Skip to content

Amplatzer™ Post Infarct Muscular VSD Occluder (REF 9-VSDMPIHDE-018)Recalls & reported problems

Aga Medical Corporation · Post Infarct Muscular VSD Occluder · Class 3 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
1
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 05415067028112), received between Nov 18, 2020 and Sep 29, 2025 — classified as 1 death, 2 malfunctions.

The most frequently reported specific device problem is “Activation Failure” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Activation Failure1
  • Difficult to Fold, Unfold or Collapse1

Patient problems

  • Bradycardia1
  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1
  • Shock1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 29, 2025Difficult to Fold, Unfold or Collapse; Difficult to Fold, Unfold or Collapse

    IT WAS REPORTED THAT ON (B)(6) 2025, A 18MM AMPLATZER PI MUSC VSD WAS CHOSEN FOR IMPLANT USING A 10F AMPLATZER TREVISIO DELIVERY SYSTEM. DURING THE PROCEDURE, THE DEVICE FORMED A COBRA SHAPE ON THE RIGHT VENTRICULAR(RV) DISC. FOLLOWING THAT, PHYSICIAN CHOSE A 22MM AMPLATZER VASCULAR PLUG II , BUT THE DEVICE WAS PULLED THROUGH THE DEFECT AND WAS REMOVED BEFORE RELEASE FROM CABLE AS THE PHYSICIAN FELT IT WAS SMALL FOR THE DEFECT. THE PROCEDURE WAS COMPLETED USING A NEW 24MM AMPLATZER PI MUSC VSD WITH SATISFACTORY CLOSURE. IT IS REPORTED THAT THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE AND THE PATIENT REMAINED HEMODYNAMICALLY STABLE THROUGHOUT THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23174976 ↗
  • DeathFeb 15, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT ON (B)(6) 2023, A 18MM AMPLATZER PI MUSC VSD FOR EMERGENT IMPLANT IN A PATIENT WHO HAD CARDIOGENIC SHOCK WITH VENTRICULAR SEPTAL DEFECT POST MYOCARDIAL INFARCTION ON (B)(6) 2023. THE IMPLANT PROCEDURE WAS SUCCESSFUL AND DEVICE WAS IMPLANTED IN STABLE POSITION. PATIENT WAS DISCHARGED TO A SKILLED NURSING FACILITY ON (B)(6) 2024. ON (B)(6) 2023, THE PATIENT WAS RE-ADMITTED IN CARDIOGENIC SHOCK AND BRADYCARDIA. THE PATIENT EXPIRED ON THE SAME DAY. THE DEATH IS NOT THOUGHT TO BE DEVICE RELATED.

    Patient problems: Bradycardia; Shock

    MAUDE report 18711658 ↗
  • MalfunctionNov 18, 2020Adverse Event Without Identified Device or Use Problem; Activation Failure; Adverse Event Without Identified Device or Use Problem; Activation Failure

    ON (B)(6) 2020 THE PHYSICIAN ATTEMPTED TO PLACE A 9-VSDMPIHDE-018. HOWEVER, THE PLACEMENT WAS UNSUCCESSFUL SO THE DEVICE WAS REMOVED AND CASE ABORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10862796 ↗

Other Transcatheter Septal Occluder devices

All Transcatheter Septal Occluder devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page