. CARDIOVERSION WAS PERFORMED TO RESOLVE THE ATRIAL FIBRILLATION. THE PROCEDURE WAS COMPLETED WITH NO OTHER CONSEQUENCES.
Patient problems: Atrial Fibrillation
MAUDE report 16948120 ↗MalfunctionDec 20, 2022No Tactile Prompts/Feedback; No Tactile Prompts/Feedback
DURING AN ATRIAL FIBRILLATION PROCEDURE, IT WAS NOTED THAT THE EP-4 STIMULATOR WAS SHOWING AS OFF AND NOT RESPONDING TO TOUCHSCREEN INPUTS WHICH CAUSED THE PROCEDURE TO BE CANCELLED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16015272 ↗MalfunctionDec 5, 2021Power Problem; Power Problem
DURING THE ATRIAL FLUTTER PROCEDURE, THE EP4 TOUCHSCREEN KEPT LOOPING INTO WINDOWS BOOT SO THAT IT COULD NOT STIMULATE OR PACE. THE TOUCHSCREEN WAS BYPASSED AND CONNECTED DIRECTLY TO CLARIS CPU AND CONFIGURED TO WORK WITH CLARIS PROGRAM. THE CASE WAS CANCELLED DUE TO THIS ISSUE. A REPLACEMENT TOUCHSCREEN WAS RECEIVED AND INSTALLED TO RESOLVE THE ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12931247 ↗InjurySep 23, 2021Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
DURING THE PROCEDURE, THE EP-4 STIMULATOR STARTED TO PACE AND WAS THOUGHT TO HAVE INDUCED THE PATIENT INTO VENTRICULAR FIBRILLATION DUE TO THE ISSUE. THE UNINTENDED STIMULATION WAS ABLE TO BE STOPPED ONCE IT WAS DISCOVERED AND DEFIBRILLATION WAS USED TO RESTORE THE PATIENT INTO SINUS RHYTHM. THE INDUCTION OF THE VENTRICULAR FIBRILLATION WAS NOTED TO HAVE BEEN CAUSED BY THE USER AND AN IMPROPER WORKFLOW OF THE DEVICE AND THERE WAS NO PERFORMANCE ISSUE WITH THE DEVICE. FURTHER INFORMATION ON THE EVENT IS UNAVAILABLE DESPITE ATTEMPTS TO GATHER THE INFORMATION.
Patient problems: Arrhythmia
MAUDE report 12518050 ↗MalfunctionMar 8, 2021Output Problem; Output Problem
DURING A SUPRAVENTRICULAR TACHYCARDIA PROCEDURE, A DELAY IN THE PROCEDURE OCCURRED DUE TO AN ERROR MESSAGE DISPLAYED ON THE EP-4 TOUCHSCREEN. ALL CABLE CONNECTIONS WERE VERIFIED AND THE ISSUE REMAINED. THE PROCEDURE WAS COMPLETED THE SAME DAY FOLLOWING A VISIT FROM THE ENGINEER TO FIX A LOOSE POWER SUPPLY ON THE EP-4 STIMULATOR. THERE WERE NO CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11440177 ↗InjuryJul 24, 2019Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure
DURING A PULMONARY VEIN ISOLATION PROCEDURE, WITH THE PATIENT PREPPED FOR THE PROCEDURE, THE EP-4 TOUCHPAD DID NOT TURN ON DURING SET UP. A DIRECT CONNECTION TO THE COMPUTER WAS ATTEMPTED TO RESOLVE THE ISSUE BUT IT WAS DECIDED TO CANCEL THE CASE DUE TO NO STIMULATION FROM THE SYSTEM. THERE WERE NO ADVERSE PATIENT CONSEQUENCES.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8823133 ↗InjuryMay 6, 2019Smoking; Smoking
THE TOUCH SCREEN INTERFACE WAS SMOKING FROM THE BACK TOP RIGHT OF THE MONITOR PRIOR TO THE PROCEDURE. THE SYSTEM HAD BEEN ON FOR ABOUT 15-20 MINUTES AND ALL THE SUDDEN STARTED SMOKING. IT WAS UNPLUGGED FROM THE POWER AND BYPASSED DIRECTLY TO THE WORKMATE CPU AND THE CASE WAS COMPLETED WITH NO FURTHER ISSUES.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8582516 ↗InjuryApr 5, 2019Pacing Problem; Pacing Problem
RELATED MANUFACTURING REF: 2184149-2019-00040. NEAR THE END OF A PREMATURE VENTRICULAR CONTRACTION (PVC) PROCEDURE, THE TOUCHSCREEN COMPUTER DELIVERED SIX STIMULI TO THE PATIENT INSTEAD OF THREE AND INDUCED VENTRICULAR FIBRILLATION. THE PATIENT WENT INTO VENTRICULAR FIBRILLATION AND CARDIOVERSION WAS REQUIRED TO STABILIZE THE PATIENT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY AND THE PATIENT IS IN STABLE CONDITION.
