Skip to content

Inion FreedomPin™ 1 (REF OPA-1550)Recalls & reported problems

Inion Oy · Pin 1.5 x 50 mm · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 06438408004155), received between Jul 14, 2025 and Jul 14, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Biocompatibility” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Biocompatibility1

Patient problems

  • Inflammation1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 14, 2025Biocompatibility; Adverse Event Without Identified Device or Use Problem; Biocompatibility; Adverse Event Without Identified Device or Use Problem

    A 6-YEAR-OLD MALE PATIENT WITH BIG TOE FRACTURE (PROXIMAL PHALANG) WAS OPERATED WITH INION FREEDOMPIN Ø1,5 MM. 2 YEARS AFTER OPERATION PATIENT SUFFERED OF REDNESS & SWELLING OF THE IP-JOINT. X-RAY, MRI, LABORATORIES (LEUK, CRP) WERE TAKEN. LABORATORY RESULTS WERE NORMAL. IN MRI THE PIN WAS PARTIALLY ABSORBED, DISTAL PART STILL INSIDE THE PHALANGE. THERE WAS EDEMA AROUND THE PIN COVERING THE WHOLE PHALANGE WHICH LED THE SURGEON TO SUSPECT ROOT CAUSES SUCH AS OSTEOMYELITIS/ARTHRITIS AND FOREIGN BODY REACTION. PATIENT HAD ANTIBIOTICS FOR 10 DAYS, HOWEVER NO SAMPLES TAKEN, BECAUSE THE PATIENT HAD NO FEVER, OR PAIN IN THE TOE AND LABORATORY SAMPLES WERE NORMAL. THE SWELLING DISAPPEARED IN A MONTH. RE-OPERATION WAS NOT NEEDED. X-RAY 4 MONTHS LATER SHOWED A LARGE EROSION (LYTIC LESION) INSIDE THE FIRST PHALANGE. X-RAYS AND MRI WERE SENT TO UNIVERSITY HOSPITAL FOR FURTHER EXAMINATIONS IF NEEDED. PATIENT'S CURRENT CONDITION IS NORMAL.

    Patient problems: Inflammation

    MAUDE report 22492918 ↗

Other PIN, FIXATION, SMOOTH devices

All PIN, FIXATION, SMOOTH devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page