Advisor™ VL, Sensor Enabled™ D-AVSE-DF20-V1525Recalls & reported problems
St. Jude Medical, Inc. · Circular Mapping Catheter · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 9
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 05415067027283), received between Jul 9, 2020 and Jun 24, 2025 — classified as 9 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Entrapment of Device” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Entrapment of Device7
- Adverse Event Without Identified Device or Use Problem3
- Difficult to Remove2
- Activation, Positioning or Separation Problem1
- Fracture1
- Patient Device Interaction Problem1
Patient problems
- Cardiac Perforation3
- No Clinical Signs, Symptoms or Conditions3
- Intraoperative Cardiac Valve Injury2
- Injury1
- Low Oxygen Saturation1
- No Consequences Or Impact To Patient1
- Regurgitation, Valvular1
- Respiratory Distress1
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 24, 2025Entrapment of Device; Fracture; Activation, Positioning or Separation Problem; Entrapment of Device; Fracture; Activation, Positioning or Separation Problem
DURING AN ATRIAL FIBRILLATION PROCEDURE, THE LOOP OF THE CATHETER WAS DAMAGED AND GOT CAUGHT ON HEART TISSUE. WHEN THE CATHETER WAS INSERTED INTO THE LEFT ATRIUM, IT WAS FOUND THAT THE CATHETER MANEUVERING AND COLLECTION OF GEOMETRY WERE DIFFICULT. THE CATHETER WAS REMOVED WITH NO INTERVENTION NEEDED, AND IT WAS NOTED TO BE DAMAGED. THE CATHETER WAS REPLACED, AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22323573 ↗InjuryFeb 21, 2025Entrapment of Device; Entrapment of Device
DURING AN ATRIAL FIBRILLATION PROCEDURE, THE ADVISOR VL CATHETER WAS ADVANCED INTO THE CORDAGE OF THE MITRAL VALVE. AFTER MANIPULATION ATTEMPTS, IT WAS NOT POSSIBLE TO REMOVE THE CATHETER. THE PATIENT WAS TRANSFERRED TO ANOTHER HOSPITAL FOR SURGICAL REMOVAL OF THE DEVICE BY CARDIAC MICROSURGERY WITHOUT ANY DAMAGE AND THE PATIENT IS STABLE. THERE WERE NO ALLEGED PERFORMANCE ISSUES WITH ANY ABBOTT DEVICES.
Patient problems: Intraoperative Cardiac Valve Injury
MAUDE report 21437925 ↗InjuryFeb 6, 2025Entrapment of Device; Entrapment of Device
DURING A PROCEDURE, THE CATHETER BECAME ENTANGLED IN THE MITRAL VALVE AND COULD NOT BE FREED. THE CATHETER HAD TO BE SURGICALLY REMOVED, AND THE MITRAL VALVE HAD TO BE REPLACED. THE PATIENT IS STABLE.
Patient problems: Intraoperative Cardiac Valve Injury
MAUDE report 21327534 ↗InjuryAug 5, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
DURING AN ATRIAL FIBRILLATION PROCEDURE, A CARDIAC PERFORATION OCCURRED. MAPPING WAS PERFORMED USING THE ENSITE. DURING MAPPING, THE SYSTOLIC BLOOD PRESSURE TEMPORARILY DROPPED INTO THE 40S. AN INTRACARDIAC ECHO CONFIRMED A SMALL AMOUNT OF PERICARDIAL FLUID, BUT DRAINAGE WAS NOT REQUIRED. SEDATION WAS RELEASED, AND A FOLLOW-UP OBSERVATION WAS PERFORMED. AFTER THAT, THE SYSTOLIC BLOOD PRESSURE INCREASED TO THE NORMAL RANGE, BUT THE PROCEDURE WAS DISCONTINUED, AND FOLLOW-UP OBSERVATION WAS PERFORMED. THE CATHETER WAS INSERTED DEEP INTO THE LEFT ATRIAL APPENDAGE DURING MAPPING, AND THERE WAS A POSSIBILITY THAT THE CARDIAC PERFORATION HAPPENED AT THAT TIME. THE PHYSICIAN ALLEGES THAT THE MAPPING CATHETER CAUSED THE CARDIAC PERFORATION. THERE HAVE BEEN NO SIGNIFICANT CHANGES IN THE PATIENT'S CONDITION SINCE THE DAY OF THE EVENT AND THE PATIENT IS STILL HOSPITALIZED. THERE WERE NO PERFORMANCE ISSUES WITH ANY ABBOTT DEVICES IN THIS EVENT.
