Nexstim NBS System 5Recalls & reported problems
Nexstim Oyj · Nexstim NBS System · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 06430063200057), received between May 5, 2017 and May 5, 2017 — classified as 1 malfunction.
The most frequently reported specific device problem is “Image Reversal” (1 report).
1 other recall cover devices cleared under the same FDA submission (K112881); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Image Reversal1
- Improper or Incorrect Procedure or Method1
Patient problems
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 5, 2017Image Reversal; Improper or Incorrect Procedure or Method; Image Reversal; Improper or Incorrect Procedure or Method
THIS IS NEXSTIM'S FIRST MDR AND NEXSTIM WANTS TO SELECT MOST CONSERVATIVE APPROACH. NEXSTIM WANTS TO REPORT IT DESPITE NEXSTIM'S RISK AND REPORTABILITY ASSESSMENT REF: (B)(4) HAS DEMONSTRATED THE EVENT DOES NOT MEET THE DEFINITION OF A REPORTABLE EVENT TO FDA BECAUSE IT HAS BEEN ALSO REPORTED TO FINNISH COMPETENT AUTHORITY VALVIRA. THE REPORT WAS INITIALLY SENT TO FDA ON MARCH 15, 2017 ON A PAPER FORMAT DUE TO TECHNICAL DIFFICULTIES NEXSTIM HAS HAD WITH SETTING UP FULLY FUNCTIONAL AND VALIDATED ESG PRODUCTION ACCOUNT. THE PAPER FORMAT WAS REJECTED BY FDA AND IT WAS RETURNED TO NEXSTIM (RECEIVED ON APRIL 3, 2017). THE PROBLEM: NEXSTIM HAS BECOME AWARE OF AN ISSUE WITH THE NBS SYSTEM 4 SOFTWARE WHICH ALSO AFFECTS NBS SYSTEM 5 SOFTWARE. WHEN A USER, WHILE COPYING MR IMAGE SERIES INTO THE NBS SYSTEM, ACCIDENTALLY GENERATES DUPLICATE COPY OF ONE OR SEVERAL FILES, NBS SOFTWARE WILL IN CERTAIN CIRCUMSTANCES, INCORRECTLY INTERPRET LEFT-RIGHT ORIENTATION AND DISPLAY THE HEAD 3D MODEL MIRRORED ON THE NBS SYSTEM DISPLAY. IF MR IMAGES ARE TRANSFERRED AS INSTRUCTED AND TRAINED, WITHOUT DUPLICATE COPIES, MIRRORING ON THE NBS DISPLAY WILL NOT HAPPEN. MIRRORING IS LIMITED ONLY TO HOW NBS DISPLAYS
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6546956 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K112881. One submission often covers many models, so check the product and lot details in each record.
- Z-2320-2017 · Mar 14, 2017 · Open, ClassifiedNBS System 4 (sw version 4.0 or higher), Software update to 4.3.3 and NBS System 5 (sw version 5.0 or higher), Software update to 5.1.1. The Nexstim Navigated Brain Stimulation Sys
Other Stimulator, Electrical, Evoked Response devices
- NIM Vital™ NIM4CM01Medtronic Xomed, Inc. · STIMULATOR, ELECTRICAL, EVOKED RESPONSE3 recalls795 reports
- NIM Vital™ NIM4CPB1Medtronic Xomed, Inc. · STIMULATOR, ELECTRICAL, EVOKED RESPONSE3 recalls331 reports
- NIM Vital NIM4CPB1Medtronic Xomed, Inc. · Stimulator, electrical, evoked response2 recalls208 reports
- POWEREASE™ Instruments 2300000Medtronic Sofamor Danek, Inc. · STIMULATOR, ELECTRICAL, EVOKED RESPONSE1 recall203 reports
- NIM Vital™ NIM4CM01Medtronic Xomed, Inc. · STIMULATOR, ELECTRICAL, EVOKED RESPONSE3 recalls47 reports
- NIM-Response® 8253200Medtronic Xomed, Inc. · STIMULATOR, ELECTRICAL, EVOKED RESPONSE113 reports
- NA DAQ916Medtronic Xomed, Inc. · STIMULATOR, ELECTRICAL, EVOKED RESPONSE80 reports
- Neuromaster G1 DC-200BNihon Kohden Corporation · STIMULATOR, ELECTRICAL, EVOKED RESPONSE73 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page