AQURE 933-599Recalls & reported problems
Radiometer Medical ApS · AQURE · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 05700699335999), received between Aug 29, 2022 and Nov 21, 2025 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Communication or Transmission Problem” (5 reports).
2 FDA recall records name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Mar 8, 2023Class IIOpen, Classified
Due to potential software issue that may result in patient mix-up information.
Z-1394-2023 · Root cause: Software design · Recall details · FDA record ↗
- Nov 23, 2016Class IITerminated
Design error when displaying additional information in the patient view window; error may result in misreading a parameter and its value.
Z-0899-2017 · Root cause: Software design · Recall details · FDA record ↗
Most reported problems
Device problems
- Communication or Transmission Problem5
- Computer Software Problem4
- Human-Device Interface Problem1
- Patient Data Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 21, 2025Communication or Transmission Problem; Patient Data Problem; Communication or Transmission Problem; Patient Data Problem
RADIOMETER COMPLAINT REFERENCE: (B)(4). ACCORDING TO THE COMPLAINT, ON (B)(6) 2025, THE CUSTOMER EXPERIENCED INCORRECT VALUE DISPLAY IN AQURE. SAMPLE TYPE PARAMETER "FIBTEM C" WAS SENT AS "---" (INDICATING AN INVALID OR MISSING RESULT) IS DISPLAYED AS AN INCORRECT NUMERICAL VALUE IN THE AQURE WEB INTERFACE. NO REPORTS OF DEATH OR SERIOUS INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23616065 ↗MalfunctionSep 15, 2025Computer Software Problem; Communication or Transmission Problem; Computer Software Problem; Communication or Transmission Problem
RADIOMETER COMPLAINT REFERENCE: (B)(4). ACCORDING TO THE COMPLAINT, THE RESULT VALUES DID NOT MATCH AFTER UNIT CONVERSION FACTOR WAS APPLIED BETWEEN THE ANALYZER PATIENT DATA LOGS AND AQURE (SERIAL NUMBER: (B)(6)). OUT OF 583 RESULTS, 21 OF THEM DID NOT MATCH. THIS WAS NOT USED DURING TREATMENT OF PATIENTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23050156 ↗MalfunctionAug 11, 2025Computer Software Problem; Computer Software Problem
RADIOMETER COMPLAINT REFERENCE: (B)(4). ACCORDING TO THE COMPLAINT, THE CUSTOMER IS HAVING ISSUES REPROCESSING RESULTS, AS AN ERROR HAPPENS AFTER ADT LOOK UP WITH AQURE. THE LOOK UP HAS BEEN SUCCESSFUL AS THE LOOK UP BOX IS BELOW THE ERROR BOX. THIS HAPPENS WHEN THE USER PRESSED "OK". [AFTER CLICKING ON INITIAL 'OK', IT RETURNS, 'SAVE ERROR': 'THE SERVER DID NOT SAVE THE SAMPLE. CHECK YOUR NETWORK CONNECTION OR TRY AGAIN LATER.'] THE RESULT THEN STAYS AS PENDING BUT CANNOT BE REPROCESSED [AFTER REPROCESSING, PATIENT DEMOGRAPHICS WAS UPDATED BUT RESULT IS IN "APPROVAL STATUS: PENDING", BUT THE PATIENT DEMOGRAPHICS ARE POPULATED]. NO REPORTS OF DEATH OR SERIOUS INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22768940 ↗MalfunctionNov 29, 2023Human-Device Interface Problem; Human-Device Interface Problem
ACCORDING TO THE COMPLAINT, THE CUSTOMER ENTERED THE WRONG PATIENT DATA THROUGH MANUAL ENTRY ON THE ABL90 FLEX PLUS ANALYZER (SERIAL NUMBER: (B)(6) . THE CUSTOMER THEN APPROVED THE SAMPLE RESULTS FOR THE WRONG PATIENT. THE RESULTS CROSSED OVER TO AQURE AND THEN TO EHR/HIS TO THE WRONG PATIENT. THE CUSTOMER THEN CORRECTED THE PATIENT DEMOGRAPHICS ON THE SAMPLE ON THE MACHINE, LINKING THE RESULTS TO THE CORRECT PATIENT, BUT THIS WAS ONLY CHANGED IN AQURE AND NOT OVER TO THEIR EHR/HIS SYSTEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18225889 ↗MalfunctionMay 15, 2023Communication or Transmission Problem; Communication or Transmission Problem
ACCORDING TO THE COMPLAINT, THE CUSTOMER IS FACING AN ISSUE ON AQURE (VERSION 2.6.1) WHEN RECEIVING PARAMETER CRP FROM LUMIRADX(THIRD-PARTY-DEVICE). WHEN LUMIRADX SENDS <5,0 TO AQURE, THE AQURE SYSTEM SHOWS 50 MG/L. ANOTHER EXAMPLE SHOWS A MEASUREMENT OF 33,4 MG/L WHERE AQURE SHOWS 334 MG/L.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16929025 ↗MalfunctionDec 13, 2022Computer Software Problem; Communication or Transmission Problem; Computer Software Problem; Communication or Transmission Problem
ACCORDING TO THE COMPLAINT THE CUSTOMER EXPERIENCED A MPI (MASTER PATIENT INDEX) ISSUE WITH NONE RADIOMETER DEVICE LUMIRADX ON AQURE (VERSION 2.6.0.). THE QUERY ON LUMIRADX WITH ABNORMAL QUERY SUCCESS ON THE DEVICE WITH ERROR MESSAGE (ON DEVICE) FOLLOWING QUERY. DEVICE SOFTWARE VERSION: DEVICE:V1.21-HUB:V4.2.14.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15968383 ↗MalfunctionAug 29, 2022Computer Software Problem; Communication or Transmission Problem; Computer Software Problem; Communication or Transmission Problem
THE CUSTOMER COMPLAINS ABOUT A PROBLEM WITH AQURE. ACCORDING TO THE COMPLAINT THE CUSTOMER MADE SOME TESTS BASED ON A PROBLEM THEY HAVE WITH A NOT RADIOMETER DEVICE (LUMIRADX). ONCE BY SCANNING A PATIENT TEST ID THE CUSTOMER RECEIVED INFORMATION FROM ANOTHER PATIENT TEST ID. SCENARIO OUTLINED: A) QUERY FROM ANALYZER FOR PATIENT 1234 B) QUERY FAILS C) QUERY FOR PATIENT ABCD D) ANALYZER RECEIVES INFORMATION FOR PATIENT 1234 IMMEDIATELY AFTER THE INCIDENT THE QUERY FUNCTIONALITY WAS DISABLED, HENCE PREVENTING FURTHER INCIDENTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15309348 ↗
Other Medical device data system devices
- Trilogy100 Ventilator, Japan-RentalRespironics, Inc. · Medical device data system2 deaths334 reports
- Trilogy 100 Trilogy100 Ventilator, Bluetooth, ItalyRespironics, Inc. · Medical device data system146 reports
- N/A A901Medtronic, Inc. · Medical device data system1 recall0 reports
- NA 9208000000Stryker Corporation · Medical device data system27 reports
- NA 9205000000Stryker Corporation · Medical device data system19 reports
- Trilogy 200 Trilogy 200 Ventilator, Italy, BluetoothRespironics, Inc. · Medical device data system13 reports
- NA 9206000000Stryker Corporation · Medical device data system9 reports
- NA 9207000000Stryker Corporation · Medical device data system7 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page