Autosoft XC 1001680Recalls & reported problems

Unomedical A/S · Single use infusion set for subcutaneous infusion. The infusion set must be used as described in the Instruction for Use. · Class 2

All models of Autosoft XC (11)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Autosoft XC infusion set has 13,437 adverse event reports, including 2,617 injuries and no deaths. Many describe bent or twisted tubing, parts that loosen or separate, and leaks, and many patient reports mention high blood sugar or diabetic ketoacidosis. Reports rose from 2,878 in the prior 12 months to 5,071 in the latest 12 months. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
14,517
Deaths reported
1
Injuries reported
3,032
Recalls naming it
0

Summary

The FDA's MAUDE database contains 14,517 adverse event reports that list this exact device (DI 05705244016620), received between Dec 13, 2018 and Sep 30, 2026 — classified as 1 death, 3,032 injuries, 11,484 malfunctions.

The most frequently reported specific device problem is “Material Twisted/Bent” (6,340 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 3 reportsJan 2024: 2 reportsFeb 2024: 5 reportsMar 2024: 4 reportsApr 2024: 3 reportsMay 2024: 596 reportsJun 2024: 1,350 reportsJul 2024: 1,139 reportsAug 2024: 750 reportsSep 2024: 665 reportsOct 2024: 552 reportsNov 2024: 487 reportsDec 2024: 167 reportsJan 2025: 370 reportsFeb 2025: 278 reportsMar 2025: 400 reportsApr 2025: 297 reportsMay 2025: 306 reportsJun 2025: 129 reportsJul 2025: 96 reportsAug 2025: 153 reportsSep 2025: 92 reportsOct 2025: 103 reportsNov 2025: 171 reportsDec 2025: 97 reportsJan 2026: 97 reportsFeb 2026: 108 reportsMar 2026: 715 reportsApr 2026: 1,224 reportsMay 2026: 5 reportsJun 2026: 2 reportsJul 2026: 215 reportsAug 2026: 938 reportsSep 2026: 2,579 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520262,579
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Twisted/Bent6,340
  • Loss of or Failure to Bond2,590
  • Premature Separation1,672
  • Leak/Splash1,283
  • Detachment of Device or Device Component940
  • Complete Blockage822
  • Infusion or Flow Problem792
  • Device Handling Problem473

Patient problems

  • No Clinical Signs, Symptoms or Conditions7,071
  • Hyperglycemia6,153
  • Elevated Ketones/ Diabetic Ketoacidosis1,038
  • Hemorrhage/Blood Loss/Bleeding87
  • Skin Inflammation/ Irritation61
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available49
  • Hypoglycemia49
  • Pain36

Recent adverse event reports

The 25 most recent of 14,517 reports, in the reporter's own words as published by the FDA.

  • InjurySep 30, 2026Complete Blockage; Material Twisted/Bent

    It was reported that the patient experienced hyperglycemia caused by three separate bent cannula events at their abdomen insertion site, with each individual infusion set failing in less than 24 hours. Two of these events occurred within the span of a single month, during which the patient noticed symptoms within 3 hours of insertion for one set, and 3 or more hours after insertion for the other. The final bent cannula incident took place on 28-Aug-2026. All resulting high blood glucose (BG) episodes were managed and treated using correction injections via multiple daily injections (MDI).

    Patient problems: Hyperglycemia

    MAUDE report 50951256 ↗
  • InjurySep 30, 2026Complete Blockage; Material Twisted/Bent

    It was reported that the patient experienced an infusion set cannula kink that resulted in severe hyperglycemia, with a blood glucose level of 556 mg/dL. Due to this event, the patient required hospitalization. The patient was admitted on (b)(6)2026 and treated with intravenous (IV) saline fluids and insulin.

    Patient problems: Elevated Ketones/ Diabetic Ketoacidosis

    MAUDE report 50951083 ↗
  • MalfunctionSep 30, 2026Loss of or Failure to Bond; Premature Separation

    It was reported that patient experienced an event on (b)(6) 2026 in which infusion set fell off during use.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50950115 ↗
  • MalfunctionSep 30, 2026Loss of or Failure to Bond; Premature Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50950112 ↗
  • InjurySep 30, 2026Complete Blockage

    It was reported that the patient experienced recurrent occlusion alarms due to a blockage in the infusion set tubing on (b)(6)2026, resulting in hyperglycemia and was treated with correction insulin injection.

    Patient problems: Hyperglycemia

    MAUDE report 50949088 ↗
  • InjurySep 30, 2026Complete Blockage

    It was reported that the patient experienced recurring occlusion alarms associated with the infusion set identifying a blockage in the tubing, resulting high blood glucose and was treated with correction bolus via the pump on (b)(6) 2026.

    Patient problems: Hyperglycemia

    MAUDE report 50948984 ↗
  • InjurySep 30, 2026Complete Blockage

    It was reported that the patient experienced occlusion alarms due to a blockage in the infusion set tubing, resulting in hyperglycemia and was treated with a correction insulin injection via MDI on (b)(6) 2026.

    Patient problems: Hyperglycemia

    MAUDE report 50948981 ↗
  • MalfunctionSep 30, 2026Complete Blockage

    It was reported that the patient experienced an occlusion alarm (Alarm 2) caused by a blockage in the infusion set tubing on (b)(6) 2026.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50948971 ↗
  • InjurySep 30, 2026Complete Blockage; Material Twisted/Bent

    Patient problems: Hyperglycemia

    MAUDE report 50948614 ↗
  • InjurySep 30, 2026Complete Blockage; Material Twisted/Bent

    It was reported that the patient experienced bent cannula event resulting in High Blood Glucose level treated with Changed site, took a bolus on (b)(6) 2026.

