Skip to content

HeartMate Touch™ HMT1100-RRecalls & reported problems

Thoratec Corporation · Communication System · Class 3

All models of HeartMate Touch (3)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 05415067033758), received between Sep 12, 2022 and Feb 13, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Communication or Transmission Problem” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

31 other recalls cover devices cleared under the same FDA submission (P060040, P160054); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jan 3, 2024Class IOpen, Classified

    Due to software and controller systems that results in unexpected pump start or pump stop.

    Z-0842-2024 · Root cause: Software design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Communication or Transmission Problem2
  • Display or Visual Feedback Problem2
  • Unintended Power Up1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 13, 2024Display or Visual Feedback Problem; Communication or Transmission Problem; Display or Visual Feedback Problem; Communication or Transmission Problem

    IT WAS REPORTED THAT THE HEARTMATE TOUCH KIT LOST CONNECTION TO THE SYSTEM CONTROLLER DURING THE PRIMING OF THE PUMP, PRIOR TO IMPLANT. THE SESSION WAS ENDED AND RESTARTED. THE PRIMING CONTINUED AT 3000 RPM, BUT THE PRIMING WINDOW WAS NOT VIEWABLE. THE PUMP AUTOMATICALLY STOPPED POST COMPLETION OF THE ORIGINAL TIMER DESPITE THIS DATA NOT BEING VISIBLE. THERE WAS ALSO NO DISPLAY THAT THE PUMP STOP WAS COMPLETE AND THERE WAS NO RED LINE TO SHOW THIS EITHER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18697320 ↗
  • MalfunctionDec 12, 2023Unintended Power Up; Unintended Power Up

    IT WAS REPORTED THAT DURING THE PRIMING OF THE PUMP, AND WHILE THE PUMP WAS CONNECTED TO THE HEARTMATE TOUCH MONITOR, THE VENTRICULAR ASSIST DEVICE (VAD) COORDINATOR ASSUMED THE PRIMING PROCESS WAS DONE AND DISCONNECTED THE SYSTEM CONTROLLER POWER CABLES WHILE THE RED PUMP STOP LINE WAS FILLING. LATER, THE POWER CABLES WERE RECONNECTED WITH THE POWER MODULE AND AS SOON AS THE CONTROLLER WAS RECOGNIZED BY THE HEARTMATE TOUCH SYSTEM THE PUMP AUTO-STARTED AT 3000 RPM. THERE WAS NO IMPACT TO THE PATIENT AT THIS TIME, AS THE OUTFLOW GRAFT HAD ALREADY BEEN ATTACHED AND THE SURGEON HAD REQUESTED FOR THE PUMP TO BE STARTED WHEN THE RECONNECTION AND UNEXPECTED AUTOSTART HAPPENED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18314144 ↗
  • MalfunctionSep 12, 2022Display or Visual Feedback Problem; Communication or Transmission Problem; Display or Visual Feedback Problem; Communication or Transmission Problem

    ADDITIONAL INFORMATION PROVIDED STATED THAT THE HEARTMATE TOUCH APP COULD NOT BE FORCE-CLOSED DUE TO IT BEING FROZEN. IT WAS REPORTED THAT THE ACCOUNT SWITCHED TO THE SYSTEM MONITOR AFTER THE ISSUE WAS OBSERVED. NO FURTHER ISSUES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15404957 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P060040, P160054. One submission often covers many models, so check the product and lot details in each record.

  • Z-2519-2026 · May 13, 2026 · Open, Classified
    Brand Name: HeartMate Product Name: HeartMate 3" System Controller Model/Catalog Number: 106531US Software Version: N/A Product Description: The HeartMate 3" System Controller acts
  • Z-2520-2026 · May 13, 2026 · Open, Classified
    Brand Name: HeartMate Product Name: HM 11V Li-ion Backup Battery, ABT Model/Catalog Number: 106128 Software Version: N/A Product Description: The HeartMate 11 Volt Lithium Ion Stan
  • Z-2518-2026 · May 13, 2026 · Open, Classified
    Brand Name: HeartMate Product Name: HeartMate 3 Left Ventricular System (LVAS) Kits Model/Catalog Number: 106524US Software Version: N/A Product Description: The HeartMate 3" Left
  • Z-0507-2026 · Oct 9, 2025 · Open, Classified
    Brand Name: HeartMate Product Name: HeartMate II LVAS Implant Kit Model/Catalog Number: 106015, HeartMate II LVAS Kit, with Pocket Controller, Sealed, US; 106016, HeartMate II LV
  • Z-0506-2026 · Oct 9, 2025 · Open, Classified
    Brand Name: HeartMate Product Name: HeartMate 3 Controller (Standalone) Model/Catalog Number: 106531US, HeartMate 3 Controller, US; 106531INT, HeartMate 3 Controller, OUS; 106531
  • Z-0508-2026 · Oct 9, 2025 · Open, Classified
    Brand Name: HeartMate Product Name: HeartMate II Controller (standalone) Model/Catalog Number: 106017, HeartMate II Pocket Controller, US; 106762, HeartMate II Pocket Controller,
  • Z-0505-2026 · Oct 9, 2025 · Open, Classified
    Brand Name: HeartMate Product Name: HeartMate 3 Left Ventricular System (LVAS) Kits Model/Catalog Number: 106524US, HeartMate 3 LVAS KIT, US; 106524INT, HeartMate 3 LVAD KIT, OUS
  • Z-2290-2025 · Jun 23, 2025 · Open, Classified
    Brand Name: HeartMate Product Name: Mobile Power Unit, AC Power Cord Model/Catalog Number: 107760 Software Version: N/A Product Description: The HeartMate 3" Left Ventricular Assis
  • Z-2291-2025 · Jun 23, 2025 · Open, Classified
    Brand Name: HeartMate Product Name: Mobile Power Unit Model/Catalog Number: 107754 Software Version: N/A Product Description: The HeartMate 3" Left Ventricular Assist System (LVAS)
  • Z-1570-2025 · Mar 13, 2025 · Open, Classified
    HeartMate Mobile Power Unit: REF: 107754, NA; 107758, EU; 107758AU, AU; 107758UK, UK; L107758, EU Rental. Used with HeartMate 3 and II Left Ventricular Assist System.
  • Z-2583-2024 · Jun 25, 2024 · Open, Classified
    HeartMate 3 System Controllers provided within the following HeartMate 3 Left Ventricular Assist System (LVAS) Implant Kits or distributed separately: HeartMate 3 LVAS Implant Ki
  • Z-1921-2024 · May 8, 2024 · Open, Classified
    Thoratec HeartMate System Monitor, REF: 1286, L1286, 1286A, L1286A, 1286A-US, L1286A-US, 1286C, L1286C, L1386, 1286A-CAN, 1286INT and L1286INT. Left ventricular assist device
  • Z-1692-2024 · Mar 1, 2024 · Open, Classified
    Thoratec HeartMate 3, LVAS Implant Kit, REF:106524US, Sterile EO, RxOnly
  • Z-1693-2024 · Mar 1, 2024 · Open, Classified
    Thoratec HeartMate 3, LVAS Implant Kit, REF:106524INT, Sterile EO, RxOnly
  • Z-1300-2024 · Feb 19, 2024 · Open, Classified
    HeartMate II Left Ventricular Assist System (LVAS): 1. HeartMate II Sealed Outflow Graft with Bend Relief REF 103393 2. HeartMate II LVAS Implant Kit, US REF 106015 3. HeartMate

…and 16 more.

Other Ventricular (assist) bypass devices

All Ventricular (assist) bypass devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page