HeartMate Touch™ HMT1100-RRecalls & reported problems
Thoratec Corporation · Communication System · Class 3
All models of HeartMate Touch (3)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 05415067033758), received between Sep 12, 2022 and Feb 13, 2024 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Communication or Transmission Problem” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
31 other recalls cover devices cleared under the same FDA submission (P060040, P160054); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Jan 3, 2024Class IOpen, Classified
Due to software and controller systems that results in unexpected pump start or pump stop.
Z-0842-2024 · Root cause: Software design · Recall details · FDA record ↗
Most reported problems
Device problems
- Communication or Transmission Problem2
- Display or Visual Feedback Problem2
- Unintended Power Up1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 13, 2024Display or Visual Feedback Problem; Communication or Transmission Problem; Display or Visual Feedback Problem; Communication or Transmission Problem
IT WAS REPORTED THAT THE HEARTMATE TOUCH KIT LOST CONNECTION TO THE SYSTEM CONTROLLER DURING THE PRIMING OF THE PUMP, PRIOR TO IMPLANT. THE SESSION WAS ENDED AND RESTARTED. THE PRIMING CONTINUED AT 3000 RPM, BUT THE PRIMING WINDOW WAS NOT VIEWABLE. THE PUMP AUTOMATICALLY STOPPED POST COMPLETION OF THE ORIGINAL TIMER DESPITE THIS DATA NOT BEING VISIBLE. THERE WAS ALSO NO DISPLAY THAT THE PUMP STOP WAS COMPLETE AND THERE WAS NO RED LINE TO SHOW THIS EITHER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18697320 ↗MalfunctionDec 12, 2023Unintended Power Up; Unintended Power Up
IT WAS REPORTED THAT DURING THE PRIMING OF THE PUMP, AND WHILE THE PUMP WAS CONNECTED TO THE HEARTMATE TOUCH MONITOR, THE VENTRICULAR ASSIST DEVICE (VAD) COORDINATOR ASSUMED THE PRIMING PROCESS WAS DONE AND DISCONNECTED THE SYSTEM CONTROLLER POWER CABLES WHILE THE RED PUMP STOP LINE WAS FILLING. LATER, THE POWER CABLES WERE RECONNECTED WITH THE POWER MODULE AND AS SOON AS THE CONTROLLER WAS RECOGNIZED BY THE HEARTMATE TOUCH SYSTEM THE PUMP AUTO-STARTED AT 3000 RPM. THERE WAS NO IMPACT TO THE PATIENT AT THIS TIME, AS THE OUTFLOW GRAFT HAD ALREADY BEEN ATTACHED AND THE SURGEON HAD REQUESTED FOR THE PUMP TO BE STARTED WHEN THE RECONNECTION AND UNEXPECTED AUTOSTART HAPPENED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18314144 ↗MalfunctionSep 12, 2022Display or Visual Feedback Problem; Communication or Transmission Problem; Display or Visual Feedback Problem; Communication or Transmission Problem
ADDITIONAL INFORMATION PROVIDED STATED THAT THE HEARTMATE TOUCH APP COULD NOT BE FORCE-CLOSED DUE TO IT BEING FROZEN. IT WAS REPORTED THAT THE ACCOUNT SWITCHED TO THE SYSTEM MONITOR AFTER THE ISSUE WAS OBSERVED. NO FURTHER ISSUES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15404957 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P060040, P160054. One submission often covers many models, so check the product and lot details in each record.
