Durata™ 7120 (UDI 05414734502085)Recalls & reported problems

St. Jude Medical, Inc. · Defibrillation Lead · Class 3 · Implant

All models of Durata (28)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This St. Jude Medical Durata defibrillator lead has 346 adverse event reports on file, including 25 deaths and 189 injuries. Many reports describe the lead sensing electrical noise as heartbeats, high resistance in the lead, and breaks in it, and some mention shocks that were not needed. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
345
Deaths reported
25
Injuries reported
187
Recalls naming it
0

Summary

The FDA's MAUDE database contains 345 adverse event reports that list this exact device (DI 05414734502085), received between Jun 10, 2018 and Sep 10, 2026 — classified as 25 deaths, 187 injuries, 133 malfunctions.

The most frequently reported specific device problem is “Over-Sensing” (122 reports).

3 other recalls cover devices cleared under the same FDA submission (P950022); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 1 reportsJan 2024: 1 reportsFeb 2024: 2 reportsMar 2024: 5 reportsApr 2024: 4 reportsMay 2024: 0 reportsJun 2024: 4 reportsJul 2024: 1 reportsAug 2024: 2 reportsSep 2024: 4 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 3 reportsJan 2025: 3 reportsFeb 2025: 2 reportsMar 2025: 2 reportsApr 2025: 1 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 1 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 2 reportsJan 2026: 3 reportsFeb 2026: 1 reportsMar 2026: 6 reportsApr 2026: 2 reportsMay 2026: 2 reportsJun 2026: 2 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Over-Sensing122
  • Signal Artifact/Noise90
  • High impedance86
  • Adverse Event Without Identified Device or Use Problem45
  • Inappropriate/Inadequate Shock/Stimulation26
  • Break20
  • Capturing Problem20
  • Low impedance19

Patient problems

  • No Clinical Signs, Symptoms or Conditions154
  • No Known Impact Or Consequence To Patient105
  • Death21
  • Unspecified Infection15
  • Insufficient Information13
  • No Consequences Or Impact To Patient13
  • Therapeutic Effects, Unexpected11
  • Shock from Patient Lead(s)9

Recent adverse event reports

The 25 most recent of 345 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 10, 2026No Apparent Adverse Event

    Patient problems: Insufficient Information

    MAUDE report 50718194 ↗
  • MalfunctionJun 24, 2026High impedance

    IT WAS REPORTED THAT THE PATIENT PRESENTED REMOTELY VIA MERLIN.NET. REVIEW OF TRANSMISSION REVEALED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH DEFIBRILLATION IMPEDANCE. PROGRAMMING CHANGES WERE MADE. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25612269 ↗
  • MalfunctionJun 19, 2026No Apparent Adverse Event

    THIS REPORT IS TO ADVISE OF AN EVENT OCCURRING DURING ANALYSIS.

    Patient problems: Insufficient Information

    MAUDE report 25568472 ↗
  • MalfunctionMay 19, 2026Signal Artifact/Noise

    IT WAS REPORTED THAT THE PATIENT'S RIGHT VENTRICULAR (RV) LEAD EXHIBITED NOISE. FURTHER INFORMATION WAS REQUESTED BUT NOT YET RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25224871 ↗
  • DeathMay 5, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT EXPIRED. THERE IS NO KNOWN ALLEGATION FROM A HEALTH PROFESSIONAL THAT SUGGESTS THE DEATH WAS RELATED TO THE DEVICE. IT WAS REPORTED THAT THE CAUSE OF DEATH IS UNKNOWN. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Insufficient Information

    MAUDE report 25081507 ↗
  • InjuryApr 30, 2026Over-Sensing; Inappropriate/Inadequate Shock/Stimulation

    IT WAS REPORTED THAT THE PATIENT RECEIVED INAPPROPRIATE SHOCK DUE TO RIGHT VENTRICULAR (RV) LEAD NOISE. THE RV LEAD WAS CAPPED AND REPLACED ON (B)(6) 2026. THE PATIENT WAS IN STABLE CONDITION WITH NO ADVERSE EVENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25050324 ↗
  • InjuryApr 30, 2026Fracture

