Durata® 7131-65Recalls & reported problems
St. Jude Medical, Inc. · Defibrillation Lead · Class 3 · Implant
All models of Durata (28)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 05414734505123), received between Aug 5, 2018 and Mar 22, 2023 — classified as 4 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Over-Sensing” (4 reports).
3 other recalls cover devices cleared under the same FDA submission (P950022); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Over-Sensing4
- Failure to Capture2
- High Capture Threshold1
- High impedance1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Arrhythmia2
- Discomfort1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjuryMar 22, 2023Over-Sensing; Over-Sensing
IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD EXHIBITED OVERSENSING NOISE. THE LEAD WAS CAPPED AND REPLACED ON (B)(6) 2023. THE PATIENT WAS IN STABLE CONDITION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16595610 ↗InjuryJan 17, 2023Over-Sensing; Over-Sensing
RELATED MANUFACTURER REFERENCE NUMBER: 2017865-2023-01215. IT WAS REPORTED THAT A PATIENT PRESENTED IN-CLINIC. UPON EXAMINATION, IT WAS NOTED THAT THE PATIENT'S IMPLANTABLE CARDIOVERTER DEFIBRILLATOR WAS FOUND TO HAVE EXHIBITED AN UNSPECIFIED OUTPUT ISSUE, AND THE PATIENT'S RIGHT VENTRICULAR LEAD WAS FOUND TO HAVE OVER-SENSING OF NOISE. CORRECTIVE PROGRAMMING CHANGES WERE MADE, AND FURTHER SURGICAL INTERVENTION WAS PLANNED. THE PATIENT WAS STABLE THROUGHOUT.
Patient problems: Arrhythmia; Discomfort
MAUDE report 16180336 ↗MalfunctionNov 9, 2022Over-Sensing; Over-Sensing
IT WAS REPORTED THAT A PATIENT PRESENTED REMOTELY VIA MERLIN.NET. REVIEW OF THE TRANSMISSION REVEALED POST-PACED T-WAVE OVERSENSING ON THE RIGHT VENTRICULAR (RV) LEAD. NO CHANGES OR INTERVENTION WAS PERFORMED; THE DEVICE WILL CONTINUE TO BE MONITORED. THERE WERE NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15762806 ↗InjuryJun 9, 2021High impedance; High impedance
IT WAS REPORTED THAT A PATIENT PRESENTED IN-CLINIC. UPON INTERROGATION, THE PATIENT'S RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH DEFIBRILLATION IMPEDANCE. THE RV LEAD WAS CAPPED AND REPLACED ON (B)(6) 2021. THE PATIENT DID NOT SUFFER ANY ADVERSE CONSEQUENCES THROUGHOUT THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11967735 ↗InjuryJul 10, 2020Failure to Capture; Over-Sensing; Failure to Capture; Over-Sensing
RELATED MANUFACTURER REFERENCE NUMBER: 2017865-2020-08953. IT WAS REPORTED THAT PATIENT UNDERWENT A PROCEDURE UNRELATED TO THE DEVICES ON (B)(6) 2020. TOWARDS THE END OF THE PROCEDURE, PATIENT'S RIGHT VENTRICULAR LEAD EXHIBITED CROSSTALK OVER-SENSING AND FAILED TO CAPTURE. PHYSICIAN ATTEMPTED TO REPROGRAM THE DEVICE. PATIENT THEN WENT ASYSTOLE. DEVICE WAS REPROGRAMMED AGAIN AND CAPTURE RESUMED. FURTHER INVESTIGATION FOUND THAT PATIENT'S RIGHT ATRIAL LEAD MAY HAVE ALSO BEEN OVER-SENSING AND INHIBITING PACING. BOTH LEADS WERE CAPPED AND REPLACED ON (B)(6) 2020. PATIENT CONDITION AFTER THE PROCEDURE WAS STABLE.
Patient problems: Arrhythmia
MAUDE report 10258764 ↗MalfunctionAug 5, 2018Failure to Capture; High Capture Threshold; Failure to Capture; High Capture Threshold
IT WAS REPORTED THAT THE ASYMPTOMATIC PATIENT PRESENTED FOR SURGERY DUE TO A ROUTINE GENERATOR CHANGE. IT WAS NOTED THAT THE RIGHT VENTRICULAR LEAD EXHIBITED LOSS OF CAPTURE AND UNACCEPTABLE THRESHOLDS. THE LEAD WAS CAPPED 07/10/2018 AND REPLACED. THE PATIENT WAS STABLE BEFORE AND AFTER THE PROCEDURE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7749790 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P950022. One submission often covers many models, so check the product and lot details in each record.
- Z-0660-2016 · Nov 3, 2015 · TerminatedOptisure Dual Coil Defibrillation Leads, Model No. LDA220, LDA220Q, LDA230Q, LDP220Q. St. Jude Medical. Designed for long term attachment to an Implantable Car
- Z-0458-2012 · Nov 28, 2011 · TerminatedSt. Jude Medical Riata ST Silicone Endocardial Defibrillation Leads Riata(7Fr), Models: 7000, 7001, 7002, 7010, 7011, 7040, 7041, 7042 Product Usage: These Silicone transvenous
- Z-0457-2012 · Nov 28, 2011 · TerminatedSt. Jude Medical Riata (8Fr), Models: 1560, 1561, 1562, 1570, 1571, 1572, 1580, 1581, 1582, 1590, 1591, 1592 Product Usage: These Silicone transvenous high voltage (HV) leads co
Other Implantable cardioverter defibrillator (non-CRT) devices
- EMBLEM™ MRI S-ICD A219Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)2 recalls36 deaths15,404 reports
- EMBLEM™ S-ICD A209Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)1 recall7 deaths6,788 reports
- EMBLEM™ S-ICD 3501Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)7 deaths4,936 reports
- Cobalt™ XT DR MRI SureScan™ DDPA2D4Medtronic, Inc. · Implantable cardioverter defibrillator (non-CRT)3 recalls30 deaths1,603 reports
- ENDOTAK RELIANCE® G 0185Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)9 deaths1,632 reports
- VIGILANT™ EL ICD DR D233Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)9 deaths1,563 reports
- Evera MRI™ XT DR SureScan® DDMB1D4Medtronic, Inc. · Implantable cardioverter defibrillator (non-CRT)1 recall35 deaths1,426 reports
- Durata™ 7122St. Jude Medical, Inc. · Implantable cardioverter defibrillator (non-CRT)27 deaths1,377 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page