NeuroTherm™ DACUK-NTRecalls & reported problems
Neurotherm, Inc. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 05415067021335), received between Jun 15, 2018 and Apr 3, 2019 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Device Displays Incorrect Message” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device Displays Incorrect Message2
- Low impedance1
Patient problems
- No Consequences Or Impact To Patient3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryApr 3, 2019Low impedance; Low impedance
DURING A BIPOLAR PALISADE ABLATION PROCEDURE, THE IMPEDANCE WAS LOW AND THE GENERATOR EMITTED A BURNING SMELL SO THE PROCEDURE WAS CANCELLED. FOLLOWING THE CANCELLATION, FURTHER TROUBLESHOOTING DETERMINED THE ISSUE WAS DUE TO THE ADAPTER CABLE. A NEW CABLE WAS USED IN ALL PORTS AND THE IMPEDANCE DISPLAYED AS INTENDED AND NO BURNING SMELL WAS NOTED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT DUE TO THE CANCELLATION AND THE PROCEDURE WAS RESCHEDULED FOR A LATER DATE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8477692 ↗MalfunctionJun 15, 2018Device Displays Incorrect Message; Device Displays Incorrect Message
RELATED MANUFACTURER REFERENCE NUMBER: (B)(4). DURING A PROCEDURE PORT 1 ON THE GENERATOR DISPLAYED AN UNSAFE PROBE ERROR MESSAGE WHEN GOING INTO LESION MODE. THE PROCEDURE WAS CANCELLED AND THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. FURTHER TROUBLESHOOTING DETERMINED THE ISSUE WAS DUE TO TWO GENERATOR ADAPTER CABLES.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7604827 ↗MalfunctionJun 15, 2018Device Displays Incorrect Message; Device Displays Incorrect Message
RELATED MANUFACTURER REFERENCE NUMBER: 3002953813-2018-00036. DURING A PROCEDURE PORT 1 ON THE GENERATOR DISPLAYED AN UNSAFE PROBE ERROR MESSAGE WHEN GOING INTO LESION MODE. THE PROCEDURE WAS CANCELLED AND THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. FURTHER TROUBLESHOOTING DETERMINED THE ISSUE WAS DUE TO TWO GENERATOR ADAPTER CABLES.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7604823 ↗
Other CABLE, ELECTRODE devices
- N/A ECG Leadwire set, 5-lead, grouped, grabber, AHA, mix 74 cm/ 29 in, 130 cm/ 51 inGE Healthcare Finland Oy · CABLE, ELECTRODE1 recall1 report
- N/A ECG TRUNK CABLE, 6-LEAD, AHA, 1.2 M/4 FTGE Healthcare Finland Oy · CABLE, ELECTRODE1 recall0 reports
- N/A ECG Leadwire set, 5-lead V2-6, grabber, AHA, 74 cm/ 29 inGE Healthcare Finland Oy · CABLE, ELECTRODE1 recall0 reports
- N/A ECG TRUNK CABLE, 3-LD W/ INTEGRATED GRABBER LEAD WIRE, IEC, 3.6 M/12 FTGE Healthcare Finland Oy · CABLE, ELECTRODE1 recall0 reports
- N/A ECG Leadwire set, 3-lead, snap, AHA, 74 cm/ 29 inGE Healthcare Finland Oy · CABLE, ELECTRODE1 recall0 reports
- N/A ECG TRUNK CABLE, 3/5-LEAD, AHA, 6 M/20 FTGE Healthcare Finland Oy · CABLE, ELECTRODE1 recall0 reports
- N/A ECG Leadwire set, 5-lead, grabber, AHA, mix 74 cm/ 29 in, 130 cm/ 51 inGE Healthcare Finland Oy · CABLE, ELECTRODE1 recall0 reports
- N/A ECG Leadwire set, 5-lead, grouped, snap, AHA, 74 cm/ 29 inGE Healthcare Finland Oy · CABLE, ELECTRODE1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page