Skip to content

TactiFlex™ Sensor Enabled™ A-TFSE-FJ-XRecalls & reported problems

St. Jude Medical, Inc. · Ablation Catheter · Class 3

All models of TactiFlex Sensor Enabled (8)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 05415067052018), received between Mar 31, 2026 and Mar 31, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Signal Artifact/Noise” (1 report).

5 other recalls cover devices cleared under the same FDA submission (P220013); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Signal Artifact/Noise1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 31, 2026Signal Artifact/Noise

    DURING A TYPICAL ATRIAL FLUTTER PROCEDURE, THERE WAS A PROCEDURAL DELAY DUE TO A NOISE ISSUE. ELECTRODE 3 ON THE CATHETER WAS NOISY AND DID NOT DISPLAY AS EXPECTED ON THE MAPPING SYSTEM. THE CATHETER, GENERATOR AND TACTISYS CABLES WERE ALL RECONNECTED WITH NO RESOLUTION. THE MOLEX CABLE, RF CABLE, AND TACTISYS WERE REPLACED WITH NO RESOLUTION. THE CATHETER WAS REPLACED AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24749126 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P220013. One submission often covers many models, so check the product and lot details in each record.

  • Z-0113-2026 · Sep 10, 2025 · Open, Classified
    TactiFlex Sensor Enabled Ablation Catheter, 8F, 115 cm, F, (UniD Curve F), REF A-TFSE-F; cardiac catheter
  • Z-0110-2026 · Sep 10, 2025 · Open, Classified
    TactiFlex Sensor Enabled Ablation Catheter, 8F, 115 cm, DD, (BiD Curve D-D), REF A-TFSE-DD; cardiac catheter
  • Z-0111-2026 · Sep 10, 2025 · Open, Classified
    TactiFlex Sensor Enabled Ablation Catheter, 8F, 115 cm, DF, (BiD Curve D-F), REF A-TFSE-DF; cardiac catheter
  • Z-0112-2026 · Sep 10, 2025 · Open, Classified
    TactiFlex Sensor Enabled Ablation Catheter, 8F, 115 cm, FJ, (BiD Curve F-J), REF A-TFSE-FJ; cardiac catheter
  • Z-0109-2026 · Sep 10, 2025 · Open, Classified
    TactiFlex Sensor Enabled Ablation Catheter, 8F, 115 cm, D, (UniD Curve D), REF A-TFSE-D; cardiac catheter

Other Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation devices

All Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page