St. Jude Medical, Inc. · Slittable Outer Guide Catheter · Class 2

All models of CPS DIRECT (16)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
1
Injuries reported
8
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 05415067004208), received between Mar 7, 2019 and Nov 25, 2024 — classified as 1 death, 8 injuries.

The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem8
  • Improper or Incorrect Procedure or Method1

Patient problems

  • Vascular Dissection6
  • Cardiac Perforation4
  • Pericardial Effusion2
  • Arrhythmia1
  • Cardiac Tamponade1
  • Death1
  • Syncope/Fainting1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 25, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DURING THE INITIAL IMPLANT PROCEDURE, THE PATIENT DEVELOPED A PERFORATION OF THE EPICARDIUM WHILE MANEUVERING THE CATHETER. LATER DURING THE PROCEDURE, THE PERFORATION RESULTED IN A PERICARDIAL EFFUSION WHICH RESULTED IN THE PATIENT LOSING CONSCIOUSNESS AND CEASING BREATHING. CARDIOPULMONARY RESUSCITATION WAS ADMINISTERED UNTIL THE PATIENT STABILIZED. THE IMPLANT PROCEDURE WAS ABANDONED. NO ALLEGATION OF MALFUNCTION WAS MADE AGAINST THE CATHETER. LEAD IMPLANTATION IS STILL BEING CONSIDERED. HOWEVER, NO FURTHER INTERVENTION WAS PERFORMED AT THIS TIME. THE PATIENT WAS IN STABLE CONDITION.

    Patient problems: Cardiac Perforation; Pericardial Effusion; Syncope/Fainting

    MAUDE report 20775986 ↗
  • InjurySep 6, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DURING IMPLANT, IT WAS REPORTED THAT THE PATIENT EXPERIENCED CHEST PAIN AND BRADYCARDIA FOLLOWING A VASCULAR DISSECTION DUE TO USE OF THE CATHETER. A PERICARDIAL EFFUSION WAS OBSERVED. A PERICARDIOCENTESIS WAS PERFORMED. THE PATIENT WILL CONTINUE TO BE MONITORED.

    Patient problems: Arrhythmia; Cardiac Perforation; Pericardial Effusion; Vascular Dissection

    MAUDE report 20158180 ↗
  • InjuryJan 27, 2022Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method

    IT WAS REPORTED THAT DURING INITIAL IMPLANT, WHILE THE LEAD WAS BEING POSITIONED IN THE RIGHT ATRIUM (RA) A DISSECTION WAS NOTED DUE TO INCORRECT INSERTION TECHNIQUE USED BY THE PHYSICIAN. THE DISSECTION SPONTANEOUSLY RESOLVED AND NO ADDITIONAL INTERVENTION WAS REQUIRED. THE PATIENT WAS STABLE AND WILL CONTINUE TO BE MONITORED.

    Patient problems: Vascular Dissection

    MAUDE report 13366296 ↗
  • InjuryNov 18, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT PRESENTED TO THE HOSPITAL ON (B)(6) 2021. UPON EXAMINATION, IT WAS NOTED THAT THE CATHETER HAD DISSECTED THE CORONARY SINUS. NO INTERVENTION WAS PERFORMED. THE PATIENT WAS IN STABLE CONDITION.

    Patient problems: Vascular Dissection

    MAUDE report 12834167 ↗
  • InjuryApr 1, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING IMPLANT OF THE LEFT VENTRICULAR (LV) LEAD, DISSECTION OF THE CORONARY SINUS BY THE CATHETER OCCURRED. THE PROCEDURE WAS ENDED AND NO LV LEAD WAS IMPLANTED. NO INTERVENTION WAS PERFORMED TO TREAT THE DISSECTION; THE PATIENT WAS STABLE AND WILL CONTINUE TO BE MONITORED. THERE WERE NO ADVERSE CONSEQUENCES.

