Ampere™ Remote ControlRecalls & reported problems

St. Jude Medical, Inc. · Remote Control · Class 3

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05415067004901), received between Aug 30, 2018 and Mar 5, 2024 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Communication or Transmission Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Communication or Transmission Problem1
  • High impedance1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 5, 2024Communication or Transmission Problem; Communication or Transmission Problem

    DURING THE ATRIOVENTRICULAR NODAL REENTRANT TACHYCARDIA (AVNRT) PROCEDURE, COMMUNICATION ISSUES RESULTED IN A PROCEDURAL DELAY. THE AMPERE GENERATOR WAS UNABLE TO CONNECT TO THE ENSITE X. ATTEMPTED MULTIPLE REBOOTS, RESEATED CONNECTIONS, REBOOTED THE ENTIRE SYSTEM, AND STARTED A NEW STUDY WITH NO RESOLVE. REPLACED THE AMPERE FIBER, THE AMPERE, THE AMPLIFIER, AND IT WAS STILL NOT POSSIBLE TO CONNECT THE AMPERE TO THE ENSITE X. THE AMPERE REMOTE WAS REMOVED, AND THE AMPERE CONNECTED. THE PROCEDURE WAS COMPLETED WITHOUT ADVERSE CONSEQUENCES FOR THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18841865 ↗
  • InjuryAug 30, 2018High impedance; High impedance

    RELATED MANUFACTURING REFERENCE: 2182269-2018-00107, 9680001-2018-00127, 2184149-2018-00145. DURING AN ATRIAL FIBRILLATION ABLATION PROCEDURE A CANCELLATION OCCURRED. WHILE ABLATING IN THE LEFT ATRIUM, APPROXIMATELY 1 HOUR INTO THE PROCEDURE, AN IMPEDANCE OF 999 WAS DISPLAYED ON THE REMOTE CONTROL AND THE GENERATOR. ADDITIONALLY A "POP" WAS FELT BY THE USER FROM EITHER THE CATHETER OR THE SHEATH WITHIN THE PATIENT. THE SHEATH AND THE CATHETER WERE EXAMINED FOR ABNORMALITIES. THE SHEATH WAS NOTED TO BE DIFFICULT TO MANEUVER BUT NO VISIBLE ABNORMALITIES WERE OBSERVED. THE IMPEDANCE FROM THE GENERATOR HOWEVER DID NOT IMPROVE AND THE CASE WAS CANCELLED. THE PATIENT WAS STABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7833355 ↗