Durata™ 7121-65Recalls & reported problems

St. Jude Medical, Inc. · Defibrillation Lead · Class 3 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This St. Jude Medical Durata defibrillation lead has 427 reports, including 23 deaths and 231 injuries. Common problems described include the lead misreading heart signals, high electrical resistance, lead fractures, and failure to stimulate the heart, and some patients needed the lead removed and replaced. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
429
Deaths reported
23
Injuries reported
229
Recalls naming it
0

Summary

The FDA's MAUDE database contains 429 adverse event reports that list this exact device (DI 05414734502122), received between Jun 11, 2018 and Sep 16, 2026 — classified as 23 deaths, 229 injuries, 177 malfunctions.

The most frequently reported specific device problem is “Over-Sensing” (162 reports).

3 other recalls cover devices cleared under the same FDA submission (P950022); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 3 reportsDec 2023: 2 reportsJan 2024: 5 reportsFeb 2024: 2 reportsMar 2024: 3 reportsApr 2024: 3 reportsMay 2024: 5 reportsJun 2024: 1 reportsJul 2024: 3 reportsAug 2024: 0 reportsSep 2024: 4 reportsOct 2024: 3 reportsNov 2024: 3 reportsDec 2024: 2 reportsJan 2025: 1 reportsFeb 2025: 2 reportsMar 2025: 0 reportsApr 2025: 2 reportsMay 2025: 5 reportsJun 2025: 2 reportsJul 2025: 1 reportsAug 2025: 1 reportsSep 2025: 2 reportsOct 2025: 1 reportsNov 2025: 1 reportsDec 2025: 4 reportsJan 2026: 1 reportsFeb 2026: 4 reportsMar 2026: 4 reportsApr 2026: 4 reportsMay 2026: 2 reportsJun 2026: 1 reportsJul 2026: 1 reportsAug 2026: 2 reportsSep 2026: 3 reportsOct 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Over-Sensing162
  • High impedance121
  • Signal Artifact/Noise90
  • Adverse Event Without Identified Device or Use Problem45
  • Fracture39
  • Capturing Problem36
  • Inappropriate/Inadequate Shock/Stimulation31
  • Failure to Capture29

Patient problems

  • No Clinical Signs, Symptoms or Conditions225
  • No Known Impact Or Consequence To Patient112
  • Unspecified Infection16
  • No Consequences Or Impact To Patient15
  • Death14
  • Therapeutic Effects, Unexpected13
  • No Information11
  • Insufficient Information10

Recent adverse event reports

The 25 most recent of 429 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 16, 2026Over-Sensing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50794288 ↗
  • MalfunctionSep 10, 2026Over-Sensing; Capturing Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50721100 ↗
  • MalfunctionSep 3, 2026High impedance

    Patient problems: Insufficient Information

    MAUDE report 50599630 ↗
  • MalfunctionAug 25, 2026High impedance

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50438447 ↗
  • MalfunctionAug 21, 2026Low impedance

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50380476 ↗
  • MalfunctionJul 30, 2026Failure to Capture; Over-Sensing

    New information noted that the right ventricular lead exhibited loss of capture. The lead was explanted and replaced on (b)(6)2026. The patient was recovering.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50078447 ↗
  • MalfunctionJun 18, 2026Over-Sensing; No Apparent Adverse Event

    IT WAS REPORTED THAT THE PATIENT PRESENTED REMOTELY. A REMOTE TRANSMISSION SHOWED THAT THE PATIENT'S RIGHT VENTRICULAR (RV) LEAD EXHIBITED NOISE OVERSENSING. PROGRAMMING CHANGES WERE MADE. THE PATIENT HAD NO CONSEQUENCES DUE TO THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25548486 ↗
  • InjuryMay 18, 2026Failure to Capture; Fracture

    IT WAS REPORTED THAT RIGHT VENTRICULAR (RV) LEAD WAS EXPLANTED AND REPLACED DUE TO LEAD FRACTURE AND FAILURE TO CAPTURE. THE PATIENT IS RECOVERING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25201569 ↗
  • InjuryMay 14, 2026Failure to Capture; High impedance

    IT WAS REPORTED THAT THE PATIENT PRESENTED IN CLINIC. A DEVICE INTERROGATION SHOWED THAT THEIR RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH PACING IMPEDANCE CORRELATED WITH INCREASED CAPTURE THRESHOLD AND NONCAPTURE. THE RV LEAD WAS CAPPED AND REPLACED ON (B)(6) 2026. THE PATIENT WAS STABLE THROUGHOUT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25176821 ↗
  • MalfunctionApr 17, 2026Intermittent Capture

