Liberta RC™ 62400Recalls & reported problems
Advanced Neuromodulation Systems, Inc. · Implantable Pulse Generator · Class 3 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Liberta RC deep brain stimulator is an implanted device used to treat essential tremor. Its 104 reports, 28 of them injuries, often describe the stimulator shutting down unexpectedly, and some describe infections at the implant site that required removal of the device. Reports describe 1 death, and 2 open recalls name this device: stimulation can switch off on a repeating schedule, and the device may lose communication with the clinician programmer.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 103
- Deaths reported
- 1
- Injuries reported
- 27
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 103 adverse event reports that list this exact device (DI 05415067047663), received between Jun 7, 2024 and Sep 15, 2026 — classified as 1 death, 27 injuries, 75 malfunctions.
The most frequently reported specific device problem is “Unintended Application Program Shut Down” (73 reports).
2 FDA recall records name this device's identifier, 2 still open.
4 other recalls cover devices cleared under the same FDA submission (P140009); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Oct 6, 2025Class IIOpen, Classified
There is a potential issue affecting Eterna Spinal Cord Stimulation (SCS) Implantable Pulse Generators (IPGs), Model 32400 and Liberta RC Deep Brain Stimulation (DBS) Implantable Pulse Generators (IPGs), Model 62400 . These devicess may experience a loss of communication with the Clinician Programmer (CP) and/or the Patient Controller (PC).
Z-0459-2026 · Root cause: Process control · Recall details · FDA record ↗
- May 16, 2024Class IIOpen, Classified
Deep brain stimulation system will first turn off after approximately 50 days after the system is activated and subsequently, this same action will recur approximately every 50 days thereafter. This unexpected lost of stimulation may result in loss of therapy.
Z-2265-2024 · Root cause: Software design · Recall details · FDA record ↗
Most reported problems
Device problems
- Unintended Application Program Shut Down73
- Adverse Event Without Identified Device or Use Problem14
- Unexpected Shutdown6
- Inappropriate/Inadequate Shock/Stimulation3
- Wireless Communication Problem3
- High impedance2
- Device Displays Incorrect Message1
- Impedance Problem1
Patient problems
- Movement Disorder77
- Failure of Implant72
- No Clinical Signs, Symptoms or Conditions10
- Post Operative Wound Infection6
- Undesired Nerve Stimulation3
- Arrhythmia2
- Inadequate Pain Relief2
- Burning Sensation1
Recent adverse event reports
The 25 most recent of 103 reports, in the reporter's own words as published by the FDA.
InjurySep 15, 2026Premature Discharge of Battery
It was reported the patient¿s IPG had an increased recharge burden. As a result, surgical intervention was undertaken wherein the IPG was replaced on (b)(6) 2026 to address the issue.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50772380 ↗InjuryMar 23, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT HAD CARDIAC COMPLICATIONS FOLLOWING THEIR IMPLANT PROCEDURE AND IT RESULTED IN TWO ELECTRICAL CARDIOVERSIONS. ALL SYMPTOMS HAVE NOW RESOLVED AND THERAPY WAS RESTORED.
Patient problems: Arrhythmia
MAUDE report 24665385 ↗InjuryMar 20, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT HAD CARDIAC COMPLICATIONS FOLLOWING THEIR IMPLANT PROCEDURE AND IT RESULTED IN TWO ELECTRICAL CARDIOVERSIONS. ALL SYMPTOMS HAVE NOW RESOLVED AND THERAPY WAS RESTORED.
Patient problems: Arrhythmia
MAUDE report 24650546 ↗InjuryFeb 27, 2026High impedance
IT WAS REPORTED PATIENT EXPERIENCED HIGH IMPEDANCES. SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE LEFT EXTENSION AND IPG WERE EXPLANTED AND REPLACED TO ADDRESS THE ISSUE. THERAPY WAS RESTORED POST-OP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24476698 ↗InjuryFeb 18, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED PATIENT'S SYSTEM WAS INFECTED AT THE BURR HOLE AND IPG POCKET SITES. SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE ENTIRE SYSTEM EXPLANTED. PATIENT WAS PRESCRIBED ANTIBIOTICS TO ADDRESS THE INFECTION.
Patient problems: Post Operative Wound Infection
MAUDE report 24381141 ↗InjuryFeb 17, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THE PATIENT EXPERIENCED AN INFECTION AT THE IPG SITE. IN TURN, SURGICAL INTERVENTION TOOK PLACE WHEREIN THE IPG WAS EXPLANTED TO ADDRESS THE ISSUE.
