Skip to content

Tendyne™ TENDV-LP-33SRecalls & reported problems

St. Jude Medical, Inc. · Mitral Valve · Class 3 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 05415067032591), received between Oct 3, 2025 and Oct 3, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Off-Label Use” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Off-Label Use1
  • Perivalvular Leak1

Patient problems

  • Hemolysis1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 3, 2025Perivalvular Leak; Off-Label Use; Perivalvular Leak; Off-Label Use

    IT WAS REPORTED THAT ON (B)(6) 2025, A 33S TENDYNE MITRAL VALVE LP WAS IMPLANTED WITH A SMALL TENDYNE APICAL PAD. ON (B)(6) 2025, A HEMOLYSIS HAD OCCURRED DUE TO PERIVALVULAR LEAK (PVL). ON (B)(6) 2025, A RE-TENSIONING WAS DONE SUCCESSFULLY. AND NO PVL LEFT.

    Patient problems: Hemolysis

    MAUDE report 23209220 ↗

Other Prosthesis, mitral valve, percutaneously delivered devices

All Prosthesis, mitral valve, percutaneously delivered devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page