Skip to content

Masters Series™ 33MECJ-502Recalls & reported problems

St. Jude Medical, Inc. · Mechanical Heart Valve · Class 3 · Implant

All models of Masters Series (28)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 05415067045294), received between May 23, 2025 and Feb 5, 2026 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Incomplete Coaptation” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incomplete Coaptation2
  • Intravalvular regurgitation1
  • Patient Device Interaction Problem1
  • Physical Resistance/Sticking1

Patient problems

  • Heart Failure/Congestive Heart Failure2
  • Mitral Valve Insufficiency/ Regurgitation1
  • No Clinical Signs, Symptoms or Conditions1
  • Thrombosis/Thrombus1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 5, 2026Incomplete Coaptation; Patient Device Interaction Problem

    IT WAS REPORTED THAT ON (b)(6) 2026, A 33MM SJM MASTERS SERIES VALVE EXPANDED CUFF WAS IMPLANTED IN THE PATIENT. TWENTY FOUR HOURS LATER, PATIENT NEEDED MORE INOTROPES AND A TRANSESOPHAGEAL ECHOCARDIOGRAM (TEE) SHOWED A THROMBUS WAS BUILT ON THE VALVE. INTRAOPERATIVELY, THE PATIENT WAS HEMODYNAMICALLY STABLE AND DRY, WITH NO VISIBLE BLEEDING. HOWEVER, IN THE INTENSIVE CARE UNIT, DUE TO INCREASED CHEST TUBE DRAINAGE OUTPUT, MEDICATIONS WERE ADMINISTERED. THE PATIENT WAS REOPERATED THE NEXT DAY BECAUSE ONE LEAFLET WAS NOT MOVING. THE PATIENT WAS REPORTED RECOVERED.

    Patient problems: Heart Failure/Congestive Heart Failure; Thrombosis/Thrombus

    MAUDE report 24272476 ↗
  • InjuryJan 16, 2026Incomplete Coaptation; Intravalvular regurgitation

    IT WAS REPORTED THAT ON (B)(6) 2025, A 33MM SJM MASTERS SERIES VALVE EXPANDED CUFF WAS CHOSEN FOR A PROCEDURE. POST OPERATIVE TRANSESOPHAGEAL ECHOCARDIOGRAM (TOE) SHOWED REGURGITATION. THE ANTERIOR LEAFLET WAS NOT FUNCTIONING PROPERLY. THE PATIENT RETURNED TO THEATRE FOR UNPLANNED SURGERY TO REPLACE WITH A FUNCTIONING VALVE. THE PATIENT WAS IN LOW CARDIAC OUTPUT STATE OVERNIGHT ON DAY 0 POST OPERATIVELY, CONFOUNDED BY ASPIRATION EVENT. THE PATIENT SUFFERED A PERIOD OF LOW CARDIAC OUTPUT WITH RESULTANT MULTI ORGAN FAILURE. ON 18 JAN 2026, THE PATIENT WAS REPORTED PASSED AWAY FROM NEUROLOGICAL INJURY, MULTI ORGAN FAILURE, GASTROINTESTINAL ISCHEMIA/INFRACTION AND UPPER GASTROINTESTINAL (GI) BLEEDING.

    Patient problems: Heart Failure/Congestive Heart Failure; Mitral Valve Insufficiency/ Regurgitation

    MAUDE report 24093020 ↗
  • MalfunctionMay 23, 2025Physical Resistance/Sticking; Physical Resistance/Sticking

    SUBSEQUENT TO THE PREVIOUSLY FILED REPORT, ADDITIONAL INFORMATION WAS RECEIVED: THE VALVE FUNCTIONED NORMALLY WHEN TESTED DURING PREPARATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22078611 ↗

Other HEART-VALVE, MECHANICAL devices

All HEART-VALVE, MECHANICAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page