uMI uMI 550 (UDI 06971576832057)Recalls & reported problems
Shanghai United Imaging Healthcare Co., Ltd. · Positron Emission Tomography and Computed Tomography System, mobile configuration · Class 2
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 06971576832057), received between Aug 18, 2024 and Aug 18, 2024 — classified as 1 malfunction.
The most frequently reported specific device problem is “Smoking” (1 report).
2 FDA recall records name this device's identifier, 2 still open.
8 other recalls cover devices cleared under the same FDA submission (K182237, K193210, K193241, K210001, K210418); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Mar 1, 2023Open, Classified
Due to a software issue where the process of patient scanning, the scatter correction may occasionally fail with will potentially cause a failure of the PET image reconstruction and generation of PET images.
Z-1353-2023 · Root cause: Software design · FDA record ↗
- Aug 2, 2021Open, Classified
The EXAM acquisition workflow could occasionally freeze after canceling the PET scan.
Z-2593-2021 · Root cause: Device Design · FDA record ↗
Most reported problems
Device problems
- Smoking1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 18, 2024Smoking; Smoking
07:15 AM OF J(B)(6)2024, UIHS CALL CENTER RECEIVED CALL ABOUT PET COOLING ERRORS, UIHS CALL CENTER CSE WORKED WITH SITE TO SHUT DOWN SYSTEM TO RECOVER THE ERROR 2 TIMES WITH FULL POWER OFF. 09:20 AM, LOCAL CSE ARRIVED AT THE SITE FOR DISPATCH. WHEN CSE ARRIVED, HE NOTICED A STRONG SMELL OF ELECTRICAL OZONE AND THE SYSTEM WAS ON. CSE FOUND THAT KC1 RELAY CABLE ASSY WAS CHARRED. THE HALON GAS FIRE SUPPRESSION WAS DEPLOYED. THERE IS NO DISCOLORATION AND DISTORTION ON THE COVER. THERE WAS NO ONE INJURED DUE TO THIS FAILURE. LOCAL CSE CHECKED SYSTEM FAILURE PARTS AND REPLACED WITH SPARE PARTS, AND PASSED QUALITY CHECK ON JULY 3RD,2024. SYSTEM HAS BEEN DELIVERED TO CUSTOMER NORMALLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19999027 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K182237, K193210, K193241, K210001, K210418. One submission often covers many models, so check the product and lot details in each record.
- Z-3283-2026 · Aug 11, 2026 · Open, ClassifiedPositron Emission Tomography and Computed Tomography System, Model: uMI 550 with software
- Z-0223-2026 · Sep 12, 2025 · Open, ClassifiedShanghai United Imaging Healthcare Co., Ltd. Positron Emission Tomography and Computed Tomography System Model:uMI 550 MD, Rx Only
- Z-0087-2023 · May 30, 2022 · Open, ClassifiedPositron Emission Tomography and Computed Tomography System, Model: uMI 550
- Z-2594-2021 · Aug 2, 2021 · Open, ClassifieduEXPLORER PET/CT System - Product Usage: intended to be operated by qualified healthcare professionals to assist in the detection, localization, diagnosis, staging, restaging, trea
- Z-0448-2021 · Aug 5, 2020 · Open, ClassifiedPositron Emission Tomography and Computed Tomography System, Model uMI 550, Rx, CE, GTIN: 06971576832026 - Product Usage: intended to be operated by qualified healthcare professio
- Z-2800-2020 · Jul 16, 2020 · TerminateduMI 550 PET/CT Diagnostic Imaging System - Product Usage: This system is intended to be operated by qualified healthcare professionals to assist in the detection, localization, dia
- Z-2819-2020 · Mar 26, 2020 · TerminatedPositron Emission Tomography and Computed Tomography System, Model: uMI 550, Material Number: 88000057 - Product Usage: This system is intended to be operated by qualified healthca
- Z-1708-2020 · Aug 26, 2019 · TerminatedPositron Emission Tomograpy and Computed Tomography System, Model: uMI 550, with software version R001.3.0.0.750505, Rx Only, CE 0197, UDI: (01)06971576832026 - Product Usage: Thi
Other System, Tomography, Computed, Emission devices
- uMI uMI 550Shanghai United Imaging Healthcare Co., Ltd. · System, Tomography, Computed, Emission6 recalls0 reports
- Discovery 670 PRO H3100AHG E Medical Systems Israel Ltd. · System, tomography, computed, emission2 recalls0 reports
- uMI uMI Panorama 35Shanghai United Imaging Healthcare Co., Ltd. · System, Tomography, Computed, Emission2 recalls0 reports
- NMCT 860G E Medical Systems Israel Ltd. · System, tomography, computed, emission2 recalls0 reports
- Vereos PET/CTPhilips Medical Systems Nederland B.V. · System, tomography, computed, emission1 recall1 report
- NM 830G E Medical Systems Israel Ltd. · System, tomography, computed, emission1 recall1 report
- Discovery 670 ES H3100AGG E Medical Systems Israel Ltd. · System, tomography, computed, emission1 recall0 reports
- Discovery MIGE Medical Systems, L.L.C. · System, tomography, computed, emission1 recall0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.