Skip to content

Spray Catheter WP-18/2200Recalls & reported problems

Micro-Tech (Nanjing) Co., Ltd. · Blue handle 1.8mm OD 2200mm length · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 06932503562249), received between Dec 31, 2024 and Apr 23, 2026 — classified as 8 malfunctions.

The most frequently reported specific device problem is “Crack” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 2 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Crack2
  • Obstruction of Flow2
  • Defective Device1
  • Deformation Due to Compressive Stress1
  • Difficult to Remove1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • Hemorrhage/Blood Loss/Bleeding1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 23, 2026Difficult to Remove

    IT WAS REPORTED BY THE HOSPITAL STAFF THAT THE PRODUCT WAS UNABLE TO EXPEL FROM CATH.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24963329 ↗
  • MalfunctionApr 16, 2026

    FROM STAFF: UNABLE TO EXPEL FROM CATH.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24895066 ↗
  • MalfunctionMar 20, 2026Deformation Due to Compressive Stress

    IT WAS REPORTED THAT KINK IN CATHETER PRIOR TO OPENING; UNABLE TO PUSH PRODUCT PROPERLY AND CAUSES LEAKS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24646892 ↗
  • MalfunctionMar 4, 2026

    FROM STAFF: KINK IN CATHETER PRIOR TO OPENING; UNABLE TO PUSH PRODUCT PROPERLY AND CAUSES LEAKS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24505660 ↗
  • MalfunctionAug 26, 2025Obstruction of Flow; Obstruction of Flow

    FROM STAFF: DURING AN EGD [ESOPHAGOGASTRODUODENOSCOPY] NEEDED TO USE PURASTAT TO STOP BLEEDING AND THE SPRAY CATHETER WOULD NOT PUSH THE MEDICATION THROUGH. SPRAY CATHETER REF#: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22892517 ↗
  • MalfunctionJul 15, 2025Obstruction of Flow; Defective Device

    FROM STAFF: DURING AN EGD [ESOPHAGOGASTRODUODENOSCOPY] NEEDED TO USE PURASTAT TO STOP BLEEDING AND THE SPRAY CATHETER WOULD NOT PUSH THE MEDICATION THROUGH. SPRAY CATHETER REF# WP-18/2200.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 22506476 ↗
  • MalfunctionJan 17, 2025Crack; Crack

    ON 12/30/2024, MICRO-TECH RECEIVED 1 INCIDENT REPORT REGARDING SPRAY CATHETER FROM FDA. UF REPORT#: (B)(4). THIS IS THE SECOND REPORT FOR ANOTHER DEVICE USED IN OTHER DIFFERENT PATIENT. OUR FIRST REPORT REFERENCE NUMBER: (B)(4). IT WAS REPORTED THAT -D MATRIX PURASTAT SPRAY CATHETER LOT#: M230925311 REF#: WP-18/2200, ADAPTER APPLIED TO PURASTAT GEL SYRINGE AND SPRAY CATHETER AND PURASTAT GEL INJECTED INTO CATHETER SYRINGE THEN DISCONNECTED AND FILLED WITH AIR AS IFU INSTRUCTS. AFTER SYRINGE APPLIED BACK ONTO SPRAY CATHETER TO DELIVER THE GEL FROM THE CATHETER IT WAS FOUND TO HAVE NO GEL EXPELLING CATHETER. UPON FURTHER EXAMINATION IT APPEARED THAT BLUE CONNECTION PORT ON SPRAY CATHETER WAS CRACKED WHICH KEPT THE PRODUCT FROM DISPENSING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21176849 ↗
  • MalfunctionDec 31, 2024Crack; Crack

    ON 12/30/2024, MICRO-TECH RECEIVED 1 INCIDENT REPORT REGARDING SPRAY CATHETER FROM FDA. UF REPORT#: (B)(4). IT WAS REPORTED THAT -D MATRIX PURASTAT SPRAY CATHETER LOT# M230925311 REF#(B)(4) ADAPTER APPLIED TO PURASTAT GEL SYRINGE AND SPRAY CATHETER AND PURASTAT GEL INJECTED INTO CATHETER SYRINGE THEN DISCONNECTED AND FILLED WITH AIR AS IFU INSTRUCTS. AFTER SYRINGE APPLIED BACK ONTO SPRAY CATHETER TO DELIVER THE GEL FROM THE CATHETER IT WAS FOUND TO HAVE NO GEL EXPELLING CATHETER. UPON FURTHER EXAMINATION IT APPEARED THAT BLUE CONNECTION PORT ON SPRAY CATHETER WAS CRACKED WHICH KEPT THE PRODUCT FROM DISPENSING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21046059 ↗

Other Adaptor, Catheter devices

All Adaptor, Catheter devices →

Sources and updates