BeneVision N1 Patient monitorRecalls & reported problems

Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
7
Deaths reported
3
Injuries reported
2
Recalls naming it
1

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 06944904094922), received between Oct 4, 2024 and Sep 26, 2025 — classified as 3 deaths, 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Device Alarm System” (5 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 2 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 2 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 15, 2025Class IIOpen, Classified

    Potential for activation of an abnormal alarm pause.

    Z-2622-2025 · Root cause: Software design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Device Alarm System5
  • No Audible Alarm5
  • Failure to Deflate2
  • Alarm Not Visible1
  • Protective Measures Problem1

Patient problems

  • Insufficient Information4
  • Ecchymosis2
  • Ischemia2
  • Neuropathy2
  • Thrombosis/Thrombus2
  • Asystole1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • InjurySep 26, 2025Failure to Deflate; Failure to Deflate

    ON (B)(6) 2025, DURING A SURGICAL PROCEDURE BEGINNING AT 07:30 AM AND ENDING AT 11:22 AM, THE BLOOD PRESSURE CUFF REMAINED INFLATED FOR A PERIOD EXCEEDING THREE HOURS. ON (B)(6) 2025, THE CUSTOMER ADDITIONALLY REPORTED THE PATIENT'S ARM WAS TUCKED AND INACCESSIBLE. THE ARTERIAL LINE WAS PLACED IN THE CONTRALATERAL ARM AND PROVIDED RELIABLE BLOOD PRESSURE READINGS. THE BLOOD PRESSURE CUFF WAS REMOVED FROM THE PATIENT. THE PATIENT EXPERIENCED ISCHEMIA OF THE UPPER EXTREMITY FOR APPROXIMATELY THREE HOURS, INCLUDED DEEP VEIN THROMBOSIS (DVT) IN THE ARM VEINS, SKIN DISCOLORATION, AND NEUROPATHY.

    Patient problems: Ecchymosis; Ischemia; Neuropathy; Thrombosis/Thrombus

    MAUDE report 23156387 ↗
  • InjurySep 26, 2025Failure to Deflate; Failure to Deflate

    ON (B)(6) 2025, DURING A SURGICAL PROCEDURE BEGINNING AT 07:30 AM AND ENDING AT 11:22 AM, THE BLOOD PRESSURE CUFF REMAINED INFLATED FOR A PERIOD EXCEEDING THREE HOURS. ON SEPTEMBER 18, 2025, THE CUSTOMER ADDITIONALLY REPORTED THE PATIENT'S ARM WAS TUCKED AND INACCESSIBLE. THE ARTERIAL LINE WAS PLACED IN THE CONTRALATERAL ARM AND PROVIDED RELIABLE BLOOD PRESSURE READINGS. THE BLOOD PRESSURE CUFF WAS REMOVED FROM THE PATIENT. THE PATIENT EXPERIENCED ISCHEMIA OF THE UPPER EXTREMITY FOR APPROXIMATELY THREE HOURS, INCLUDED DEEP VEIN THROMBOSIS (DVT) IN THE ARM VEINS, SKIN DISCOLORATION, AND NEUROPATHY.

    Patient problems: Ecchymosis; Ischemia; Neuropathy; Thrombosis/Thrombus

    MAUDE report 23156368 ↗
  • DeathJul 30, 2025Device Alarm System; No Audible Alarm; Alarm Not Visible; Device Alarm System; No Audible Alarm; Alarm Not Visible

    IT WAS REPORTED THAT ON JULY 7, 2025, AT 10:00 P.M., A PATIENT WAS CONNECTED TO A BENEVISION N12 PATIENT MONITOR (SERIAL NUMBER (B)(6)) IN USE WITH A BENEVISION N1 (SERIAL NUMBER (B)(6)). THE PATIENT EXPERIENCED ASYSTOLE; HOWEVER, NO ALARM WAS GENERATED.

    Patient problems: Asystole

    MAUDE report 22664532 ↗
  • DeathDec 22, 2024Device Alarm System; No Audible Alarm; Protective Measures Problem; Device Alarm System; No Audible Alarm; Protective Measures Problem

    THE CUSTOMER REPORTED THAT A PATIENT WAS CONNECTED TO SPO2 AND THE ECG DISPLAYED "LEADS OFF" AS THE PATIENT WAS MOVING TOO MUCH. ECG ALARM WAS MISSED AND THE PATIENT WAS FOUND UNRESPONSIVE AND CPR WAS INITIATED. RESUSCITATION EFFORTS WERE NOT SUCCESSFUL AND THE PATIENT EXPIRED ON (B)(6) 2024 AT 2:30AM.

    Patient problems: Insufficient Information

    MAUDE report 20997805 ↗
  • DeathDec 22, 2024Device Alarm System; No Audible Alarm; Device Alarm System; No Audible Alarm

    THE CUSTOMER REPORTED THAT A PATIENT WAS CONNECTED TO SPO2 AND THE ECG DISPLAYED "LEADS OFF" AS THE PATIENT WAS MOVING TOO MUCH. ECG ALARM WAS MISSED AND THE PATIENT WAS FOUND UNRESPONSIVE AND CPR WAS INITIATED. RESUSCITATION EFFORTS WERE NOT SUCCESSFUL AND THE PATIENT EXPIRED ON (B)(6) 2024 AT 2:30AM.

    Patient problems: Insufficient Information

    MAUDE report 20997804 ↗
  • MalfunctionOct 4, 2024Device Alarm System; No Audible Alarm; Device Alarm System; No Audible Alarm

    IT WAS REPORTED THAT ON (B)(6) 2024, AT 10:54 P.M., THE BENEVISION N1 PATIENT MONITOR DID NOT GENERATE A LOW SPO2 AUDIO ALARM. NO PATIENT INJURY WAS REPORTED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 20376024 ↗
  • MalfunctionOct 4, 2024Device Alarm System; No Audible Alarm; Device Alarm System; No Audible Alarm

    IT WAS REPORTED THAT ON (B)(6) 2024, AT 10:54 P.M., THE BENEVISION N1 PATIENT MONITOR DID NOT GENERATE A LOW SPO2 AUDIO ALARM. NO PATIENT INJURY WAS REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 20373686 ↗

Other Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms) devices

All Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms) devices →