Patient problems: Post Operative Wound Infection
MAUDE report 24363570 ↗InjuryJan 27, 2026Inappropriate/Inadequate Shock/Stimulation
IT WAS REPORTED THAT FOLLOWING SEVERAL FALLS, THE PATIENT EXPERIENCED UNCOMFORTABLE STIMULATION AT THE IPG SITE. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN THE IPG WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE.
Patient problems: Undesired Nerve Stimulation
MAUDE report 24195724 ↗DeathJan 15, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT PASSED AWAY WITHIN 30 DAYS OF AN ABBOTT PROCEDURE. THE CAUSE OF DEATH IS UNKNOWN.
Patient problems: Chronic Obstructive Pulmonary Disease (COPD); No Clinical Signs, Symptoms or Conditions
MAUDE report 24090144 ↗InjuryDec 19, 2025Wireless Communication Problem; Wireless Communication Problem
IT WAS REPORTED THAT THE PATIENT'S SYMPTOMS RETURNED. THE IPG WAS UNABLE TO PROPERLY COMMUNICATE WITH EXTERNAL DEVICES AND DISPLAYED A CHARGE DISCREPANCY. TROUBLESHOOTING HAS BEEN UNABLE TO RESOLVE THE ISSUE. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2025 WHEREIN THE IPG WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE.
Patient problems: Movement Disorder
MAUDE report 23851465 ↗InjuryDec 12, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED PATIENT'S IPG SITE WAS SWELLING. PATIENT UNDERWENT SURGICAL INTERVENTION ON (B)(6) 2025 WHEREIN IPG SITE WAS OPENED AND CLEANED OUT WHAT APPEARED TO BE A HEMATOMA. THERAPY WAS RESTORED POST-OP.
Patient problems: Hematoma
MAUDE report 23794292 ↗InjuryDec 9, 2025High impedance; High impedance
IT WAS REPORTED THAT THE PATIENT'S SYSTEM NEVER ESTABLISHED EFFECTIVE THERAPY DUE TO INCREASING IMPEDANCES. SURGICAL INTERVENTION WAS UNDERTAKEN ON AN UNKNOWN DATE WHEREIN THE SYSTEM WAS EXPLANTED.
Patient problems: Movement Disorder
MAUDE report 23760855 ↗InjuryDec 2, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT THE PATIENT HAD A IPG REPLACEMENT ON (B)(6) 2025 FOR AN UNKNOWN REASON. NO FURTHER INFORMATION IS AVAILABLE AT THIS TIME.
Patient problems: Insufficient Information
MAUDE report 23697433 ↗InjuryNov 18, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ADDITIONAL INFORMATION RECEIVED INDICATES THE LEAD HAD ERODED AT THE BURR HOLE SITE. THIS DEVICE IS NO LONGER CONSIDERED REPORTABLE.
Patient problems: Erosion
MAUDE report 23582788 ↗InjuryNov 11, 2025Low impedance; Low impedance
IT WAS REPORTED THAT THE PATIENT HAD LOW IMPEDANCES FOLLOWING THEIR IPG REPLACEMENT ON (B)(6) 2025. THE PATIENT THEN UNDERWENT SURGICAL INTERVENTION IN WHICH THEIR STITCHES WERE REOPENED AND THE IMPEDANCES RESOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23525935 ↗InjuryNov 5, 2025Wireless Communication Problem; Wireless Communication Problem
IT WAS REPORTED THAT THE IPG BECAME UNABLE TO COMMUNICATE WITH EXTERNAL DEVICES. TROUBLESHOOTING HAS BEEN UNABLE TO RESOLVE THE ISSUE. AS A RESULT, SURGICAL INTERVENTION MAY BE UNDERTAKEN TO ADDRESS THE ISSUE.
Patient problems: Movement Disorder
MAUDE report 23480090 ↗InjuryNov 4, 2025Temperature Problem; Temperature Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED HEATING AT THE IPG SITE. SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE IPG WAS EXPLANTED AND REPLACED.
Patient problems: Burning Sensation
MAUDE report 23469159 ↗InjurySep 10, 2025Unexpected Shutdown; Unexpected Shutdown
IT WAS REPORTED THAT THE PATIENT'S THERAPY WAS TURNING OFF AND ON. SURGICAL INTERVENTION WAS UNDERTAKEN AND THE IPG WAS EXPLANTED AND REPLACED.
Patient problems: Inadequate Pain Relief; Movement Disorder
MAUDE report 23019975 ↗InjuryJul 21, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED INFECTION AT THE IPG SITE. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2025 WHEREIN THE IPG WAS EXPLANTED. THE PATIENT RECEIVED IV ANTIBIOTICS AND WAS HOSPITALIZED.
