BeneVision N12 Patient MonitorRecalls & reported problems

Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 06944904095592), received between Dec 10, 2025 and Dec 10, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Device Alarm System” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 2 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Alarm System2

Patient problems

  • Insufficient Information2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 10, 2025Device Alarm System; Device Alarm System

    PER MEDWATCH REPORT REFERENCE # MW5177989: NURSE WAS GOING THROUGH EVENT LOGS TO PRINT OFF A STRIP FROM THE PATIENT'S RECORDS. WHILE GOING THROUGH THE RECORDS, A RUN OF PREMATURE VENTRICULAR COMPLEXES (PVCS) WERE FOUND AT A TACHYCARDIC RATE AND THERE WAS NO HIGHLIGHTING OF THAT TRACE TO INDICATE ANY ALARMS. THERE WAS ALSO NO CHANGE IN CALCULATED HEART RATE. THERE WERE NO ALARMS THAT SOUNDED AT THE TIME OF EVENT. FROM THE ATTACHED IMAGE OF THE STRIP, IT IS EVIDENT THAT THERE WAS A VENTRICULAR TACHYCARDIA (V-TACH) EVENT. AFTER THIS WAS FOUND, STAFF GOT BIOMED INVOLVED TO DO AN INVESTIGATION. FROM BIOMED'S FINDINGS, THE MULTIPARAMETER MODULE (MPM) MODULE THAT WAS IN THE NEMESIS PRIME (N12) MONITOR WAS UNABLE TO ANNOTATE THE QRS COMPLEX. THE COMPATIBILITY ISSUE IS BELIEVED TO BE WHY THERE WAS NO ALARM. MEANWHILE, THE PATIENT WAS MOVED TO A DIFFERENT ROOM WITH A NEW N12 AND N1 MODULE CONNECTED AND THERE WAS A V-TACH ALARM FOR A SIMILAR RUN OF PVCS. MINDRAY WAS CALLED AND CREATED AN INVESTIGATION OF THEIR OWN. THEIR INITIAL FINDINGS WERE THAT THE MONITOR WAS CORRECT IN ITS CALCULATIONS AND THE WAVEFORM FROM THE INCIDENT DID NOT MEET THE REQUIREMENTS TO TRIGGER A V-TACH ALARM. THERE HAS N

    Patient problems: Insufficient Information

    MAUDE report 23768211 ↗
  • MalfunctionDec 10, 2025Device Alarm System; Device Alarm System

    PER MEDWATCH REPORT REFERENCE # MW5177989: NURSE WAS GOING THROUGH EVENT LOGS TO PRINT OFF A STRIP FROM THE PATIENT'S RECORDS. WHILE GOING THROUGH THE RECORDS, A RUN OF PREMATURE VENTRICULAR COMPLEXES (PVCS) WERE FOUND AT A TACHYCARDIC RATE AND THERE WAS NO HIGHLIGHTING OF THAT TRACE TO INDICATE ANY ALARMS. THERE WAS ALSO NO CHANGE IN CALCULATED HEART RATE. THERE WERE NO ALARMS THAT SOUNDED AT THE TIME OF EVENT. FROM THE ATTACHED IMAGE OF THE STRIP, IT IS EVIDENT THAT THERE WAS A VENTRICULAR TACHYCARDIA (V-TACH) EVENT. AFTER THIS WAS FOUND, STAFF GOT BIOMED INVOLVED TO DO AN INVESTIGATION. FROM BIOMED'S FINDINGS, THE MULTIPARAMETER MODULE (MPM) MODULE THAT WAS IN THE NEMESIS PRIME (N12) MONITOR WAS UNABLE TO ANNOTATE THE QRS COMPLEX. THE COMPATIBILITY ISSUE IS BELIEVED TO BE WHY THERE WAS NO ALARM. MEANWHILE, THE PATIENT WAS MOVED TO A DIFFERENT ROOM WITH A NEW N12 AND N1 MODULE CONNECTED AND THERE WAS A V-TACH ALARM FOR A SIMILAR RUN OF PVCS. MINDRAY WAS CALLED AND CREATED AN INVESTIGATION OF THEIR OWN. THEIR INITIAL FINDINGS WERE THAT THE MONITOR WAS CORRECT IN ITS CALCULATIONS AND THE WAVEFORM FROM THE INCIDENT DID NOT MEET THE REQUIREMENTS TO TRIGGER A V-TACH ALARM. THERE HAS N

    Patient problems: Insufficient Information

    MAUDE report 23768201 ↗

Other Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms) devices

All Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms) devices →