ViewMate™ H701336Recalls & reported problems

St. Jude Medical, Inc. · Ultrasound Console · Class 2

Adverse event reports
6
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 05415067013361), received between Nov 6, 2019 and Dec 3, 2025 — classified as 5 injuries, 1 malfunction.

The most frequently reported specific device problem is “Power Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Power Problem2
  • Communication or Transmission Problem1
  • Output Problem1
  • Smoking1
  • Sparking1
  • Tactile Prompts/Feedback1

Patient problems

  • No Consequences Or Impact To Patient4
  • No Clinical Signs, Symptoms or Conditions3
  • No Patient Involvement1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 3, 2025Sparking; Power Problem; Sparking; Power Problem

    DURING DEVICE PREPARATION WHEN POWERING UP THE VIEWMATE ULTRASOUND DEVICE, A SPARKING SOUND WAS HEARD AND SPARKING WAS VISIBLE. THE DEVICE WOULD NOT POWER ON. NO PATIENT WAS INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23709117 ↗
  • InjuryAug 26, 2020Smoking; Smoking

    WHEN THE SYSTEM WAS PLUGGED IN PRIOR TO A PROCEDURE, VISIBLE SMOKE WAS NOTED FROM THE SYSTEM. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Patient Involvement

    MAUDE report 10455364 ↗
  • InjuryAug 21, 2020Communication or Transmission Problem; Communication or Transmission Problem

    DURING THE PROCEDURE, THE TRANSDUCER COULD NOT BE SELECTED, A BLACK SCREEN WAS NOTED AND THE CASE WAS CANCELLED. THERE WERE NO CONSEQUENCES TO THE PATIENT AND THE CASE WAS RESCHEDULED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10437235 ↗
  • InjuryApr 15, 2020Power Problem; Power Problem

    IT WAS CONFIRMED THE FREEZE ISSUE OCCURRED PRIOR TO THE PATIENT ENTERING THE ROOM AND NO PATIENT WAS INVOLVED. THEREFORE THERE WAS NO CLINICALLY SIGNIFICANT DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 9963293 ↗
  • InjuryJan 10, 2020Tactile Prompts/Feedback; Tactile Prompts/Feedback

    DURING A PROCEDURE, WHEN THE VIEWMATE SYSTEM WAS STARTED, THE CONSOLE KEYS WERE NOT WORKING AS THE PATIENT INFORMATION COULD NOT BE ENTERED. WHEN THE CATHETER WAS CONNECTED, IT WOULD NOT REGISTER. THE SYSTEM WAS POWER CYCLED, THE PROBES WERE REPLACED, AND THE BATTERY WAS RE-INSERTED, BUT THE ISSUE REMAINED. THE PROCEDURE WAS ULTIMATELY CONTINUED BY SWITCHING TO TEE IMAGING, BUT THIS TRANSITION REQUIRED INCREASED SEDATION TIME FOR THE PATIENT. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9577319 ↗
  • InjuryNov 6, 2019Output Problem; Output Problem

    RELATED MANUFACTURING REF: 3008452825-2019-00561, 2184149-2019-00209. DURING A LEFT ATRIAL APPENDAGE CLOSURE PROCEDURE, THE SYSTEM DISPLAYED A "OVERHEATING ERROR" MESSAGE AND THE PROCEDURE WAS CANCELED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT DUE TO THE CANCELLATION.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9284089 ↗

Other System, imaging, pulsed echo, ultrasonic devices

All System, imaging, pulsed echo, ultrasonic devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.