    IT WAS REPORTED THAT RIGHT VENTRICULAR (RV) LEAD WAS CAPPED AND REPLACED ON (B)(6) 2026, DUE TO FRACTURE. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25044353 ↗
  • InjuryMar 19, 2026Over-Sensing

    IT WAS REPORTED THAT THE PATIENT PRESENTED REMOTELY VIA MERLIN.NET DUE TO NOISE OVERSENSING EXHIBITED BY THEIR RIGHT VENTRICULAR (RV) LEAD. THE PATIENT WAS BROUGHT INTO CLINIC AND PROGRAMMING CHANGES WERE MADE. THE RV LEAD WAS THEN EXPLANTED AND SUCCESSFULLY REPLACED. THE PATIENT WAS STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24643328 ↗
  • InjuryMar 19, 2026Over-Sensing; Inappropriate/Inadequate Shock/Stimulation

    IT WAS REPORTED THAT INTERMITTENT OVERSENSING OF RIGHT VENTRICULAR (RV) LEAD NOISE WAS NOTED ON A REMOTE TRANSMISSION. THE PATIENT RECEIVED INAPPROPRIATE ANTI-TACHYCARDIA PACING (ATP) THERAPY AS A RESULT OF OVERSENSING THE RV LEAD NOISE. THE RV LEAD WAS CAPPED AND REPLACED WITH A COMPETITOR PRODUCT ON (B)(6) 2026. THE PATIENT WAS STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24639352 ↗
  • InjuryMar 17, 2026Capturing Problem

    IT WAS REPORTED THAT ELEVATED CAPTURE THRESHOLDS WERE OBSERVED ON THE RIGHT VENTRICULAR (RV) LEAD. OTHER PARAMETERS WERE STABLE. THE RV LEAD WAS CAPPED ON (B)(6) 2026 AND REPLACED. THE PATIENT WAS STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24622232 ↗
  • InjuryMar 9, 2026Failure to Capture

    IT WAS REPORTED DURING GENERATOR CHANGE THAT THE RIGHT VENTRICULAR LEAD DISPLAYED LOSS OF CAPTURE. THE LEAD WAS CAPPED ON (B)(6) 2026 AND SUCCESSFULLY REPLACED. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24547907 ↗
  • MalfunctionMar 3, 2026High impedance

    IT WAS REPORTED DURING REMOTE FOLLOW UP THAT THE RIGHT VENTRICULAR LEAD DISPLAYED HIGH DEFIBRILLATION IMPEDANCE. THE PRODUCT WILL CONTINUE TO BE MONITORED. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24503111 ↗
  • MalfunctionMar 3, 2026Signal Artifact/Noise

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED NOISE. NO INTERVENTION WAS REPORTED. THE PATIENT CONDITION WAS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24497577 ↗
  • InjuryFeb 2, 2026High impedance

    IT WAS REPORTED THAT A PATIENT PRESENTED WITH HIGH DEFIBRILLATION IMPEDANCE NOTED ON THE PATIENT'S RIGHT VENTRICULAR LEAD. A FRACTURE OF THE LEAD WAS NOT ALLEGED BY THE HEALTHCARE PROVIDER. NO INFORMATION REGARDING INTERVENTION OR ADVERSE PATIENT CONSEQUENCES WERE REPORTED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 24241697 ↗
  • MalfunctionJan 19, 2026Break; Over-Sensing

    IT WAS REPORTED THAT THE PATIENT PRESENTED REMOTELY VIA MERLIN.NET. REVIEW OF TRANSMISSION REVEALED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED NOISE OVER-SENSING, WHICH RESULTED IN PACING INHIBITION. NO INTERVENTION WAS PERFORMED. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24109508 ↗
  • MalfunctionJan 15, 2026Failure to Deliver Shock/Stimulation; Inappropriate/Inadequate Shock/Stimulation; Low impedance

    NEW INFORMATION RECEIVED INDICATES THAT THE PATIENT PRESENTED REMOTELY VIA MERLIN.NET. REVIEW OF TRANSMISSION REVEALED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) FAILED TO DELIVER HIGH VOLTAGE THERAPY DUE TO LOW, OUT OF RANGE DEFIBRILLATION IMPEDANCE. PROGRAMMING CHANGES WERE MADE. THE PATIENT WAS IN STABLE CONDITION

