CELEROMARKRecalls & reported problems
Hologic, Inc. · CeleroMark Biopsy Site Identification System · Class 2 · Implant
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05420045503376), received between Apr 29, 2025 and Dec 22, 2025 — classified as 1 injury, 1 malfunction.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Appropriate Device Problem Term/Code Not Available1
Patient problems
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 22, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED IN A MEDICAL IMAGING CASE LETTER THAT WAS SUBMITTED ON (B)(6) 2020, TO THE JOURNAL OF MEDICAL IMAGING AND RADIATION ONCOLOGY THAT, AN ULTRASOUND GUIDED VACUUM ASSISTED EXCISION BIOPSY, WAS PERFORMED ON A 60-YEAR-OLD WOMAN ON AN UNKNOWN DATE, USING A 7G ENCORE PROBE AND ENCOR ENSPIRE BREAST BIOPSY SYSTEM. IT IS REPORTED THAT "A DUMBBELL-SHAPED BREAST BIOPSY MARKER," (B)(6) WAS INSERTED; HOWEVER, "MAMMOGRAPHY SHOWED AN AIR-FILLED CAVITY AT THE LESION SITE, COMPLETE REMOVAL OF THE RESIDUAL LESION/TWIRL MARKER AND THE DUMBBELL-SHAPED BREAST BIOPSY MARKER SEVERAL CENTIMETERS FROM THE SITE." IT WAS FURTHER REPORTED THAT "ANOTHER BREAST BIOPSY MARKER, CYLINDRICAL-SHAPED CELEROMARK WAS PLACED TARGETING THE CAVITY; HOWEVER, INACCURATE POSITIONING/MIGRATION WAS AGAIN SEEN." MALFUNCTION MDR IS BEING SUBMITTED DUE TO THE REPORTED MARKER MIGRATION. HOLOGIC AWARENESS OF THIS CASE LETTER WAS NOVEMBER 24, 2025. NO FURTHER INFORMATION IS CURRENTLY AVAILABLE.
Patient problems: Insufficient Information
MAUDE report 23869624 ↗InjuryApr 29, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT DURING A CELEROMARK PROCEDURE ON (B)(6) 2025, THE PHYSICIAN REPORTED THAT ON THE POST ULTRASOUND GUIDED BREAST BIOPSY THE MARKER COULD BE OBSERVED DEPLOYED. THE PATIENT WAS SENT FOR MAMMOGRAM IN SITU. WHILE REVIEWING THE IMAGE ONLY THE PREVIOUS CLIPS WERE VISIBLE AND THE MARKER COULD NOT BE OBSERVED. PATIENT WAS LATER RECALLED FOR ANOTHER IMPLANTATION AND ANOTHER MARKER. PHYSICIAN REPORTED THAT THE FIRST PROCEDURE WAS A SMALL POCKET OF GAS (¨PSEUDO CLIP¨) WHICH COULD NOT BE DIFFERENTIATED. ANOTHER MARKER WAS IMPLANTED ON A SEPARATE INCISION AND CONFIRMED IMPLANTATION. NO OTHER INFORMATION AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21931524 ↗
Other MARKER, RADIOGRAPHIC, IMPLANTABLE devices
- BioZorb F0202Hologic, Inc. · MARKER, RADIOGRAPHIC, IMPLANTABLE2 recalls192 reports
- BioZorb F0221Hologic, Inc. · MARKER, RADIOGRAPHIC, IMPLANTABLE3 recalls15 reports
- BioZorb F0203Hologic, Inc. · MARKER, RADIOGRAPHIC, IMPLANTABLE2 recalls51 reports
- BioZorb F0231Hologic, Inc. · MARKER, RADIOGRAPHIC, IMPLANTABLE2 recalls19 reports
- BioZorb F0304Hologic, Inc. · MARKER, RADIOGRAPHIC, IMPLANTABLE2 recalls19 reports
- BioZorb F0303Hologic, Inc. · MARKER, RADIOGRAPHIC, IMPLANTABLE2 recalls19 reports
- BioZorb F0404Hologic, Inc. · MARKER, RADIOGRAPHIC, IMPLANTABLE2 recalls6 reports
- BioZorb F0331Hologic, Inc. · MARKER, RADIOGRAPHIC, IMPLANTABLE2 recalls4 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.