Amplatzer Piccolo™ 9-PDS-04F-045Recalls & reported problems
St. Jude Medical, Inc. · Delivery System · Class 2
All models of Amplatzer Piccolo (10)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05415067052926), received between Mar 23, 2026 and Sep 25, 2026 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Air/Gas in Device” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Air/Gas in Device1
Patient problems
- Low Blood Pressure/ Hypotension1
- No Clinical Signs, Symptoms or Conditions1
- Pericardial Effusion1
- Tachycardia1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjurySep 25, 2026Adverse Event Without Identified Device or Use Problem
It was reported that on (b)(6) 2026, a 4F Amplatzer Piccolo Delivery System was chosen for a Patent Ductus Arteriosus (PDA) closure procedure on a 3week old, 1000grams premature infant. During procedure, the right femoral vein access was achieved in the usual manner and a 4F access sheath was placed. A non-Abbott wire and Piccolo Delivery System (PDS) combination was advanced up the inferior vena cava (IVC), across the tricuspid valve towards the right ventricular outflow tract (RVOT) and into the pulmonary artery(PA). The non-Abbott wire was advanced in an attempt to engage the PDA, but would not track down the descending aorta. The non-Abbott wire was then exchanged for an another non-Abbott glide wire which took an unusual course and appeared to be in the pericardial space. The baby was stable at this point. The glide-wire and PDS was then withdrawn back into the IVC at which point the baby become hypotensive and tachycardic. A transthoracic echocardiogram (TTE) assessment was done and revealed a rapidly accumulating 5mm circumferential pericardial effusion. The glide wire and PDS was then removed from the body and a pericardiocentesis was performed. A 8ccs of blood was removed
Patient problems: Low Blood Pressure/ Hypotension; Pericardial Effusion; Tachycardia
MAUDE report 50897010 ↗MalfunctionMar 23, 2026Air/Gas in Device
IT WAS REPORTED ON 11 FEBRUARY AN AMPLATZER PICCOLO WAS SELECTED FOR IMPLANT USING A 4F AMPLATZER PICCOLO DELIVERY SYSTEM. ALL STEPS WERE PERFORMED PER INSTRUCTIONS FOR USE (IFU). WHILST INSERTING THE LOADER INTO THE TUOHY CONNECTED TO THE CATHETER WITHIN THE PATIENT, PRIOR TO INSERTING THE DELIVERY SYSTEM INTO THE FEMORAL VEIN, IT WAS NOTED THAT AIR CONTINUOUSLY OBSERVED IN THE HEMOSTATIC VALVE OF THE LOADER. A WET-TO-WET CONNECTION WAS CORRECTLY ESTABLISHED. BACKFLOW WAS USED TO REMOVE AIR BUBBLES. AN ATTEMPT WAS MADE TO DISCONNECT THE COMPONENTS OF THE DELIVERY SYSTEM AND RE-DO DEVICE PREPARATION. THE DEVICE WAS REPLACED WITH A NEW 16MM AMPLATZER POST-INFARCT MUSCULAR VSD OCCLUDER. THE SAME DELIVERY SYSTEM WAS USED TO COMPLETE THE PROCEDURE. THE PATIENT STATUS WAS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24663505 ↗
Other Catheter, percutaneous devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page