The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05415067046192), received between Oct 6, 2025 and Oct 6, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Retraction Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 2 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Retraction Problem2
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 6, 2025Retraction Problem; Material Deformation; Retraction Problem; Material Deformation

    IT WAS REPORTED THAT ON (B)(6) 2025, A 35M TENDYNE MITRAL VALVE LP WAS CHOSEN FOR A PROCEDURE. DURING PROCEDURE, AFTER THE DEPLOYMENT OF VALVE, THE PATIENT DEVELOPED A VARIABLE LEFT VENTRICULAR OUTFLOW TRACT (LVOT) GRADIENT (30-50MMMHG) THAT COULD NOT BE MITIGATED BY TENSION REDUCTION. THE PHYSICIAN DECIDED TO REMOVE THE VALVE. A TENDYNE GEN II RETRIEVAL SYSTEM HAS BEEN PREPARED BY INSTRUCTIONS FOR USE (IFU). A GOOD METAL TO METAL AND ALIGNMENT ACHIEVED BEFORE STARTING TO CAPTURE THE VALVE. WHEN ROTATING THE RETRIEVAL KNOB THE VALVE STARTED TO ENTER INTO THE SHEATH BUT THEN STOPPED DESPITE RETRIEVAL KNOB ROTATIONS. THE CATHETER BUCKLING HAS BEEN NOTICED AND NO PRESSURE APPLIED ON THE CATHETER BY THE PHYSICIAN DURING THE RETRIEVAL. THE RETRIEVAL HAS BEEN REMOVED AND A NEW TENDYNE GEN II RETRIEVAL SYSTEM HAS BEEN PREPARED FOLLOWING IFU. A GOOD METAL TO METAL AND ALIGNMENT ACHIEVED BUT, EXTREME RESISTANCE NOTICED DURING THE RETRIEVAL KNOB ROTATION AND THE KNOB WAS SPINNING BACK IF HAND REMOVED FROM THE WHEEL. THE CATHETER SHOWED SEVERE BOWING DURING VALVE RETRIEVAL. FINALLY THE VALVE HAS BEEN RETRIEVED. THE PATIENT HEMODYNAMIC WAS STABLE AND PATIENT AWAKE THE DAY AFTER. THE PATIENT WI

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23222177 ↗
  • MalfunctionOct 6, 2025Retraction Problem; Retraction Problem

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23222174 ↗

Other Prosthesis, mitral valve, percutaneously delivered devices