    IT WAS REPORTED THAT A PATIENT PRESENTED REMOTELY VIA MERLIN.NET. REVIEW OF THE TRANSMISSION REVEALED RIGHT VENTRICULAR (RV) LEAD EXHIBITED INTERMITTENT CAPTURE. NO CHANGES OR INTERVENTIONS WERE REPORTED. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24917607 ↗
  • InjuryApr 17, 2026Failure to Capture; High impedance

    IT WAS REPORTED THAT THE PATIENT'S RIGHT VENTRICULAR (RV) LEAD EXHIBITED A HIGH PACING IMPEDANCE AND FAILURE TO CAPTURE. THE RV LEAD WAS CAPPED ON (B)(6) 2026 AND SUCCESSFULLY REPLACED. THE PATIENT WAS DISCHARGED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24916132 ↗
  • InjuryApr 11, 2026High impedance; Over-Sensing; Capturing Problem

    IT WAS REPORTED THAT THE PATIENT PRESENTED TO THE HOSPITAL FOR RIGHT VENTRICULAR (RV) LEAD EXPLANT. DEVICE INTERROGATION REVEALED THAT THE RV LEAD EXHIBITED VARYING CAPTURE THRESHOLD, HIGH PACING IMPEDANCE, AND OVERSENSING NOISE. FLUOROSCOPY WAS PERFORMED, AND NO DEFECTS WERE OBSERVED. DURING THE PROCEDURE, INSULATION DAMAGE WAS NOTED ON THE RIGHT ATRIAL (RA) LEAD. BOTH LEADS WERE EXPLANTED AND REPLACED WITH NEW ONES. THE PATIENT WAS STABLE BEFORE, DURING, AND AFTER THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24849153 ↗
  • InjuryApr 8, 2026High impedance; Capturing Problem

    NEW INFORMATION RECEIVED NOTES THIS RIGHT VENTRICULAR (RV) LEAD ALSO EXHIBITED HIGH CAPTURE THRESHOLDS PRIOR TO EXPLANT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24824507 ↗
  • MalfunctionMar 30, 2026Over-Sensing

    IT WAS REPORTED THAT THE ASYMPTOMATIC PATIENT PRESENTED VIA MERLIN.NET. IT WAS NOTED THAT NOISE RESULTING IN OVERSENSING WAS OBSERVED ON THE RIGHT VENTRICULAR (RV) LEAD ON A STORED VENTRICULAR FIBRILLATION (VF) EPISODE. NO INAPPROPRIATE THERAPY WAS DELIVERED, AS A RETURN TO SINUS WAS DETECTED DURING CAPACITOR CHARGING. THE PATIENT PRESENTED FOR ADDITIONAL TESTING IN CLINIC. THE NOISE WAS NOT REPRODUCIBLE IN CLINIC. PROGRAMMING CHANGES WERE COMPLETED. THE PATIENT WAS JUST BEING MONITORED. THE PATIENT WAS STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24728398 ↗
  • InjuryMar 20, 2026Fracture; Over-Sensing

    IT WAS REPORTED THAT RIGHT VENTRICULAR (RV) LEAD WAS CAPPED AND REPLACED ON (B)(6) 2026 DUE FRACTURE. LEAD FRACTURE WAS CONFIRMED BY NOISE AND THE NOISE WAS OVER-SENSED. THE PATIENT IS IN STABLE CONDITION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24652929 ↗
  • InjuryMar 17, 2026Over-Sensing

    IT WAS REPORTED THAT THE ASYMPTOMATIC PATIENT PRESENTED FOR A FOLLOW UP IN CLINIC. IT WAS NOTED THAT NOISE RESULTING IN OVERSENSING WAS OBSERVED ON THE RIGHT VENTRICULAR (RV) LEAD. ALL OTHER ELECTRICAL VALUES WERE NORMAL. THE RV LEAD WAS CAPPED (B)(6) 2026 AND REPLACED. THE PATIENT WAS STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24618937 ↗
  • InjuryMar 11, 2026Signal Artifact/Noise; Impedance Problem

    IT WAS REPORTED DURING IN CLINIC FOLLOW UP THAT THE RIGHT VENTRICULAR LEAD DISPLAYED NOISE AND VARYING IMPEDANCE. THE PRODUCT WAS EXPLANTED AND SUCCESSFULLY REPLACED. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24572336 ↗
  • InjuryFeb 27, 2026Insufficient Device Problem Information

    IT WAS REPORTED THAT THE RV LEAD WAS SURGICALLY REPAIRED DUE TO UNSPECIFIED REASON. FURTHER INFORMATION WAS REQUESTED BUT NOT RECEIVED.