    Patient problems: Hyperglycemia

    MAUDE report 50948607 ↗
  • MalfunctionSep 30, 2026Partial Blockage; Material Twisted/Bent

    It was reported that the patient experienced an occlusion alarm associated with a kinked cannula after more than 3 hours of use on (b)(6) 2026.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50948583 ↗
  • MalfunctionSep 30, 2026Partial Blockage; Material Twisted/Bent

    It was reported that the patient experienced occlusion alarms associated with a kinked Cannula on (b)(6) 2026.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50948576 ↗
  • MalfunctionSep 30, 2026Partial Blockage; Material Twisted/Bent

    It was reported that the Patient had experienced infusion set cannula was bent which led to occlusion at site on (b)(6) 2026.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50948331 ↗
  • MalfunctionSep 30, 2026Partial Blockage; Material Twisted/Bent

    It was reported that the patient faced an event where the infusion set which was used for three hours resulted in kinked cannula led to occlusion at site on (b)(6) 2026.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50947960 ↗
  • MalfunctionSep 30, 2026Partial Blockage; Material Twisted/Bent

    It was reported that the patient faced an event where the infusion set which was in use for three hours resulted in kinked cannula led to occlusion and experienced Hyperglycemia resolved by infusion set change on (b)(6) 2026.

    Patient problems: Hyperglycemia

    MAUDE report 50947956 ↗
  • MalfunctionSep 30, 2026Break

    It was reported that the patient experienced Infusion set spring detached from outer ring of inserter prior to insertion event on (b)(6) 2026.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50946885 ↗
  • MalfunctionSep 30, 2026Break

    It was reported that the patient experienced Infusion set spring detached from outer ring of inserter prior to insertion event on (b)(6) 2026.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50946877 ↗
  • MalfunctionSep 30, 2026Break

    It was reported that the patient experienced Infusion set spring detached from outer ring of inserter prior to insertion event on (b)(6) 2026.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50946876 ↗
  • InjurySep 30, 2026Complete Blockage; Material Twisted/Bent

    It was reported that the patient experienced a kinked infusion set cannula and high blood glucose reported. The event was treated with a correction insulin injection via Multiple Daily Injections on (b)(6) 2026.

    Patient problems: Hyperglycemia

    MAUDE report 50946384 ↗
  • InjurySep 30, 2026Complete Blockage; Material Twisted/Bent

    It was reported that the patient experienced a kinked infusion set cannula and high blood glucose reported. The event was treated with a correction insulin injection via Multiple Daily Injections on (b)(6) 2026.

    Patient problems: Hyperglycemia

    MAUDE report 50946382 ↗
  • InjurySep 30, 2026Material Twisted/Bent

    It was reported that the patient experienced a kinked infusion set cannula and high blood glucose reported .The event was treated with a correction bolus via the insulin pump on (b)(6) 2026.

    Patient problems: Hyperglycemia

    MAUDE report 50946085 ↗
  • InjurySep 30, 2026Complete Blockage; Material Twisted/Bent

    It was reported that the patient experienced a bent cannula event on (b)(6)2026. The issue resulted in a high blood glucose level, which was treated with a correction injection via multiple daily injections (MDI).

    Patient problems: Hyperglycemia

    MAUDE report 50945932 ↗
  • InjurySep 30, 2026Complete Blockage; Material Twisted/Bent

    It was reported that the patient experienced a kinked infusion set cannula and high blood glucose reported. The event was treated with a correction insulin injection via Multiple Daily Injections on (b)(6) 2026.

    Patient problems: Hyperglycemia

    MAUDE report 50945926 ↗
  • MalfunctionSep 30, 2026Partial Blockage; Material Twisted/Bent

    It was reported that patient experienced an event on 28-Aug-2026 where infusion set cannula was kinked which leads to occlusion.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50945777 ↗
  • MalfunctionSep 30, 2026Fluid/Blood Leak

    It was reported that the patient faced infusion set tubing leakage issue at the site on (b)(6) 2026. The issue occurred with four infusion sets.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50945561 ↗

FDA clearance and classification

The FDA regulates product code FPA (set, administration, intravascular) under 21 CFR 880.5440 as a Class II device. 21 CFR 880.5440 ↗

Devices of this type normally need 510(k) clearance before sale.

Common questions about the Autosoft XC 1001680

Has the Autosoft XC 1001680 been recalled?

No FDA recall record in our data names this exact device identifier (DI 05705244016620), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Autosoft XC 1001680?

The FDA's MAUDE database lists 14,517 adverse event reports for this device, received between Dec 13, 2018 and Sep 30, 2026. In those reports, the most common specific device problems are “Material Twisted/Bent” (6,340), “Loss of or Failure to Bond” (2,590) and “Premature Separation” (1,672); the most common patient problems are “Hyperglycemia” (6,153), “Elevated Ketones/ Diabetic Ketoacidosis” (1,038) and “Hemorrhage/Blood Loss/Bleeding” (87). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Autosoft XC 1001680?

1 report is classified as a death, 3,032 as injuries and 11,484 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the Autosoft XC 1001680, and is it still sold?

The FDA's device database lists Unomedical A/S as the company responsible for it (brand Autosoft XC, model 1001680). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Set, Administration, Intravascular devices