Amplatzer™ Post Infarct Muscular VSD Occluder (REF 9-VSDMPIHDE-016)Recalls & reported problems

Aga Medical Corporation · Post Infarct Muscular VSD Occluder · Class 3 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05415067028105), received between Mar 31, 2022 and Jan 27, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Difficult to Fold, Unfold or Collapse” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Fold, Unfold or Collapse2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 27, 2025Difficult to Fold, Unfold or Collapse; Difficult to Fold, Unfold or Collapse

    IT WAS REPORTED THAT ON (B)(6) 2025, A 16MM AMPLATZER POST-INFARCT MUSCULAR VSD OCCLUDER WAS CHOSEN FOR IMPLANT USING AN UNKNOWN DELIVERY SYSTEM. DURING PROCEDURE, IT WAS NOTED THAT THE DEVICE HAD A COBRA HEAD ON THE RIGHT DISC AFTER DEPLOYMENT. THE DEVICE WAS RECAPTURED AND TAKEN OUT OF THE BODY. THE DEFORMED DEVICE WAS NEVER RELEASED FROM THE DELIVERY CABLE WHILE INSIDE THE PATIENT. THERE WAS NO REPORT OF ANY INTERACTION WITH CARDIAC STRUCTURES DURING DEPLOYMENT AND NO REPORT OF ANY ANGULATION/KINK NOTICED WITH THE DELIVERY SYSTEM. A NEW UNKNOWN DEVICE WAS SUCCESSFULLY IMPLANTED. THE PATIENT REMAINED HEMODYNAMICALLY STABLE THROUGHOUT THE PROCEDURE. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN PROCEDURE AND NO ADVERSE PATIENT EFFECTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21242577 ↗
  • MalfunctionMar 31, 2022Difficult to Fold, Unfold or Collapse; Difficult to Fold, Unfold or Collapse

    THIS EVENT IS BEING REPORTED FOR MALFUNCTION DUE TO THE DEFORMATION ON THE 16MM AMPLATZER POST-INFARCT MUSCULAR VSD OCCLUDER. THE PERICARDIAL EFFUSION/CARDIAC TAMPONADE IS RELATED TO THE NON-ABBOTT 0.035 GLIDE CATHETER. IT WAS REPORTED ON (B)(6) 2022, A 16MM AMPLATZER POST-INFARCT MUSCULAR VSD OCCLUDER WAS SELECTED FOR IMPLANT FOR POST INFARCT VENTRICULAR SEPTAL DEFECT (PI VSD). IT WAS NOTED THAT APPROXIMATELY 6 WEEKS PRIOR TO IMPLANT PROCEDURE THE PATIENT EXPERIENCED A SIGNIFICANT LEFT VENTRICULAR (LV) MYOCARDIAL INFARCTION RESULTING IN A FORMATION OF PI VSD NEAR THE APEX OF BOTH THE LEFT AND RIGHT VENTRICLES. THE PATIENT ALSO SUFFERED A NUMBER OF PERICARDIAL EFFUSIONS THAT WERE NOTED IMMEDIATELY FOLLOWING THE INFARCTION. DURING DEPLOYMENT A COBRA DEFORMATION WAS NOTED ON THE INITIAL ATTEMPT. THE DEVICE WAS RECAPTURED AND A SECOND DEPLOYMENT WAS ATTEMPTED HOWEVER A DEFORMATION WAS NOTED AGAIN. DURING THIS TIME, UNDER TRANSESOPHAGEAL ECHOCARDIOGRAM (TEE), AN EFFUSION WAS NOTED AND WAS GETTING LARGER. IT WAS DECIDED TO RECAPTURE THE DEVICE AGAIN AND DEPLOY FOR A THIRD TIME HOWEVER, THE COBRA SHAPE APPEARED AGAIN AND THE PROXIMAL DISK WAS STILL CAPTURED IN THE SHEATH. THE PATIENT BLO

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13967308 ↗

Other Transcatheter Septal Occluder devices

All Transcatheter Septal Occluder devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page