Patient problems: Post Operative Wound Infection
MAUDE report 22561601 ↗MalfunctionJun 16, 2025Device Displays Incorrect Message; Device Displays Incorrect Message
IT WAS REPORTED THE PATIENTS IPG IS DISPLAYING INCORRECT CHARGE VALUES. TROUBLESHOOTING HAS BEEN UNABLE TO RESOLVE THE ISSUE. THE NEXT COURSE OF ACTION IS UNKNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22214373 ↗MalfunctionMay 29, 2025Wireless Communication Problem; Wireless Communication Problem
IT WAS REPORTED THAT THE PATIENT'S IPG IS UNABLE TO COMMUNICATE WITH EXTERNAL DEVICES. THE PATIENT HAS THERAPY AND THERE IS NO INTERVENTION PLANNED AT THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22106377 ↗InjuryApr 28, 2025Impedance Problem; Impedance Problem
IT WAS REPORTED THAT THE PATIENT'S SYSTEM HAD INVALID IMPEDANCES. SURGICAL INTERVENTION WAS UNDERTAKEN AND THE IPG WAS EXPLANTED AND REPLACED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21922527 ↗InjuryApr 25, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED AN MRSA INFECTION AT THE LEFT IPG SITE. AS A RESULT, SURGICAL INTERVENTION TOOK PLACE ON (B)(6) 2025, WHEREIN THE IPG WAS EXPLANTED. ADDITIONALLY, PATIENT WAS GIVEN ORAL ANTIBIOTICS TO ADDRESS THE ISSUE.
Patient problems: Post Operative Wound Infection
MAUDE report 21910940 ↗InjuryApr 24, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ADDITIONAL INFORMATION WAS RECEIVED STATING THE INFECTION HAS RESOLVED.
Patient problems: Unspecified Infection
MAUDE report 21900925 ↗InjuryApr 22, 2025Material Discolored; Material Discolored
IT WAS REPORTED THAT PATIENT EXPERIENCED INEFFECTIVE THERAPY, PATIENT'S LEAD HAD HIGH IMPEDANCES. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2025 WHEREIN EXTENSION WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE. DURING THE PROCEDURE, IT WAS NOTICED THAT THE IPG WAS DISCOLORED, IPG WAS EXPLANTED AND REPLACED DURING THE PROCEDURE TO ADDRESS THE ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21886618 ↗InjuryMar 11, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED DISCOMFORT AT THE IPG SITE. SURGICAL INTERVENTION MAY TAKE PLACE IN THE FUTURE TO ADDRESS THE ISSUE.
Patient problems: Implant Pain
MAUDE report 21575390 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P140009. One submission often covers many models, so check the product and lot details in each record.
- Z-2694-2017 · Jun 2, 2017 · TerminatedInfinity 5, Implantable Pulse Generator, REF 6661, Rx only, STERILE EO Product Usage: The Infinity IPG is an electronic device designed to be connected to one or two extensions. I
- Z-2695-2017 · Jun 2, 2017 · TerminatedInfinity 7, Implantable Pulse Generator, REF 6662, Rx only, STERILE EO Product Usage: The Infinity IPG is an electronic device designed to be connected to one or two extensions. I
- Z-2693-2017 · Jun 2, 2017 · TerminatedInfinity 5, Implantable Pulse Generator, REF 6660, Rx only, STERILE EO, Product Usage: The Infinity IPG is an electronic device designed to be connected to one or two extensions.
- Z-2696-2017 · Jun 2, 2017 · TerminatedInfinity 7, Implantable Pulse Generator, REF 6663, Rx only, STERILE EO Product Usage: The Infinity IPG is an electronic device designed to be connected to one or two extensions. I
Other Stimulator, electrical, implanted, for essential tremor devices
- Vercise Genus™ DB-1216Boston Scientific Neuromodulation Corporation · Stimulator, electrical, implanted, for essential tremor2 recalls23 deaths584 reports
- SureTek™ DB-4600-CBoston Scientific Neuromodulation Corporation · Stimulator, electrical, implanted, for essential tremor1 recall138 reports
- Activa® SC 37603Medtronic, Inc. · Stimulator, electrical, implanted, for essential tremor1 recall111 reports
- Activa® SC 37602Medtronic, Inc. · Stimulator, electrical, implanted, for essential tremor1 recall85 reports
- SenSight™ B3301542Medtronic, Inc. · Stimulator, electrical, implanted, for essential tremor3 deaths77 reports
- SenSight™ B3301542MMedtronic, Inc. · Stimulator, electrical, implanted, for essential tremor3 deaths43 reports
- N/A 3708660Medtronic, Inc. · Stimulator, electrical, implanted, for essential tremor43 reports
- NA DB-3128-55BBoston Scientific Neuromodulation Corporation · Stimulator, electrical, implanted, for essential tremor38 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page