Skip to content

Masters Series™ 23AJ-501Recalls & reported problems

St. Jude Medical, Inc. · Mechanical Heart Valve · Class 3 · Implant

All models of Masters Series (28)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05415067043030), received between Mar 18, 2025 and Jun 30, 2025 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Physical Resistance/Sticking” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Physical Resistance/Sticking2

Patient problems

  • Aortic Valve Stenosis1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 30, 2025Physical Resistance/Sticking; Physical Resistance/Sticking

    IT WAS REPORTED THAT ON (B)(6) 2025, A 23MM MASTERS VALVE WAS SUCCESSFULLY IMPLANTED. AFTER THE HEART RESUMED BEATING, AN ESOPHAGEAL ULTRASOUND WAS PERFORMED AND THE TRANSVALVULAR PRESSURE DIFFERENCE WAS TOO HIGH AND THE BLOOD FLOW VELOCITY WAS TOO FAST. THE MECHANICAL VALVE WAS REMOVED AND TESTED, REVEALING THAT THE VALVE LEAFLETS WERE HAVING DIFFICULTY OPENING AND CLOSING. A NEW 19MM REGENT VALVE WAS IMPLANTED TO RESOLVE THE EVENT. THE PHYSICIAN DOESN'T BELIEVE THE PATIENT OR USER EXPERIENCED ADVERSE HEALTH CONSEQUENCES DUE TO THE PERFORMANCE OF THE PRODUCT. THE PATIENT REMAINED HEMODYNAMICALLY STABLE THROUGHOUT THE PROCEDURE AND THERE WAS NO CLINICALLY SIGNIFICANT DELAY. THE PATIENT IS DOING WELL.

    Patient problems: Aortic Valve Stenosis

    MAUDE report 22359634 ↗
  • MalfunctionMar 18, 2025Physical Resistance/Sticking; Physical Resistance/Sticking

    IT WAS REPORTED THAT ON (B)(6) 2025, A 23MM MASTERS AORTIC MECHANICAL HEART VALVE WAS SELECTED FOR IMPLANT. DURING DEVICE PREPARATION, THE LEAFLETS WERE NOT MOVING NORMALLY WHEN TESTED WITH A VALVE TESTER. DURING PROCEDURE, THE VALVE LEAFLETS DID NOT "OPEN AND CLOSE FLEXIBLY" AND WAS REMOVED. UPON REMOVAL, THE LEAFLETS WERE AGAIN TESTED OUTSIDE OF THE PATIENT AND THE ISSUE PERSISTED. A NEW 23MM MASTERS AORTIC MECHANICAL HEART VALVE WAS SUCCESSFULLY IMPLANTED. THE PATIENT WAS REPORTED TO BE IN NORMAL CONDITION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21635165 ↗

Other HEART-VALVE, MECHANICAL devices