AMPLATZER™ 9-ASD-012Recalls & reported problems
Aga Medical Corporation · Septal Occluder · Class 3 · Implant
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 05415067019202), received between Jun 20, 2019 and Dec 2, 2022 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Difficult to Fold, Unfold or Collapse” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Difficult to Fold, Unfold or Collapse2
- Migration1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 2, 2022Difficult to Fold, Unfold or Collapse; Difficult to Fold, Unfold or Collapse
IT WAS REPORTED THAT ON (B)(6) 2022, A 12MM AMPLATZER SEPTAL OCCLUDER WAS SELECTED FOR IMPLANTATION USING AN UNKNOWN ABBOTT DELIVERY SYSTEM. DURING THE PROCEDURE, DURING DEPLOYMENT, THE DEVICE DEFORMED INTO A COBRA SHAPE. THE DEVICE WAS REMOVED FROM PATIENT ANATOMY AND REPLACED WITH A ANOTHER 12MM AMPLATZER DUCT OCCLUDER. THE DEVICE WAS NOT RELEASED INSIDE PATIENT ANATOMY, THERE WAS NO ANGULATION OR KINK NOTED IN THE DELIVERY SYSTEM, AND THERE WAS NO INTERACTION WITH CARDIAC STRUCTURES DURING DEPLOYMENT. THE PATIENT IS REPORTED TO BE STABLE. THE PATIENT REMAINED HEMODYNAMICALLY STABLE THROUGHOUT THE PROCEDURE. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN PROCEDURE AND NO ADVERSE PATIENT EFFECTS. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15911127 ↗MalfunctionSep 20, 2022Difficult to Fold, Unfold or Collapse; Difficult to Fold, Unfold or Collapse
IT WAS REPORTED THAT ON (B)(6) 2022, A 12 MM AMPLATZER SEPTAL OCCLUDER WAS CHOSEN FOR IMPLANT. WHEN THE DEVICE WAS DEPLOYED, IT DISPLAYED A COBRA HEAD APPEARANCE. THE DOCTOR THEN RECAPTURED THE DEVICE AND REMOVED IT FROM THE BODY. THE DEVICE WAS NOT RELEASED INTO THE BODY. NO ADVERSE EVENTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15456741 ↗InjuryJun 20, 2019Migration; Migration
ON (B)(6) 2019, A 12MM AMPLATZER SEPTAL OCCLUDER WAS IMPLANTED USING ALL ON-LABEL SIZING EFFORTS INCLUDING A SIZING BALLOON UNDER TEE GUIDANCE. A TTE PERFORMED THE FOLLOWING DAY REVEALED THE DEVICE HAD EMBOLIZED. THE PATIENT WAS BROUGHT TO THE CATH LAB AND THE DEVICE WAS REMOVED FROM THE DESCENDING AORTA SUCCESSFULLY. THE USER BELIEVED THE DEVICE EMBOLIZED SECONDARY TO PATIENT ANATOMY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8717539 ↗
Other Transcatheter septal occluder devices
- Amplatzer™ Septal OccluderAga Medical Corporation · Transcatheter septal occluder294 reports
- Amplatzer™ Septal OccluderAga Medical Corporation · Transcatheter septal occluder1 death213 reports
- Amplatzer™ Talisman™ 9-PFO-3025St. Jude Medical, Inc. · Transcatheter septal occluder1 death193 reports
- GORE CARDIOFORM Septal Occluder GSX0030AW. L. Gore & Associates, Inc. · Transcatheter Septal Occluder1 recall143 reports
- Amplatzer™ 9-ASD-MF-025St. Jude Medical, Inc. · Transcatheter septal occluder1 death183 reports
- Amplatzer™ PFO OccluderSt. Jude Medical, Inc. · Transcatheter septal occluder1 death153 reports
- Amplatzer™ 9-ASD-MF-030St. Jude Medical, Inc. · Transcatheter septal occluder149 reports
- Amplatzer™ Talisman™ 9-PFO-2518St. Jude Medical, Inc. · Transcatheter septal occluder139 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page