Patient problems: Ventricular Fibrillation
MAUDE report 8487017 ↗InjuryJan 30, 2019Communication or Transmission Problem; Communication or Transmission Problem
DURING AN ACCESSORY PATHWAY ABLATION PROCEDURE, THE STIMULATOR DISPLAYED THE ERROR MESSAGE ¿EP-4 STIMULATOR HARDWARE IS NOT CONNECTED¿ AND THE PROCEDURE WAS CANCELLED. ONLY THE MAIN SCREEN DISPLAYED ON THE TOUCHSCREEN AND IT WAS NOT POSSIBLE TO CHANGE THE DIFFERENT STIMULATION PROTOCOLS SO THE PROCEDURE WAS RESCHEDULED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT DUE TO THE CANCELLATION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8288488 ↗MalfunctionAug 1, 2018Smoking; Smoking
WHILE POSITIONING THE CATHETERS AND COLLECTING GEOMETRY WITH THE MAPPING SYSTEM, THE TOUCH SCREEN COMPUTER SHUT DOWN AND EMITTED A BURNING SMELL AND SMOKE. THE PROCEDURE WAS COMPLETED WITHOUT THE TOUCH SCREEN COMPUTER AND THERE WERE NO PATIENT CONSEQUENCES.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7737941 ↗MalfunctionJul 18, 2018Smoking; Smoking
PRIOR TO THE PROCEDURE, WHEN THE TOUCHSCREEN COMPUTER WAS POWERED ON SMOKE WAS NOTED TO BE COMING FROM THE DEVICE. THERE WERE NO ADVERSE CONSEQUENCES TO THE STAFF AND NO PATIENT WAS INVOLVED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7697921 ↗MalfunctionJun 26, 2018Smoking; Smoking
DURING DEVICE PREPARATION, SMOKE EMITTED FROM THE BACK OF THE TOUCHSCREEN. THERE WAS NO PATIENT INVOLVED, THEREFORE THERE WERE NO ADVERSE PATIENT CONSEQUENCES.
Patient problems: No Patient Involvement
MAUDE report 7637483 ↗
FDA clearance and classification
Cleared through Special 510(k) K092913 on Oct 16, 2009 (decision: Substantially Equivalent). Submitted by Ep Medsystems as “Ep-4tm Cardiac Stimulator, Models: Ep-04-02, Ep-04-04”. The FDA review took 24 days. FDA record ↗
The FDA regulates product code JOQ (generator, pulse, pacemaker, external programmable) under 21 CFR 870.1750 as a Class II device. 21 CFR 870.1750 ↗
The FDA lists this type of device as life-sustaining or life-supporting.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the EP-4™ H21911
Has the EP-4™ H21911 been recalled?
No FDA recall record in our data names this exact device identifier (DI 05414734219112), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the EP-4™ H21911?
The FDA's MAUDE database lists 12 adverse event reports for this device, received between Jun 26, 2018 and May 17, 2023. In those reports, the most common specific device problems are “Smoking” (4), “No Tactile Prompts/Feedback” (2) and “Communication or Transmission Problem” (1); the most common patient problems are “Arrhythmia” (1), “Atrial Fibrillation” (1) and “Ventricular Fibrillation” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the EP-4™ H21911?
None of the reports are classified as deaths, 6 as injuries and 6 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the EP-4™ H21911 FDA approved?
It was cleared through Special 510(k) K092913 on Oct 16, 2009. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the EP-4™ H21911, and is it still sold?
The FDA's device database lists St. Jude Medical, Inc. as the company responsible for it (brand EP-4™, model H21911). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other GENERATOR, PULSE, PACEMAKER, EXTERNAL PROGRAMMABLE devices
- EP-4™ H403023St. Jude Medical, Inc. · Generator, pulse, pacemaker, external programmable (for electrophysiological studies only)42 reports
- EP-4™ Cardiac StimulatorSt. Jude Medical, Inc. · GENERATOR, PULSE, PACEMAKER, EXTERNAL PROGRAMMABLE3 reports
- StimLabMicropace Pty Limited · Generator, Pulse, Pacemaker, External Programmable1 report
- WorkMate Claris™ H701806St. Jude Medical, Inc. · Generator, pulse, pacemaker, external programmable (for electrophysiological studies only)1 report
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page