Skip to content

CureWrap™ Infant, 2.5-4Kg 500-03518Recalls & reported problems

Belmont Instrument Corporation · Accessory for the CritiCool system. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 07290012127182), received between May 12, 2020 and Mar 5, 2026 — classified as 8 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (6 reports).

1 other recall cover devices cleared under the same FDA submission (K083662); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak6
  • Leak/Splash2
  • Product Quality Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8
  • No Information1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 5, 2026Leak/Splash

    IT WAS REPORTED THAT A BABY UNDERGOING THERAPEUTIC HYPOTHERMIA EXPERIENCED LEAKAGE FROM THE COOLING MATTRESS/WRAP SYSTEM DURING TREATMENT. IT WAS NOTED THAT LEAKAGE HAD OCCURRED MULTIPLE TIMES DURING THE COOLING PERIOD. IN TOTAL, THREE COOLING WRAPS WERE USED PRIOR TO THE FINAL INCIDENT DURING THE REWARMING PHASE. THE INFANT WAS FOUND WET ON FOUR OCCASIONS DURING TREATMENT. DURING THE REWARMING PHASE, WITH APPROXIMATELY 45 MINUTES REMAINING, LEAKAGE WAS AGAIN IDENTIFIED, WHICH SLOWED THE REWARMING PROCESS. THE MEDICAL TEAM MADE THE DECISION TO REMOVE THE COOLING MATTRESS, DRY THE INFANT, AND ALLOW REWARMING TO CONTINUE WITHOUT THE DEVICE FOR THE REMAINING DURATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24524171 ↗
  • MalfunctionJan 27, 2026Fluid/Blood Leak; Product Quality Problem

    RECEIVED BABY ON THERAPEUTIC COOLING WITH CUREWRAP COOLING JACKET ON. THE STAFF OBSERVED THE JACKET WAS WET ON THE HEAD AREA AND SOAKING THROUGH THE SHEETS. ALL OTHER AREAS OF THE JACKET WAS DRY AND THE JACKET WAS NOT FILLED WITH WATER. AFTER THE STAFF NOTICED THIS, THEY REPLACED THE COOLING JACKET.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24192377 ↗
  • MalfunctionAug 4, 2025Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT, " COOLING BLANKETS WERE CHANGED THREE TIMES DUE TO DIFFUSE LEAKING ON THE SURFACE OF THE BLANKETS".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22705914 ↗
  • MalfunctionMay 18, 2023Fluid/Blood Leak; Fluid/Blood Leak

    THE USER REPORTED THAT, WHILE ASSESSING THE BABY, THE COOLING BLANKET WAS NOTED TO BE LEAKING AND A PUDDLE WAS FORMED. INFANT IMMEDIATELY CHANGED TO NEW COOLING BLANKET, DRIED, AND STABILIZED. LEAKING BLANKET WAS SET ASIDE FOR AN INVESTIGATION. THERE IS NO PATIENT HARM ASSOCIATED WITH THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16952302 ↗
  • MalfunctionJun 18, 2021Fluid/Blood Leak; Fluid/Blood Leak

    BELMONT MEDICAL TECHNOLOGIES RECEIVED A REPORT FROM THE USER FACILITY THAT AN INFANT CUREWRAP LEAK STARTED AS A SLOW DRIP AND BECAME A CONSISTENT SPRAY WHEN CLAMPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12028775 ↗
  • MalfunctionMay 13, 2021Fluid/Blood Leak; Fluid/Blood Leak

    THE USER FACILITY REPORTED THAT THE INFANT CUREWRAP WAS LEAKING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11823271 ↗
  • MalfunctionNov 7, 2020Fluid/Blood Leak; Fluid/Blood Leak

    BELMONT'S DISTRIBUTOR RELAYED A REPORT FROM THE USER FACILITY THAT THE CURE WRAP LEAKED DURING A PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10805965 ↗
  • MalfunctionMay 12, 2020Fluid/Blood Leak; Fluid/Blood Leak

    THE USER FACILITY REPORTED THAT AN INFANT CUREWRAP WAS LEAKING AT THE PLASTIC CONNECTION SIDE OF THE BLANKET.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Information

    MAUDE report 10047512 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K083662. One submission often covers many models, so check the product and lot details in each record.

  • Z-0153-2023 · Jul 20, 2021 · Open, Classified
    CritiCool, thermoregulation device

Other System, Thermal Regulating devices

All System, Thermal Regulating devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page