Advisor™ D-AVSE-D10-F20Recalls & reported problems

St. Jude Medical, Inc. · FL, Sensor Enabled Circular Mapping Catheter · Class 2

All models of Advisor (4)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
1
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 05415067024817), received between Jul 3, 2018 and Jun 13, 2025 — classified as 1 death, 5 injuries, 4 malfunctions.

The most frequently reported specific device problem is “Communication or Transmission Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem6
  • Communication or Transmission Problem2
  • Expiration Date Error1
  • Fracture1
  • Use of Device Problem1

Patient problems

  • Cardiac Perforation5
  • No Clinical Signs, Symptoms or Conditions4
  • Death1
  • Heart Block1
  • Stroke/CVA1

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 13, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THIS REPORT INCLUDES AN UPDATED EVENT DESCRIPTION AND UPDATED HEALTH CLINICAL CODE AND HEALTH IMPACT CODE. DURING AN ATRIAL FIBRILLATION PROCEDURE, WHILE MAPPING, AV BLOCK OCCURRED WHICH WAS TREATED WITH STIMULATION TO STABILIZE THE PATIENT. THE PATIENT WAS SEDATED WITH DEXDOR AND MAPPING WAS PERFORMED USING BOTH THE TACTICATH SE CATHETER AND THE ADVISOR FL SE CATHETER. TOWARDS THE END OF THE MAPPING THE PATIENT WENT INTO COMPLETE AV BLOCK WITH ST SEGMENT ELEVATION IN THE ABSENCE OF HEMODYNAMIC LESIONS ON CORONARY IMAGING. STIMULATION WAS ADMINISTERED IN THE CORONARY SINUS WITH THE CATHETER AND AFTER A FEW SECONDS SINUS RHYTHM WAS RESTORED. A CARDIOGRAM WAS DONE WHICH WAS NEGATIVE. THE PATIENT WAS UNABLE TO BE AWAKENED AND DEVELOPED A COMA, SO CT AND MRI EXAMS WERE DONE WHICH WERE NEGATIVE. THE PATIENT WAS ADMITTED TO ICU AND GRADUALLY RECOVERED CONSCIOUSNESS THE NEXT DAY. CT SCANS WERE REPEATED, AND AN ISCHEMIC ENCEPHALIC LESION APPEARED WITH CONCURRENT MOTOR DEFICIT OF THE UPPER LIMB. THE PHYSICIANS DO NOT THINK THAT ANY ABBOTT PRODUCTS ARE INVOLVED IN THE REPORTED EVENT. NO ABLATIONS HAD BEEN PERFORMED PRIOR TO THE REPORTED AV BLOCK. THERE WERE NO PERFORMANCE ISSUES WITH ANY ABB

    Patient problems: Heart Block; Stroke/CVA

    MAUDE report 22202557 ↗
  • MalfunctionApr 18, 2023Communication or Transmission Problem; Communication or Transmission Problem

    RELATED MANUFACTURING REF: 3008452825-2023-00167. DURING AN ATRIAL TACHYCARDIA PROCEDURE, A COMMUNICATION ISSUE OCCURRED WITH THE ADVISOR HD GRID CATHETER RESULTING IN A DELAY TO THE PROCEDURE. THE ADVISOR HD GRID CATHETER WAS INSERTED INTO THE RIGHT ATRIUM AND MAPPING WAS PERFORMED. THEN THE ADVISOR FL CATHETER WAS INSERTED INTO THE LEFT ATRIUM AND THE MAPPING CATHETER OF THE ENSITE X SETTING WAS SET TO THE ADVISOR FL. THE REFERENCE (CS) COULD NOT BE COLLECTED, THE MAPPING COULD NOT BE PERFORMED WITH THE ADVISOR FL CATHETER WHILE THE ELECTRIC POTENTIALS WAS NOT DISPLAYED. THE ELECTRIC POTENTIALS OF THE ADVISOR FL CATHETER WERE DISPLAYED ON THE RECORDING SYSTEM. THE CATHETER AND THE CABLE WERE RE-CONNECTED, AND RE-SET TO THE CATHETER SET UP SCREEN OF THE ENSITE X, THE DWS WAS RE-STARTED, AND THE ADVISOR FL CATHETER WAS REPLACED WITH A SECOND ONE, BUT WITH NO RESOLUTION. THE LEFT ATRIUM WAS MAPPED WITH THE ADVISOR HD GRID CATHETER. THE PROCEDURE WAS COMPLETED AND THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. THE HOSPITAL DISCARDED ONLY SECOND CATHETER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16766377 ↗
  • MalfunctionApr 18, 2023Communication or Transmission Problem; Communication or Transmission Problem

    RELATED MANUFACTURING REF: 3008452825-2023-00168. DURING A ATRIAL TACHYCARDIA PROCEDURE, A COMMUNICATION ISSUE OCCURRED WITH THE ADVISOR HD GRID CATHETER RESULTING IN A DELAY TO THE PROCEDURE. THE ADVISOR HD GRID CATHETER WAS INSERTED INTO THE RIGHT ATRIUM AND MAPPING WAS PERFORMED. THEN THE ADVISOR FL CATHETER WAS INSERTED INTO THE LEFT ATRIUM AND THE MAPPING CATHETER OF THE ENSITE X SETTING WAS SET TO THE ADVISOR FL. THE REFERENCE (CS) COULD NOT BE COLLECTED, THE MAPPING COULD NOT BE PERFORMED WITH THE ADVISOR FL CATHETER WHILE THE ELECTRIC POTENTIALS WAS NOT DISPLAYED. THE ELECTRIC POTENTIALS OF THE ADVISOR FL CATHETER WERE DISPLAYED ON THE RECORDING SYSTEM. THE CATHETER AND THE CABLE WERE RE-CONNECTED, AND RE-SET TO THE CATHETER SET UP SCREEN OF THE ENSITE X, THE DWS WAS RE-STARTED, AND THE ADVISOR FL CATHETER WAS REPLACED WITH A SECOND ONE, BUT WITH NO RESOLUTION. THE LEFT ATRIUM WAS MAPPED WITH THE ADVISOR HD GRID CATHETER. THE PROCEDURE WAS COMPLETED AND THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. THE HOSPITAL DISCARDED ONLY SECOND CATHETER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16766374 ↗
  • MalfunctionApr 7, 2022Fracture; Fracture

