BeneVision N1 Patient monitorRecalls & reported problems
Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 7
- Deaths reported
- 3
- Injuries reported
- 2
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 06944904094922), received between Oct 4, 2024 and Sep 26, 2025 — classified as 3 deaths, 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Device Alarm System” (5 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Aug 15, 2025Class IIOpen, Classified
Potential for activation of an abnormal alarm pause.
Z-2622-2025 · Root cause: Software design · Recall details · FDA record ↗
Most reported problems
Device problems
- Device Alarm System5
- No Audible Alarm5
- Failure to Deflate2
- Alarm Not Visible1
- Protective Measures Problem1
Patient problems
- Insufficient Information4
- Ecchymosis2
- Ischemia2
- Neuropathy2
- Thrombosis/Thrombus2
- Asystole1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjurySep 26, 2025Failure to Deflate; Failure to Deflate
ON (B)(6) 2025, DURING A SURGICAL PROCEDURE BEGINNING AT 07:30 AM AND ENDING AT 11:22 AM, THE BLOOD PRESSURE CUFF REMAINED INFLATED FOR A PERIOD EXCEEDING THREE HOURS. ON (B)(6) 2025, THE CUSTOMER ADDITIONALLY REPORTED THE PATIENT'S ARM WAS TUCKED AND INACCESSIBLE. THE ARTERIAL LINE WAS PLACED IN THE CONTRALATERAL ARM AND PROVIDED RELIABLE BLOOD PRESSURE READINGS. THE BLOOD PRESSURE CUFF WAS REMOVED FROM THE PATIENT. THE PATIENT EXPERIENCED ISCHEMIA OF THE UPPER EXTREMITY FOR APPROXIMATELY THREE HOURS, INCLUDED DEEP VEIN THROMBOSIS (DVT) IN THE ARM VEINS, SKIN DISCOLORATION, AND NEUROPATHY.
Patient problems: Ecchymosis; Ischemia; Neuropathy; Thrombosis/Thrombus
MAUDE report 23156387 ↗InjurySep 26, 2025Failure to Deflate; Failure to Deflate
ON (B)(6) 2025, DURING A SURGICAL PROCEDURE BEGINNING AT 07:30 AM AND ENDING AT 11:22 AM, THE BLOOD PRESSURE CUFF REMAINED INFLATED FOR A PERIOD EXCEEDING THREE HOURS. ON SEPTEMBER 18, 2025, THE CUSTOMER ADDITIONALLY REPORTED THE PATIENT'S ARM WAS TUCKED AND INACCESSIBLE. THE ARTERIAL LINE WAS PLACED IN THE CONTRALATERAL ARM AND PROVIDED RELIABLE BLOOD PRESSURE READINGS. THE BLOOD PRESSURE CUFF WAS REMOVED FROM THE PATIENT. THE PATIENT EXPERIENCED ISCHEMIA OF THE UPPER EXTREMITY FOR APPROXIMATELY THREE HOURS, INCLUDED DEEP VEIN THROMBOSIS (DVT) IN THE ARM VEINS, SKIN DISCOLORATION, AND NEUROPATHY.
Patient problems: Ecchymosis; Ischemia; Neuropathy; Thrombosis/Thrombus
MAUDE report 23156368 ↗DeathJul 30, 2025Device Alarm System; No Audible Alarm; Alarm Not Visible; Device Alarm System; No Audible Alarm; Alarm Not Visible
IT WAS REPORTED THAT ON JULY 7, 2025, AT 10:00 P.M., A PATIENT WAS CONNECTED TO A BENEVISION N12 PATIENT MONITOR (SERIAL NUMBER (B)(6)) IN USE WITH A BENEVISION N1 (SERIAL NUMBER (B)(6)). THE PATIENT EXPERIENCED ASYSTOLE; HOWEVER, NO ALARM WAS GENERATED.
Patient problems: Asystole
MAUDE report 22664532 ↗DeathDec 22, 2024Device Alarm System; No Audible Alarm; Protective Measures Problem; Device Alarm System; No Audible Alarm; Protective Measures Problem
THE CUSTOMER REPORTED THAT A PATIENT WAS CONNECTED TO SPO2 AND THE ECG DISPLAYED "LEADS OFF" AS THE PATIENT WAS MOVING TOO MUCH. ECG ALARM WAS MISSED AND THE PATIENT WAS FOUND UNRESPONSIVE AND CPR WAS INITIATED. RESUSCITATION EFFORTS WERE NOT SUCCESSFUL AND THE PATIENT EXPIRED ON (B)(6) 2024 AT 2:30AM.
Patient problems: Insufficient Information
MAUDE report 20997805 ↗DeathDec 22, 2024Device Alarm System; No Audible Alarm; Device Alarm System; No Audible Alarm
THE CUSTOMER REPORTED THAT A PATIENT WAS CONNECTED TO SPO2 AND THE ECG DISPLAYED "LEADS OFF" AS THE PATIENT WAS MOVING TOO MUCH. ECG ALARM WAS MISSED AND THE PATIENT WAS FOUND UNRESPONSIVE AND CPR WAS INITIATED. RESUSCITATION EFFORTS WERE NOT SUCCESSFUL AND THE PATIENT EXPIRED ON (B)(6) 2024 AT 2:30AM.
Patient problems: Insufficient Information
MAUDE report 20997804 ↗MalfunctionOct 4, 2024Device Alarm System; No Audible Alarm; Device Alarm System; No Audible Alarm
IT WAS REPORTED THAT ON (B)(6) 2024, AT 10:54 P.M., THE BENEVISION N1 PATIENT MONITOR DID NOT GENERATE A LOW SPO2 AUDIO ALARM. NO PATIENT INJURY WAS REPORTED.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 20376024 ↗MalfunctionOct 4, 2024Device Alarm System; No Audible Alarm; Device Alarm System; No Audible Alarm
IT WAS REPORTED THAT ON (B)(6) 2024, AT 10:54 P.M., THE BENEVISION N1 PATIENT MONITOR DID NOT GENERATE A LOW SPO2 AUDIO ALARM. NO PATIENT INJURY WAS REPORTED.
Patient problems: Insufficient Information
MAUDE report 20373686 ↗
Other Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms) devices
- IntelliVue MX40 1.4GHz MX40 1.4 GHz Smart HoppingPhilips Medical Systems Hsg · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)1 recall47 deaths2,541 reports
- NA PU-621RNihon Kohden Corporation · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)17 deaths1,666 reports
- Tempus Pro 00-1024-RRemote Diagnostic Technologies Limited · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)7 recalls8 deaths1,217 reports
- NA PU-681RANihon Kohden Corporation · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)22 deaths1,493 reports
- IntelliVue Multi-Measurement Module X3 IntelliVue X3Philips Medizin Systeme Böblingen GmbH · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)1 recall19 deaths1,172 reports
- Patient Information Center IXPhilips Medical Systems Hsg · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)82 deaths941 reports
- Patient Information Center iXPhilips Medical Systems Hsg · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)83 deaths814 reports
- Tempus Pro 00-1024-RRemote Diagnostic Technologies Limited · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)5 recalls541 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page