ABL80 FLEX Analyzer 393-839Recalls & reported problems

Radiometer Medical ApS · ABL80 FLEX Full panel, Analyzer · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05700693938394), received between Sep 27, 2018 and Oct 15, 2025 — classified as 2 injuries.

The most frequently reported specific device problem is “Contamination” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Contamination1
  • Contamination of Device Ingredient or Reagent1
  • Fracture1
  • High Test Results1

Patient problems

  • Abnormal Blood Gases1
  • Hyponatremia1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 15, 2025High Test Results; High Test Results

    THE HOSPITAL REPORTED, ON (B)(6) 2025, AT 15:29, TWO BLOOD SAMPLES WERE SIMULTANEOUSLY DRAWN FROM THE PATIENT. ONE SAMPLE WAS USED FOR BLOOD GAS ANALYSIS ON ABL80 ANALYZER, WHILE THE OTHER WAS SENT TO THE LABORATORY FOR STAT RENAL FUNCTION TESTING. AT 15:49, THE ABL80 RESULT SHOWED A CRITICAL SODIUM VALUE OF 200 MMOL/L. UPON RECEIVING THE REPORT, THE PHYSICIAN IMMEDIATELY ASSESSED THE PATIENT, CONSIDERING POTENTIAL COMPLICATIONS SUCH AS ELECTROLYTE IMBALANCE, MULTIPLE ORGAN DYSFUNCTION, ACUTE KIDNEY INJURY, DEHYDRATION, ARRHYTHMIA, AND SUDDEN CARDIAC ARREST. IMMEDIATE ACTIONS INCLUDED UPGRADING THE PATIENT'S CARE LEVEL TO INTENSIVE, INITIATING CARDIAC MONITORING, DECLARING THE PATIENT'S CONDITION AS CRITICAL, AND ADMINISTERING DIURETICS AND FLUID REPLACEMENT TO LOWER SODIUM LEVELS, WHILE CLOSELY MONITORING THE PATIENT'S STATUS. AT 16:52, THE STAT RENAL FUNCTION TEST REPORT (FROM THE SAMPLE DRAWN AT THE SAME TIME AS THE BLOOD GAS ANALYSIS) SHOWED A SODIUM LEVEL OF 130 MMOL/L, INDICATING HYPONATREMIA. THE PHYSICIAN IMMEDIATELY DISCONTINUED THE DIURETICS AND INITIATED SODIUM SUPPLEMENTATION THERAPY. AT 22:43 THAT SAME DAY, A REPEAT STAT ELECTROLYTE TEST SHOWED A SODIUM LEVEL OF 132.0

    Patient problems: Hyponatremia

    MAUDE report 23290747 ↗
  • InjurySep 27, 2018Contamination; Fracture; Contamination of Device Ingredient or Reagent; Contamination; Fracture; Contamination of Device Ingredient or Reagent

    A PATIENT ANALYSIS WAS PERFORMED ON THE SUBJECT ANALYZER, ABL80 FLEX SERIAL NUMBER (B)(4), AND BASED ON THE REPORTED HCT/THB RESULT THE PATIENT RECEIVED A BLOOD TRANSFUSION. THE USER WAS SUSPICIOUS OF THE SUBJECT ANALYZER'S RESULT AND THEREFORE PERFORMED ANOTHER MEASUREMENT ON A COMPARISON ABL80 FLEX THAT WAS ON-SITE. MEASURED RESULTS: SUBJECT ANALYZER: THB = 7.1G/DL HCT = 22%. COMPARISON ANALYZER: THB = 13.7 G/DL HCT = 42%. BASED ON A COMPARISON OF THE RESULTS BETWEEN THE SUBJECT AND COMPARISON ANALYZERS, THE USER CONCLUDED THAT THE SUBJECT ANALYZER RESULTS WERE INCORRECT AND THE BLOOD TRANSFUSION WAS UNNECESSARY. A SERVICE TECHNICIAN EVALUATED THE ANALYZER AND FOUND THAT THE ROLLER WHEEL ASSEMBLY WAS NOT WORKING PROPERLY. FURTHER INVESTIGATION REVEALED THE ROLLER WHEEL ASSEMBLY HOUSING WAS CRACKED AROUND A SET SCREW THAT IS USED TO SECURE THE ROLLER WHEEL ASSEMBLY TO THE ROLLER MOTOR SHAFT. THE CRACKED HOUSING RESULTED IN ROLLER WHEEL ASSEMBLY SLIPPAGE AFFECTING SAMPLE ASPIRATION AND FLUSHING. THE ISSUE WAS RESOLVED FOLLOWING REPLACEMENT OF THE ROLLER WHEEL ASSEMBLY.

    Patient problems: Abnormal Blood Gases

    MAUDE report 7914865 ↗