Patient problems: Cardiac Perforation
MAUDE report 19911377 ↗MalfunctionJan 9, 2023Entrapment of Device; Difficult to Remove; Entrapment of Device; Difficult to Remove
DURING THE PAROXYSMAL ATRIAL FIBRILLATION PROCEDURE USING THE ADVISOR¿ VL, SENSOR ENABLED¿ CIRCULAR MAPPING CATHETER, IT WAS NOTED THAT IT BECAME ENTANGLED WITH THE CHORDAE TENDINEAE. THE CATHETER WAS INSERTED INTO THE LEFT ATRIUM, AND THE CATHETER INSERTED INTO THE LEFT VENTRICLE. WHEN THE GEOMETRY OF THE LEFT ATRIUM WAS BEING CREATED, IT WAS NOTED THE CATHETER WAS ENTANGLED WITH THE CHORDAE TENDINEAE, AND IT COULD NOT COME OUT. THE CATHETER WAS ROTATED, BUT IT WAS STILL HARD TO REMOVE, SO THE SHEATH WAS REMOVED. NO DAMAGE WAS SEEN ON THE CATHETER, BUT IT WAS REPLACED AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16122891 ↗InjuryOct 8, 2021Entrapment of Device; Entrapment of Device
NO ADDITIONAL TREATMENT FOR MYOCARDIAL DAMAGE WAS PERFORMED. THE PATIENT WAS DISCHARGED FROM THE HOSPITAL IN STABLE CONDITION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12599752 ↗InjuryDec 28, 2020Entrapment of Device; Entrapment of Device
DURING AN ATRIAL FIBRILLATION PROCEDURE, WHEN MAPPING OF THE LEFT ATRIUM WAS BEING PERFORMED ON THE ENSITE, THE CATHETER ENTERED THE LEFT VENTRICLE AND BECAME ENTANGLED IN THE CHORDAE TENDINEAE AND THE PAPILLARY MUSCLES, AND THE CATHETER COULD NOT BE REMOVED FROM THERE. TORQUE WAS APPLIED TO THE CATHETER BY VIEWING THROUGH THE ECHOCARDIOGRAM, THE CATHETER WAS PULLED INTO THE SHEATH, AND PACING WAS APPLIED WITH THE CATHETER BEING MOVED, BUT THE CATHETER COULD NOT BE REMOVED. THE PATIENT WAS MOVED TO AN OPERATING ROOM IN THE CARDIOVASCULAR SURGERY UNIT AND THE CATHETER WAS REMOVED BY THORACOSCOPY AND A VALVULOPLASTY WAS PERFORMED FOR TORN PAPILLARY MUSCLES. IT WAS FOUND THAT THE CHORDAE TENDINEAE AND THE PAPILLARY MUSCLES WERE TORN OFF AND ENTANGLED. THE CATHETER WAS DISPOSED OF AT THE HOSPITAL.
