Peristeen 29127Recalls & reported problems
Coloplast A/S · Peristeen Transanal Irrigation. Accessory unit. 15 rectal catheters Small, 1 water bag. Non sterile. Rx only. · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05708932514934), received between Oct 18, 2019 and Jan 20, 2021 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Deflation Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Deflation Problem1
Patient problems
- Hemorrhage/Blood Loss/Bleeding1
- Pain1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 20, 2021Deflation Problem; Deflation Problem
THIS CASE INVOLVES USE OF THE PERISTEEN TRANSANAL IRRIGATION SYSTEM ON A CHILD. AFTER USE OF THE DEVICE ON THE CHILD THE BALLOON ON THE CATHETER FAILED TO DEFLATE. THE BLUE TUBE WAS DISCONNECTED FROM THE DIAL; HOWEVER, THE BALLOON REMAINED INFLATED. IN ORDER TO DEFLATE THE BALLOON SUCCESSFULLY THE CLEAR PART OF THE BLUE TUBE HAD TO BE PHYSICALLY CUT BY THE USER. THE CHILD WAS NOT INJURED; HOWEVER, IT WAS STATED THAT THE CHILD WAS IN GREAT PAIN DUE TO THE INABILITY TO EVACUATE THE WATER AND STOOL PRIOR TO THE BALLOON DEFLATION. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: Pain
MAUDE report 11204398 ↗InjuryOct 18, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE AVAILABLE INFORMATION, DURING THE INITIAL TRAINING FOR PERISTEEN, THE DAD INSERTED THE CATHETER INSIDE THE END USER'S VAGINA INSTEAD OF HER RECTUM AND THEN INFLATED THE BALLOON. THIS RESULTED IN BLEEDING THAT WAS HEAVY AND LASTED QUITE A LONG TIME. THE END USER WAS TAKEN TO THE HOSPITAL AND WAS EXAMINED. THE SURGEON PACKED THE VAGINA WITH DRESSING AND THE END USER WAS ADMITTED TO THE HOSPITAL WHERE SHE STAYED FOR TWO NIGHTS. THE WOUND WAS TREATED AS A LACERATION TO THE LINING OF THE VAGINAL AREA. THE END USER EXPERIENCED PAIN AS A RESULT OF THIS INCIDENT THAT LASTED PERHAPS 10 MINUTES. AFTER THAT, SHE WAS SHARING WITH THE NURSE HER ACTIVITIES FROM EARLIER IN THE DAY. DURING HER TIME IN THE HOSPITAL, THERE WERE NO COMPLICATIONS. THE WOUND "OOZED" AND THIS WAS CONSIDERED NORMAL AND EXPECTED BY THE SURGEON. SHE HAS HAD NO SPOTTING SINCE RETURNING HOME AND IS SAID TO BE "COMPLETELY OK" BY THE NURSE. THE FAMILY WILL BE RETURNING FOR TRAINING FOR THE PERISTEEN SYSTEM BECAUSE THEY BELIEVE IT WILL BE BENEFICIAL FOR THE END USER. MOM EXPLAINED THAT THE END USER IS STILL SPOTTING AND MAY EXPERIENCE SPOTTING FOR 2 WEEKS. THE END USER PASSED A CLOT OF BLOOD WHILE AT SCHOOL. TH
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 9211644 ↗
Other TUBES, GASTROINTESTINAL (AND ACCESSORIES) devices
- Bard® DigniShield® Stool Management System SMS002C. R. Bard, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)2 deaths525 reports
- CORTRAK* 2 ENTERAL ACCESS SYSTEM 20-0950Avanos Medical, Inc. · Tubes, gastrointestinal (and accessories)1 recall22 deaths134 reports
- Bard Tri-Funnel Replacement Gastrostomy Tube 000720Bard Access Systems, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 death170 reports
- EndoVive Safety PEG Kit M00566470Boston Scientific Corporation · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 recall1 death111 reports
- Kangaroo 8884711253Cardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)152 reports
- NutraGlide Nasal Feeding Tube ENFT-05055-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
- NutraGlide Nasal Feeding Tube ENFT-06090-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
- NutraGlide Nasal Feeding Tube ENFT-06055-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page