St. Jude Medical, Inc. · Aortic Valved Graft with Hemashield Graft technology Mechanical Valved Graft · Class 3 · Implant

All models of Masters Series (28)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 05415067044099), received between Mar 4, 2024 and Mar 18, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Contamination /Decontamination Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Contamination /Decontamination Problem2
  • Physical Resistance/Sticking2

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 18, 2025Physical Resistance/Sticking; Physical Resistance/Sticking

    IT WAS REPORTED THAT ON (B)(6) 2025, A 25MM MASTERS SERIES MECHANICAL HEART VALVE WAS SELECTED FOR IMPLANT. DURING DEVICE PREPARATION, THE LEAFLETS WERE MOVING NORMALLY. AFTER THE VALVE WAS IMPLANTED AND SUTURING WAS COMPLETED, THE HEART STARTED BEATING AGAIN. HOWEVER, THIS WAS "ACCOMPANIED BY THE PROBLEM OF BLOCKING THE HEART VALVE." THE VALVE WAS UNABLE TO BE ROTATED AFTER IMPLANTATION; THE LEAFLETS WERE NOT OPENING AND CLOSING. THE CAUSE OF THE INABILITY TO ROTATE WAS REPORTED AS "THE MECHANICAL VALVE IS STUCK." A VALVE MEASURING DEVICE WAS USED TO TEST THE LEAFLET FUNCTION. THE VALVE WAS EXPLANTED AND UPON EXPLANT, THE LEAFLET MOVEMENT WAS TESTED AGAIN WITH NORMAL MOBILITY. IT WAS EXCHANGED FOR A NEW 23MM MASTERS SERIES HEART COATED AORTIC VALVED GRAFT, WHICH WAS SUCCESSFULLY IMPLANTED. THE PATIENT WAS REPORTED TO BE IN IN GOOD CONDITION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21635063 ↗
  • MalfunctionDec 13, 2024Contamination /Decontamination Problem; Physical Resistance/Sticking; Contamination /Decontamination Problem; Physical Resistance/Sticking

    IT WAS REPORTED THAT ON (B)(6) 2024, A 25MM SJM MASTERS SERIES COATED AORTIC VALVED GRAFT WAS SELECTED FOR IMPLANT. DURING THE PROCEDURE, WHILE SUTURING THE VALVE, THE PHYSICIAN NOTICED STAINS ON THE AORTIC VALVE GRAFT TUBE. THE DEVICE WAS REMOVED AND REPLACED. IT WAS ALSO NOTED THAT THE LEAFLETS DIDN'T OPEN AND CLOSE SMOOTHLY. THERE WERE NO ADVERSE HEALTH CONSEQUENCES TO THE PATIENT. THE PATIENT REMAINED HEMODYNAMICALLY STABLE THROUGHOUT THE PROCEDURE AND THERE WAS NO CLINICALLY SIGNIFICANT DELAY. THE PATIENT HAS BEEN DISCHARGED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20929042 ↗
  • MalfunctionMar 4, 2024Contamination /Decontamination Problem; Contamination /Decontamination Problem

    IT WAS REPORTED THAT ON (B)(6)2024, A 25MM SJM MASTERS SERIES COATED AORTIC VALVED GRAFT WAS CHOSEN FOR A BENTALL PROCEDURE. DURING PROCEDURE, AFTER THE ANASTOMOSIS OF THE DEVICE, THE PHYSICIAN FOUND LARGE AMOUNT OF YELLOW DOTS INSIDE THE CONDUIT. THERE WAS NO ANOMALIES WERE FOUND DURING PREPARATION. THE PHYSICIAN DECIDED NOT TO EXPLANT THE DEVICE AND TRIED TO REMOVE THEM ONE BY ONE. WHEN THE BLOOD FLOW WAS RESTORED HE REPORTED A LOST OF IMPERMEABILITY OF THE CONDUIT, THE INITIAL LOSS OF BLOOD FROM THE TISSUE WAS RESOLVED AFTER SOME MINUTES. THE PATIENT REMAINED HEMODYNAMICALLY STABLE THROUGHOUT THE PROCEDURE. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN PROCEDURE AND NO ADVERSE PATIENT EFFECTS. THE PATIENT WAS REPORTED STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18828841 ↗