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24083093 ↗
  • MalfunctionJan 9, 2026High impedance

    IT WAS REPORTED THAT A PATIENT PRESENTED WITH HIGH IMPEDANCE NOTED ON THE RIGHT VENTRICULAR LEAD. FURTHER EXAMINATION WAS RECOMMENDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24031081 ↗
  • InjuryDec 9, 2025Break; Over-Sensing; Break; Over-Sensing

    NEW INFORMATION INDICATED THAT AN INSULATION BREACH WAS OBSERVED ON THE RIGHT VENTRICULAR (RV) LEAD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23756820 ↗
  • InjuryDec 5, 2025Over-Sensing; Over-Sensing

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) AND RIGHT VENTRICULAR (RV) LEADS EXHIBITED OVERSENSING. ON (B)(6) 2025, THE PHYSICIAN CAPPED THE RA AND RV LEAD. THE RV LEAD WAS REPLACED. THE PATIENT CONDITION WAS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23724763 ↗
  • InjuryAug 8, 2025Break; Failure to Capture; Fracture; Break; Failure to Capture; Fracture

    IT WAS REPORTED THE DURING A ROUTINE DEVICE CHANGE, RIGHT VENTRICULAR (RV) AND RIGHT ATRIAL (RA) LEADS WERE NOT CAPTURING. FRAYING WAS NOTED IN BOTH LEADS ON CHEST X-RAY. BOTH LEADS WERE EXPLANTED AND REPLACED. THE PATIENT IS IN STABLE CONDITION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22749394 ↗
  • MalfunctionJul 2, 2025Defective Device; Defective Device

    THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.

    Patient problems: Insufficient Information

    MAUDE report 22387417 ↗
  • MalfunctionJun 11, 2025Signal Artifact/Noise; Signal Artifact/Noise

    IT WAS REPORTED THAT THE PATIENT PRESENTED REMOTELY VIA MERLIN.NET. IT WAS NOTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED NOISE. NO INTERVENTION WAS PERFORMED. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22185820 ↗
  • InjuryMay 19, 2025Signal Artifact/Noise; High impedance; Signal Artifact/Noise; High impedance

    IT WAS REPORTED THAT HIGH PACING IMPEDANCE AND NOISE WAS NOTED ON THE RIGHT VENTRICULAR (RV) LEAD. ON (B)(6) 2025, THE PHYSICIAN ELECTED TO CAP AND REPLACE THE RV LEAD. THE PATIENT CONDITION WAS STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22048268 ↗
  • MalfunctionApr 18, 2025High impedance; High impedance

    IT WAS REPORTED A PATIENT'S RIGHT VENTRICULAR (RV) LEAD PRESENTED WITH HIGH DEFIBRILLATION IMPEDANCE. NO CHANGES WERE REPORTED. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21866931 ↗
  • MalfunctionMar 26, 2025Signal Artifact/Noise; Signal Artifact/Noise

    IT WAS REPORTED DURING REMOTE FOLLOW UP THAT THE RIGHT VENTRICULAR LEAD EXHIBITED NOISE. THE LEAD WILL CONTINUE TO BE MONITORED. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21698436 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P950022. One submission often covers many models, so check the product and lot details in each record.

  • Z-0660-2016 · Nov 3, 2015 · Terminated
    Optisure Dual Coil Defibrillation Leads, Model No. LDA220, LDA220Q, LDA230Q, LDP220Q. St. Jude Medical. Designed for long term attachment to an Implantable Car
  • Z-0457-2012 · Nov 28, 2011 · Terminated
    St. Jude Medical Riata (8Fr), Models: 1560, 1561, 1562, 1570, 1571, 1572, 1580, 1581, 1582, 1590, 1591, 1592 Product Usage: These Silicone transvenous high voltage (HV) leads co
  • Z-0458-2012 · Nov 28, 2011 · Terminated
    St. Jude Medical Riata ST Silicone Endocardial Defibrillation Leads Riata(7Fr), Models: 7000, 7001, 7002, 7010, 7011, 7040, 7041, 7042 Product Usage: These Silicone transvenous

Other Implantable cardioverter defibrillator (non-CRT) devices

All Implantable cardioverter defibrillator (non-CRT) devices →

Sources and updates