    THIS REPORT IS TO ADVISE OF AN EVENT NOTED DURING ANALYSIS WHICH REVEALED A TEAR IN THE CATHETER SHAFT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14046118 ↗
  • MalfunctionJan 12, 2022Use of Device Problem; Expiration Date Error; Use of Device Problem; Expiration Date Error

    THIS EVENT IS BEING REPORTED AS THE DEVICE USED DURING THE PROCEDURE HAD EXPIRED. THERE WERE NO CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13236897 ↗
  • DeathMar 29, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DURING A PULMONARY VEIN ISOLATION ABLATION PROCEDURE.

    Patient problems: Cardiac Perforation; Death

    MAUDE report 8465820 ↗
  • InjuryFeb 15, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    RELATED MANUFACTURER REFERENCE NUMBER: 3005334138-2019-00078, 3008452825-2019-00067, 3005334138-2019-00079, 3005334138-2019-00080. DURING A PULMONARY VEIN ISOLATION ABLATION PROCEDURE A PERICARDIAL EFFUSION OCCURRED. JUST AFTER TRANSSEPTAL ACCESS WAS OBTAINED AND DURING LEFT ATRIAL GEOMETRY RECONSTRUCTION, THE PATIENT BECAME HYPOTENSIVE. AN ECHOCARDIOGRAM REVEALED A PERICARDIAL EFFUSION FOR WHICH A PERICARDIOCENTESIS WAS PERFORMED TO STABILIZE THE PATIENT. THERE WERE NO PERFORMANCE ISSUES WITH ANY ABBOTT DEVICE.

    Patient problems: Cardiac Perforation

    MAUDE report 8345542 ↗
  • InjuryJan 30, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    RELATED MANUFACTURING REF: 3008452825-2019-00030. DURING AN ATRIAL FIBRILLATION PROCEDURE, A PERICARDIAL EFFUSION OCCURRED. WHILE ABLATING IN THE LEFT ATRIUM, THE PATIENT BECAME HYPOTENSIVE. AN ECHOCARDIOGRAM REVEALED A PERICARDIAL EFFUSION. A PERICARDIOCENTESIS WAS PERFORMED TO STABILIZE THE PATIENT. THERE WERE NO PERFORMANCE ISSUES WITH ANY ABBOTT DEVICE.

    Patient problems: Cardiac Perforation

    MAUDE report 8288432 ↗
  • InjuryNov 15, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    RELATED MANUFACTURER REFERENCE NUMBER: 2030404-2018-00080, 3005334138-2018-00432, 3008452825-2018-00390. DURING A PULMONARY VEIN ISOLATION ABLATION PROCEDURE A PERICARDIAL EFFUSION OCCURRED. AT THE END OF THE PROCEDURE, AFTER REMOVING ALL DEVICES, THE PATIENT BECAME HYPOTENSIVE. A TRANSTHORACIC ECHOCARDIOGRAM WAS PERFORMED, REVEALING A PERICARDIAL EFFUSION MEASURING 2 CENTIMETERS. A PERICARDIOCENTESIS WAS PERFORMED TO STABILIZE THE PATIENT. THERE WERE NO PERFORMANCE ISSUES WITH ANY ABBOTT DEVICE.

    Patient problems: Cardiac Perforation

    MAUDE report 8074884 ↗
  • InjuryJul 3, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    RELATED MANUFACTURER REFERENCE NUMBER: 9680001-2018-00100, 3008452825-2018-00220, 3005334138-2018-00187, 3005334138-2018-00188, 3008452825-2018-00221. FOLLOWING A PULMONARY VEIN ISOLATION ABLATION PROCEDURE, A PERICARDIAL EFFUSION OCCURRED. AS THE DEVICES WERE REMOVED THE PATIENT BECAME HYPOTENSIVE. PHARMACOLOGICAL INTERVENTION WAS GIVEN BUT THE PATIENT¿S BLOOD PRESSURE REMAINED LOW. AN ECHOCARDIOGRAM WAS PERFORMED, REVEALING A GLOBAL EFFUSION CAUSING CARDIAC TAMPONADE. A DRAIN WAS INSERTED HOWEVER THE PERICARDIUM KEPT REFILLING SO OPEN-HEART SURGERY WAS PERFORMED. A PERFORATION WAS LOCATED AT THE BASE OF THE LEFT ATRIAL APPENDAGE WHICH REQUIRED A SUTURE TO CLOSE. THE PATIENT STABILIZED AND IS RECOVERING FROM THE SURGERY. THERE WERE NO PERFORMANCE ISSUES WITH ANY ABBOTT DEVICES.

    Patient problems: Cardiac Perforation

    MAUDE report 7658106 ↗

Other CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING devices

All CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING devices →