Amplatzer™ 9-ASD-017Recalls & reported problems
Aga Medical Corporation · Septal Occluder · Class 3 · Implant
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 05415067019257), received between Mar 24, 2020 and Apr 18, 2023 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Difficult to Fold, Unfold or Collapse” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Difficult to Fold, Unfold or Collapse4
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 18, 2023Difficult to Fold, Unfold or Collapse; Difficult to Fold, Unfold or Collapse
IT WAS REPORTED THAT ON (B)(6) 2023, A 17MM AMPLATZER SEPTAL OCCLUDER WAS CHOSEN FOR IMPLANT USING A 9F AMPLATZER TREVISIO INTRAVASCULAR DELIVERY SYSTEM. DURING DEPLOYMENT IN THE LEFT ATRIUM, THE DEVICE HAD A COBRA DEFORMATION AND THE HEALTH CARE PROFESSIONAL TRIED SEVERAL ATTEMPTS TO CORRECT THE DEVICE, BUT WAS UNSUCCESSFUL. THE DEVICE WAS NOT RELEASED FROM THE DELIVERY CABLE. THE TRANSESOPHAGEAL ECHOCARDIOGRAM (TEE) AND ANGIOGRAPHY IMAGES WERE CHECKED BUT IT WAS SEEN THAT THE DEVICE CONTINUED TO NOT DEPLOY NORMALLY. THERE WAS NO REPORT OF ANY INTERACTION WITH LUMEN TISSUE DURING DEPLOYMENT AND NO REPORT OF ANY ANGULATION/KINK NOTICED WITH THE DELIVERY SYSTEM. WHEN THE DEVICE WAS REMOVED FROM THE PATIENT, IT ALSO DID NOT DEPLOY NORMALLY. A DECISION WAS MADE TO REPLACE THE DEVICE WITH A SECOND 17MM AMPLATZER SEPTAL OCCLUDER WITH SAME DELIVERY SYSTEM AND COMPLETED THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16763786 ↗MalfunctionMar 31, 2023Difficult to Fold, Unfold or Collapse; Difficult to Fold, Unfold or Collapse
IT WAS REPORTED THAT ON (B)(6) 2023, A 17MM AMPLATZER SEPTAL OCCLUDER WAS CHOSEN FOR IMPLANT. DURING DEPLOYMENT, THE DEVICE HAD A COBRA DEFORMATION. THE DEFORMED DEVICE WAS NEVER RELEASED FROM THE DELIVERY CABLE WHILE INSIDE THE PATIENT. A DECISION WAS MADE TO REPLACE THE DEVICE WITH A SECOND 17MM AMPLATZER SEPTAL OCCLUDER. THE REPLACEMENT DEVICE WAS IMPLANTED IN THE PATIENT WITH NO ADVERSE CONSEQUENCES. THERE WAS NO REPORT OF ANY INTERACTION WITH CARDIAC STRUCTURES DURING DEPLOYMENT AND THERE WAS NO ANGULATION/KINK NOTICED IN THE DELIVERY SYSTEMS. IT WAS ALSO REPORTED THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE DUE TO THE EVENT AND THE PATIENT REMAINED HEMODYNAMICALLY STABLE THROUGHOUT THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16653331 ↗MalfunctionJun 3, 2022Difficult to Fold, Unfold or Collapse; Difficult to Fold, Unfold or Collapse
IT WAS REPORTED A 17MM AMPLATZER SEPTAL OCCLUDER HAD BEEN CHOSEN FOR PROCEDURE ON (B)(6) 2022 USING AN 8 F DELIVERY SYSTEM. DURING THE PROCEDURE, THE DEVICE PRESENTED WITH A DEFORMATION. IT WAS NOTED THAT THE DEVICE WAS RE-SHEATHED FIVE TIMES, THE DEVICE WAS RETRIEVED. THE DEVICE ATTEMPTED AGAIN BUT THE DEFORMATION PERSISTED. IT WAS NOTED THAT THERE WAS A DELAY IN PROCEDURE BUT THAT THE PATIENT REMAINED STABLE THROUGHOUT THE PROCEDURE. THE DEVICE WAS REMOVED AND EXCHANGED WITH A NEW 17MM AMPLATZER SEPTAL OCCLUDER, THAT WAS SUCCESSFULLY IMPLANTED. THE PATIENT WAS NOTED HAVE BEEN DISCHARGED AT THE TIME OF REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14598670 ↗MalfunctionApr 14, 2022Difficult to Fold, Unfold or Collapse; Difficult to Fold, Unfold or Collapse
IT WAS REPORTED ON (B)(6) 2022, A 17MM AMPLATZER SEPTAL OCCLUDER WAS SELECTED FOR IMPLANT. DURING DEPLOYMENT, THE LEFT DISC APPEARED OVAL AND WAS PROTRUDING INTO THE LEFT ATRIUM. IT WAS NOTED THAT TWO ATTEMPTS WERE MADE TO DEPLOY THE OCCLUDER BEFORE THE DEVICE WAS REMOVED AND REPLACED WITH A NEW 17MM AMPLATZER SEPTAL OCCLUDER. THE NEW DEVICE WAS SUCCESSFULLY IMPLANTED WITHOUT ANY COMPLICATIONS. THE PATIENT WAS REPORTED TO BE IN STABLE CONDITION. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN PROCEDURE AND NO ADVERSE PATIENT EFFECTS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14111083 ↗MalfunctionMar 24, 2020Material Deformation; Material Deformation
ON (B)(6) 2020 DECIDED TO ATTEMPT A DEVICE IMPLANT OF 9-ASD-017. THE PHYSICIAN UNSHEATHED THE LEFT DISC DURING THE PROCEDURE AND THE DISC APPEARED COBRA SHAPED ON THE IMAGING MODALITIES. THE PHYSICIAN THEN DECIDED TO REMOVE THE DEVICE FROM THE PATIENTS BODY AND DID NOT WANT TO PROCEED WITH THE PROCEDURE USING THE MISSHAPED DEVICE. HE RE-SHEATHED THE LEFT DISC AND REMOVED THE DEVICE FROM THE PATIENTS BODY. THE PHYSICIAN THEN DEPLOYED A 9-ASD-018 SAFELY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9874367 ↗
Other Transcatheter septal occluder devices
- Amplatzer™ Septal OccluderAga Medical Corporation · Transcatheter septal occluder294 reports
- Amplatzer™ Septal OccluderAga Medical Corporation · Transcatheter septal occluder1 death213 reports
- Amplatzer™ Talisman™ 9-PFO-3025St. Jude Medical, Inc. · Transcatheter septal occluder1 death193 reports
- GORE CARDIOFORM Septal Occluder GSX0030AW. L. Gore & Associates, Inc. · Transcatheter Septal Occluder1 recall143 reports
- Amplatzer™ 9-ASD-MF-025St. Jude Medical, Inc. · Transcatheter septal occluder1 death183 reports
- Amplatzer™ PFO OccluderSt. Jude Medical, Inc. · Transcatheter septal occluder1 death153 reports
- Amplatzer™ 9-ASD-MF-030St. Jude Medical, Inc. · Transcatheter septal occluder149 reports
- Amplatzer™ Talisman™ 9-PFO-2518St. Jude Medical, Inc. · Transcatheter septal occluder139 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page