Injection Needle IN12-255232302Recalls & reported problems
Micro-Tech (Nanjing) Co., Ltd. · POM inner tuber,Needle Length 5mm,with metal cap 25G*5*2.3*230,Shroat stucture · Class 2
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 06932503533249), received between Feb 27, 2024 and Sep 23, 2024 — classified as 1 malfunction.
The most frequently reported specific device problem is “Device Fell” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Device Fell1
Patient problems
- Foreign Body In Patient1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 23, 2024Device Fell; Device Fell
MICRO-TECH RECEIVED A MDR NOTIFICATION FROM FDA. MDR REPORT #: MW5152104. IT WAS REPORTED THAT INJECTION NEEDLE HAS A CLEAR TIP COVER THAT IS THE SAME COLOR AS THE TUBING AND EASILY FITS THROUGH THE SCOPE CHANNEL. THE USER INADVERTENTLY FORGOT TO REMOVE THE TIP COVER AND THE TIP CAME OFF IN THE PATIENTS COLON AND HAD TO BE RETRIEVED. THERE WAS NO INJURY TO PATIENT DUE TO THIS ERROR. RECOMMEND TIP COVER COLOR AND SIZE BE REVISED TO AVOID FUTURE INCIDENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20277425 ↗OtherFeb 27, 2024
INJECTION NEEDLE HAS A CLEAR TIP COVER THAT IS THE SAME COLOR AS THE TUBING AND EASILY FITS THROUGH THE SCOPE CHANNEL. THE USER INADVERTENTLY FORGOT TO REMOVE THE TIP COVER AND THE TIP CAME OFF IN THE PATIENTS COLON AND HAD TO BE RETRIEVED. THERE WAS NO INJURY TO PATIENT DUE TO THIS ERROR. RECOMMEND TIP COVER COLOR AND SIZE BE REVISED TO AVOID FUTURE INCIDENT.
Patient problems: Foreign Body In Patient
MAUDE report 18799723 ↗
Other Endoscopic Injection Needle, Gastroenterology-Urology devices
- AcuJect G22526Cook Incorporated · endoscopic injection needle, gastroenterology-urology1 recall1 report
- n.a. 27182AKarl Storz GmbH & Co. KG · endoscopic injection needle, gastroenterology-urology1 recall0 reports
- injeTAK Adjustable Tip Needle 35cm DIS199Laborie Medical Technologies Canada ULC · endoscopic injection needle, gastroenterology-urology1 recall0 reports
- n.a. 27030NKarl Storz GmbH & Co. KG · endoscopic injection needle, gastroenterology-urology1 recall0 reports
- injeTAK Adjustable Tip Needle 70cm DIS201Laborie Medical Technologies Canada ULC · endoscopic injection needle, gastroenterology-urology1 recall0 reports
- Single Use Injector NM-401L-0425Olympus Medical Systems Corp. · endoscopic injection needle, gastroenterology-urology28 reports
- Single Use Injector NM-401L-0423Olympus Medical Systems Corp. · endoscopic injection needle, gastroenterology-urology27 reports
- Single Use Injector NM-401L-0425Olympus Medical Systems Corp. · endoscopic injection needle, gastroenterology-urology26 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.