    Patient problems: Vascular Dissection

    MAUDE report 11605630 ↗
  • InjuryOct 23, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A CATHETER PUNCTURED A PATIENT'S HEART DURING A IMPLANTABLE CARDIOVERTER DEFIBRILLATOR UPGRADE PROCEDURE. THE PUNCTURE WAS DISCOVERED BY A FLUOROSCOPY. THE PATIENT WAS HEMODYNAMICALLY STABLE. CATHETER WAS REMOVED SAFELY. THE UPGRADE PROCEDURE WAS ABANDONED. PATIENT WAS STABLE AFTER THE PROCEDURE.

    Patient problems: Cardiac Perforation

    MAUDE report 10731292 ↗
  • DeathOct 10, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    RELATED MANUFACTURER REFERENCE NUMBER:2017865-2019-14938, 2017865-2019-14940, 2017865-2019-14942, 2017865-2019-14943, 2017865-2019-14944, 2017865-2019-14945. IT WAS REPORTED THAT THE PATIENT EXPIRED DURING AN IMPLANT PROCEDURE. THERE IS NO KNOWN ALLEGATION FROM A HEALTH PROFESSIONAL THAT SUGGESTS THE DEATH WAS RELATED TO THE DEVICE. IT WAS REPORTED THAT THE CAUSE OF DEATH WAS DUE TO RESPIRATORY ISSUES DURING THE PROCEDURE.

    Patient problems: Death

    MAUDE report 9177016 ↗
  • InjuryJun 4, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    RELATED MANUFACTURER REFERENCE NUMBER: 2938836-2019-04385. RELATED MANUFACTURER REFERENCE NUMBER: 2938836-2019-04386. IT WAS REPORTED THAT DURING AN INITIAL IMPLANT PROCEDURE, AFTER IMPLANTATION OF A RIGHT VENTRICULAR PACING LEAD, THE PHYSICIAN ATTEMPTED TO IMPLANT A QUADRIPOLAR PASSIVE LEAD IN THE LEFT VENTRICLE. PHYSICIAN COMPLAINED OF TIGHTNESS OF SUB-SELECTOR AROUND THE LV LEAD. THE LEAD COULDN¿T BE ADVANCED ALL THE WAY THROUGH THE SUB SELECTOR OF THE INNER CATHETER. ANOTHER INNER CATHETER WAS USED TO COMPLETE THE PROCEDURE. HOWEVER, THE DISTAL CORONARY SINUS (CS) WAS PERFORATED/ DISSECTED WHILE USING THE CPS UNIVERSAL DELIVERY SYSTEM OUTER AND INNER CATHETERS. THE PERFORATION RESULTED IN SIGNIFICANTLY LOWERED BLOOD PRESSURE, CARDIAC TAMPONADE, AND LOSS OF PULSE READINGS. THE PATIENT LOST A SIGNIFICANT AMOUNT OF BLOOD. THE PATIENT WAS ADMINISTERED CHEST COMPRESSIONS, RECEIVED MULTIPLE EXTERNAL DEFIBRILLATOR SHOCKS FOR SPONTANEOUS VF, AND WAS ADMINISTERED AN INTRATHORACIC NEEDLE AND CHEST TUBE TO EVACUATE BLOOD FROM THE PERICARDIUM. THE IMPLANT OF THE LEFT VENTRICULAR LEAD WAS DISCONTINUED AFTER THE TAMPONADE OCCURRED. UPON HEMODYNAMIC NORMALIZATION OF THE PATIENT, A RIGHT ATRIA

    Patient problems: Cardiac Perforation; Cardiac Tamponade; Vascular Dissection

    MAUDE report 8667955 ↗
  • InjuryMar 7, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DURING AN IMPLANT PROCEDURE, THERE WAS A DISSECTION NOTED ON THE CORONARY SINUS (CS) WITH ECHOCARDIOGRAM IMAGING DUE TO THE INTRODUCER TOOL. THERE WAS NO PERICARDIAL EFFUSION OR OTHER ADVERSE CONSEQUENCE AS A RESULT OF EVENT. THE PROCEDURE WAS POSTPONED TO A LATER DATE AND THE PATIENT WAS IN STABLE CONDITION AND WILL CONTINUE TO BE MONITORED.

    Patient problems: Vascular Dissection

    MAUDE report 8399780 ↗