    Patient problems: Insufficient Information

    MAUDE report 24476696 ↗
  • InjuryFeb 5, 2026Fracture; High impedance; Over-Sensing

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD EXHIBITED OVERSENSING NOISE DUE TO A SUSPECTED LEAD FRACTURE. IT WAS NOTED THAT HIGH VOLTAGE LEAD IMPEDANCE HAD AN ABRUPT INCREASE AS WELL AND TACHY THERAPIES WERE TURNED OFF PRIOR TO LEAD REPLACEMENT. THE LEAD WAS CAPPED AND REPLACED ON (B)(6) 2026. THE PATIENT WAS IN STABLE CONDITION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24269198 ↗
  • InjuryFeb 4, 2026Over-Sensing

    IT WAS REPORTED THAT THE PATIENT PRESENTED IN CLINIC. A DEVICE INTERROGATION WAS PERFORMED AND REVEALED THAT THEIR RIGHT VENTRICULAR (RV) LEAD EXHIBITED NOISE OVERSENSING. THE RV LEAD WAS EXPLANTED AND REPLACED. THE PATIENT WAS STABLE THROUGHOUT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24259367 ↗
  • MalfunctionFeb 2, 2026High impedance

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH DEFIBRILLATION IMPEDANCE. NO INTERVENTION WAS REPORTED, AND THE PATIENT WILL CONTINUE TO BE MONITORED. THERE WERE NO PATIENT CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24241946 ↗
  • MalfunctionJan 7, 2026Over-Sensing

    IT WAS REPORTED THAT MYOPOTENTIALS OVERSENSING WAS NOTED ON THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD). PROGRAMMING CHANGES WERE PERFORMED TO RESOLVE THE ISSUE. THE PATIENT CONDITION WAS STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23995796 ↗
  • InjuryDec 30, 2025Signal Artifact/Noise; Over-Sensing; Signal Artifact/Noise; Over-Sensing

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD EXHIBITED A NOISE PROBLEM. NO INTERVENTION WAS REPORTED. PATIENT STATUS WAS NOT PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23928041 ↗
  • MalfunctionDec 23, 2025Signal Artifact/Noise; Signal Artifact/Noise

    IT WAS REPORTED THAT DURING A FOLLOW-UP CLINIC CHECK, NOISE WAS OBSERVED ON THE RIGHT VENTRICULAR (RV) LEAD. THE PATIENT WAS IN STABLE CONDITION WITH NO ADVERSE HEALTH CONSEQUENCES. NO ADDITIONAL INFORMATION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23878471 ↗
  • MalfunctionDec 11, 2025High impedance; High impedance

    IT WAS REPORTED THAT AN ALERT WAS OBSERVED NOTING THE HIGH VOLTAGE LEAD IMPEDANCE (HVLI) WAS HIGH ON THE RIGHT VENTRICULAR (RV) LEAD. UPON REVIEW, IT WAS NOTED THAT THE UPPER LIMIT FOR THE ALERT HAD BEEN SET TO 101 OHMS ON THIS PATIENT FOR UNKNOWN REASONS WHILE THE NORMAL UPPER LIMIT FOR (HVLI) IS 200 OHMS. THE LIMIT WAS TO BE DISCUSSED WITH THE PHYSICIAN. THE PATIENT WAS STABLE DURING CHECKUP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23775099 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P950022. One submission often covers many models, so check the product and lot details in each record.

  • Z-0660-2016 · Nov 3, 2015 · Terminated
    Optisure Dual Coil Defibrillation Leads, Model No. LDA220, LDA220Q, LDA230Q, LDP220Q. St. Jude Medical. Designed for long term attachment to an Implantable Car
  • Z-0458-2012 · Nov 28, 2011 · Terminated
    St. Jude Medical Riata ST Silicone Endocardial Defibrillation Leads Riata(7Fr), Models: 7000, 7001, 7002, 7010, 7011, 7040, 7041, 7042 Product Usage: These Silicone transvenous
  • Z-0457-2012 · Nov 28, 2011 · Terminated
    St. Jude Medical Riata (8Fr), Models: 1560, 1561, 1562, 1570, 1571, 1572, 1580, 1581, 1582, 1590, 1591, 1592 Product Usage: These Silicone transvenous high voltage (HV) leads co

Other Implantable cardioverter defibrillator (non-CRT) devices

All Implantable cardioverter defibrillator (non-CRT) devices →