TriClip™ TCDS0801-NTWRecalls & reported problems
St. Jude Medical, Inc. · G5 Delivery System · Class 3 · Implant
All models of TriClip (7)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 05415067050816), received between Dec 4, 2025 and Dec 4, 2025 — classified as 1 injury.
The most frequently reported specific device problem is “Entrapment of Device” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Entrapment of Device1
Patient problems
- Foreign Body In Patient1
- Tricuspid Valve Insufficiency/ Regurgitation1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryDec 4, 2025Entrapment of Device; Entrapment of Device
IT WAS REPORTED THAT ON (B)(6) 2025, A PATIENT PRESENTED WITH GRADE 3+ SEVERE FOCAL FUNCTIONAL POSTERIOR SEPTAL TRICUSPID REGURGITATION (TR), RESTRICTED SHORT SEPTAL LEAFLET FOR A TRICLIP PROCEDURE. DURING THE PROCEDURE, TRICLIP WAS PLACED ON THE MEDIAL SIDE OF POSTERIOR AND SEPTAL LEAFLET (P/S) BELOW THE TRICUSPID RING. TRICLIP WAS MOVED TO THE AORTIC, GRIPPERS WERE RAISED AND CLIP WAS UNLOCKED, IT WAS DETERMINED THAT THE CLIP WAS STUCK ON THE SEPTAL LEAFLET AND RING. TROUBLE SHOOTING WAS PERFORMED, BUT THE CLIP WAS UNABLE TO BE REMOVED FROM THE SEPTAL RING. SIGNIFICANT FORCE WAS REQUIRED TO REMOVE THE CLIP. THEREFORE, THE CLIP WAS DEPLOYED ON THE SEPTAL LEAFLET. ALL STEPS WERE PERFORMED AS PER IFU DURING THE PREPARATION. IMAGING WAS DIFFICULT DUE TO TRICUSPID RING. ALL STEPS WERE PERFORMED AS PER IFU. THE FINAL TR WAS UNCHANGED. THERE WERE NO CLINICALLY SIGNIFICANT DELAYS OR ADVERSE PATIENT SEQUELA REPORTED.
Patient problems: Foreign Body In Patient; Tricuspid Valve Insufficiency/ Regurgitation
MAUDE report 23721594 ↗
Other Tricuspid valve repair device, percutaneously delivered devices
- TriClip™ TCDS0301-XTWSt. Jude Medical, Inc. · Tricuspid valve repair device, percutaneously delivered2 deaths287 reports
- TriClip™ TCDS0801-XTWSt. Jude Medical, Inc. · Tricuspid valve repair device, percutaneously delivered148 reports
- TriClip™ TCDS0301-XTSt. Jude Medical, Inc. · Tricuspid valve repair device, percutaneously delivered62 reports
- TriClip™ TCDS0801-XTSt. Jude Medical, Inc. · Tricuspid valve repair device, percutaneously delivered49 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page