Patient problems: Injury
MAUDE report 11078388 ↗InjuryNov 18, 2020Patient Device Interaction Problem; Patient Device Interaction Problem
RELATED MANUFACTURER REPORT NUMBER: 3008452825-2020-00604. AFTER A PAROXYSMAL ATRIAL FIBRILLATION ABLATION PROCEDURE, A BLOOD CLOT WAS NOTED TO BE ATTACHED TO THE ELECTRODE 7. AN ECHOCARDIOGRAM WAS PERFORMED AFTER THE PROCEDURE, AND NO FLOATING SUBSTANCE/MATERIAL LIKE A BLOOD CLOT WAS FOUND IN THE LEFT ATRIUM AND THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: Thrombus
MAUDE report 10859757 ↗InjurySep 14, 2020Difficult to Remove; Difficult to Remove
DURING RIGHT ATRIAL GEOMETRY, THE CATHETER BECAME LODGED IN THE TRICUSPID VALVE APPARATUS (CHORDAE TENDINEAE). THE CATHETER BECAME GOT CAUGHT IN THE CHORDAE TENDINEAE AND WHEN PULLING THE CATHETER BACK INTO THE SHEATH TO FREE THE CATHETER, THE PAPILLARY MUSCLE WAS RUPTURED. THE RUPTURE CAUSED A PROLAPSE OF THE TRICUSPID VALVE (TV) AND REGURGITATION. THE PATIENT WAS STABLE UNDER GENERAL ANESTHESIA, SO THE PROCEDURE CONTINUED. PULMONARY VEIN ISOLATION AND A CTI LINE WAS COMPLETED USING THE SAME ADVISOR VL AS THE DIAGNOSTIC CATHETER. THE PHYSICIAN BELIEVES THE TRICUSPID VALVE REGURGITATION WAS A RESULT OF THE CATHETER GETTING STUCK IN THE CHORDAE AND RUPTURING THE PAPILLARY MUSCLE, CAUSING PROLAPSE OF THE TV. COMBINED WITH THE PREVIOUSLY KNOWN PFO, THIS LED TO PATIENT OXYGEN DESATURATION. AFTER THE ABLATION PROCEDURE WAS COMPLETED, A PFO OCCLUDER WAS INSERTED INTO THE PFO TO PREVENT FURTHER DESATURATION. THE PATIENT MAY NEED VALVE REPAIR SURGERY IN THE FUTURE. THERE WERE NO PERFORMANCE ISSUES WITH THE ABBOTT DEVICE.
Patient problems: Low Oxygen Saturation; Regurgitation, Valvular; Respiratory Distress; Tricuspid Regurgitation
MAUDE report 10532520 ↗InjuryAug 27, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
RELATED MANUFACTURER REFERENCE NUMBER: 3008452825-2020-00444. AT THE END OF A PULMONARY VEIN ISOLATION ABLATION PROCEDURE, A PERICARDIAL EFFUSION OCCURRED. THE ABLATION PROCEDURE WAS COMPLETED WITH NO REPORTED ISSUES, BUT AT THE END OF THE PROCEDURE, AFTER ALL THE CATHETERS WERE REMOVED, THE PATIENT BECAME HYPOTENSIVE. AN ECHOCARDIOGRAM REVEALED A PERICARDIAL EFFUSION, FOR WHICH A PERICARDIOCENTESIS WAS PERFORMED TO STABILIZE THE PATIENT. THERE WERE NO PERFORMANCE ISSUES WITH ANY ABBOTT DEVICE.
Patient problems: Cardiac Perforation
MAUDE report 10462605 ↗InjuryJul 16, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
DURING AN ATRIAL FIBRILLATION AND ATRIAL FLUTTER ABLATION PROCEDURE, A PERICARDIAL EFFUSION OCCURRED. TRANSSEPTAL ACCESS WAS GAINED TO THE LEFT ATRIUM AND THE ADVISOR VL WAS INSERTED. THERE WAS DIFFICULTY NAVIGATING AND MAPPING THE LEFT ATRIAL APPENDAGE. AFTER THE MAP WAS COMPLETED, A SECOND TRANSSEPTAL PUNCTURE WAS PERFORMED AND THE ABLATION CATHETER WAS INSERTED INTO THE LEFT ATRIUM. NOISE WAS OBSERVED ON THE DISTAL TIP OF THE ABLATION CATHETER AND IT WAS NOT LOCATING PROPERLY ON ENSITE. NO ABLATION WAS CONDUCTED. SHORTLY AFTER, THE PATIENT BECAME HYPOTENSIVE. A 2 CM PERICARDIAL EFFUSION WAS OBSERVED ON ICE. THE MAPPING CATHETER AND ABLATION CATHETERS WERE MOVED OUT OF THE LEFT ATRIUM. A PERICARDIAL TAP WAS PERFORMED WHICH STABILIZED THE PATIENT. THE PATIENT'S HEMOGLOBIN AND HEMATOCRIT COUNT WAS DECREASING DESPITE DRAINING THE EFFUSION. A HEMATOMA WAS IDENTIFIED AT THE RIGHT GROIN SITE. THE HEMATOMA WAS CAUSED WHEN ACHIEVING VENOUS ACCESS AT THE BEGINNING OF THE PROCEDURE AND WAS NOT BELIEVED TO BE DUE TO AN ABBOTT DEVICE. A CT SCAN WAS ALSO PERFORMED AND RULED OUT RETROPERITONEAL BLEED. THE PATIENT WAS INTUBATED AND LEFT IN STABLE CONDITION. THE PHYSICIAN BELIEVES THE PERICARDIA
Patient problems: Cardiac Perforation
MAUDE report 10283104 ↗MalfunctionJul 9, 2020Entrapment of Device; Entrapment of Device