- Z-2519-2026 · May 13, 2026 · Open, ClassifiedBrand Name: HeartMate Product Name: HeartMate 3" System Controller Model/Catalog Number: 106531US Software Version: N/A Product Description: The HeartMate 3" System Controller acts
- Z-2520-2026 · May 13, 2026 · Open, ClassifiedBrand Name: HeartMate Product Name: HM 11V Li-ion Backup Battery, ABT Model/Catalog Number: 106128 Software Version: N/A Product Description: The HeartMate 11 Volt Lithium Ion Stan
- Z-2518-2026 · May 13, 2026 · Open, ClassifiedBrand Name: HeartMate Product Name: HeartMate 3 Left Ventricular System (LVAS) Kits Model/Catalog Number: 106524US Software Version: N/A Product Description: The HeartMate 3" Left
- Z-0507-2026 · Oct 9, 2025 · Open, ClassifiedBrand Name: HeartMate Product Name: HeartMate II LVAS Implant Kit Model/Catalog Number: 106015, HeartMate II LVAS Kit, with Pocket Controller, Sealed, US; 106016, HeartMate II LV
- Z-0506-2026 · Oct 9, 2025 · Open, ClassifiedBrand Name: HeartMate Product Name: HeartMate 3 Controller (Standalone) Model/Catalog Number: 106531US, HeartMate 3 Controller, US; 106531INT, HeartMate 3 Controller, OUS; 106531
- Z-0508-2026 · Oct 9, 2025 · Open, ClassifiedBrand Name: HeartMate Product Name: HeartMate II Controller (standalone) Model/Catalog Number: 106017, HeartMate II Pocket Controller, US; 106762, HeartMate II Pocket Controller,
- Z-0505-2026 · Oct 9, 2025 · Open, ClassifiedBrand Name: HeartMate Product Name: HeartMate 3 Left Ventricular System (LVAS) Kits Model/Catalog Number: 106524US, HeartMate 3 LVAS KIT, US; 106524INT, HeartMate 3 LVAD KIT, OUS
- Z-2290-2025 · Jun 23, 2025 · Open, ClassifiedBrand Name: HeartMate Product Name: Mobile Power Unit, AC Power Cord Model/Catalog Number: 107760 Software Version: N/A Product Description: The HeartMate 3" Left Ventricular Assis
- Z-2291-2025 · Jun 23, 2025 · Open, ClassifiedBrand Name: HeartMate Product Name: Mobile Power Unit Model/Catalog Number: 107754 Software Version: N/A Product Description: The HeartMate 3" Left Ventricular Assist System (LVAS)
- Z-1570-2025 · Mar 13, 2025 · Open, ClassifiedHeartMate Mobile Power Unit: REF: 107754, NA; 107758, EU; 107758AU, AU; 107758UK, UK; L107758, EU Rental. Used with HeartMate 3 and II Left Ventricular Assist System.
- Z-2583-2024 · Jun 25, 2024 · Open, ClassifiedHeartMate 3 System Controllers provided within the following HeartMate 3 Left Ventricular Assist System (LVAS) Implant Kits or distributed separately: HeartMate 3 LVAS Implant Ki
- Z-1921-2024 · May 8, 2024 · Open, ClassifiedThoratec HeartMate System Monitor, REF: 1286, L1286, 1286A, L1286A, 1286A-US, L1286A-US, 1286C, L1286C, L1386, 1286A-CAN, 1286INT and L1286INT. Left ventricular assist device
- Z-1692-2024 · Mar 1, 2024 · Open, ClassifiedThoratec HeartMate 3, LVAS Implant Kit, REF:106524US, Sterile EO, RxOnly
- Z-1693-2024 · Mar 1, 2024 · Open, ClassifiedThoratec HeartMate 3, LVAS Implant Kit, REF:106524INT, Sterile EO, RxOnly
- Z-1300-2024 · Feb 19, 2024 · Open, ClassifiedHeartMate II Left Ventricular Assist System (LVAS): 1. HeartMate II Sealed Outflow Graft with Bend Relief REF 103393 2. HeartMate II LVAS Implant Kit, US REF 106015 3. HeartMate
…and 16 more.
Other Ventricular (assist) bypass devices
- Thoratec® HeartMate 3™ 106524USThoratec Corporation · Ventricular (Assisst) Bypass7 recalls5,096 deaths16,478 reports
- THORATEC® HEARTMATE®, LVAS IMPLANT KIT (WITH SEALED GRAFTS) 106015Thoratec Corporation · Ventricular (Assisst) Bypass2 recalls2,141 deaths8,000 reports
- Thoratec® HeartMate 3™ 106531USThoratec Corporation · Ventricular (Assisst) Bypass1 recall11 deaths4,646 reports
- HVAD Pump Implant Kit 1103Heartware Inc. · Ventricular (Assist) Bypass6 recalls424 deaths2,952 reports
- HVAD® Pump Implant Kit (Enhanced) 1103Heartware Inc. · Ventricular (Assisst) Bypass4 recalls573 deaths2,807 reports
- HVAD® Pump Implant Kit 1103Heartware Inc. · Ventricular (Assisst) Bypass4 recalls666 deaths2,737 reports
- HVAD® Controller 1420Heartware Inc. · Ventricular (Assisst) Bypass5 recalls14 deaths2,161 reports
- HEARTMATE®, MOBILE POWER UNIT, NA 107754Thoratec Corporation · Ventricular (Assisst) Bypass1 death2,098 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page