DURING THE PROCEDURE, THE DISTAL ELECTRODE WAS NOTED TO BE STUCK IN THE CHORDA TENDINEA AROUND THE TRICUSPID VALVE. THE CATHETER COULD NO LONGER BE OPERATED. THE CATHETER WAS REMOVED FROM THE PATIENT BY APPLYING TORQUE TO THE CATHETER SHAFT SEVERAL TIMES BUT THE LOOP WAS FOUND TO BE ELONGATED VERTICALLY. NO FURTHER INTERVENTION WAS NEEDED TO REMOVE THE DEVICE. THE CATHETER WAS REPLACED AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10254492 ↗
FDA clearance and classification
Cleared through 510(k) K202066 on Nov 25, 2020 (decision: Substantially Equivalent). Submitted by Abbott Medical as “EnSite X EP System, Advisor VL Circular Mapping Catheter, Sensor Enabled, Advisor FL Circular Mapping Catheter, Sensor Enabled, Advisor HD Grid High Density Mapping Catheter, Senor Enabled”. The FDA review took 121 days. FDA record ↗
Cleared through 510(k) K192037 on Sep 17, 2019 (decision: Substantially Equivalent). Submitted by Abbott Medical as “Advisor VL Circular Mapping Catheter, Sensor Enabled”. The FDA review took 49 days. FDA record ↗
The FDA regulates product code DRF (catheter, electrode recording, or probe, electrode recording) under 21 CFR 870.1220 as a Class II device. 21 CFR 870.1220 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Advisor™ VL, Sensor Enabled™ D-AVSE-DF20-V1525
Has the Advisor™ VL, Sensor Enabled™ D-AVSE-DF20-V1525 been recalled?
No FDA recall record in our data names this exact device identifier (DI 05415067027283), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Advisor™ VL, Sensor Enabled™ D-AVSE-DF20-V1525?
The FDA's MAUDE database lists 12 adverse event reports for this device, received between Jul 9, 2020 and Jun 24, 2025. In those reports, the most common specific device problems are “Entrapment of Device” (7), “Difficult to Remove” (2) and “Activation, Positioning or Separation Problem” (1); the most common patient problems are “Cardiac Perforation” (3), “Intraoperative Cardiac Valve Injury” (2) and “Injury” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Advisor™ VL, Sensor Enabled™ D-AVSE-DF20-V1525?
None of the reports are classified as deaths, 9 as injuries and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Advisor™ VL, Sensor Enabled™ D-AVSE-DF20-V1525 FDA approved?
It was cleared through 510(k) K192037 on Sep 17, 2019; its record lists 2 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the Advisor™ VL, Sensor Enabled™ D-AVSE-DF20-V1525, and is it still sold?
The FDA's device database lists St. Jude Medical, Inc. as the company responsible for it (brand Advisor™ VL, Sensor Enabled™, model D-AVSE-DF20-V1525). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING devices
- IntellaMap Orion™ M004RC64S0Boston Scientific Corporation · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING6 deaths356 reports
- Advisor™ HD Grid, Sensor Enabled™ D-AVHD-DF16St. Jude Medical, Inc. · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING6 deaths356 reports
- ACHIEVE™ 990063-020Medtronic, Inc. · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING7 deaths129 reports
- Inquiry™ IBI-81104Irvine Biomedical, Inc. · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING1 recall2 deaths64 reports
- DECANAV R7F282CTBiosense Webster Inc · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING4 deaths108 reports
- LASSO NAV D134301Biosense Webster Inc · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING2 deaths82 reports
- DECANAV R7D282CTBiosense Webster Inc · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING1 death76 reports
- ACHIEVE ADVANCE™ 2ACH20Medtronic, Inc. · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING3 deaths69 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record