Skip to content

CureWrap™ Pediatric 122–135cm (single) 500-03548Recalls & reported problems

Belmont Instrument Corporation · Accessory for the CritiCool system. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07290012127380), received between Jun 27, 2024 and Jun 27, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Device Handling Problem” (1 report).

1 other recall cover devices cleared under the same FDA submission (K083662); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Handling Problem1
  • Device-Device Incompatibility1
  • Leak/Splash1
  • Material Puncture/Hole1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 27, 2024Leak/Splash; Material Puncture/Hole; Device-Device Incompatibility; Device Handling Problem; Leak/Splash; Material Puncture/Hole; Device-Device Incompatibility; Device Handling Problem

    THE USER FACILITY IN UK REPORTED THAT PATIENT BEING ACTIVELY COOLED FOR NEUROPROTECTION. PATIENT PLACED ON CRITICOOL COOLING MATTRESS, CONNECTED TO MACHINE AND COOLING COMMENCED. LARGE AMOUNT OF LEAKING FROM COOLING MATTRESS NOTED COVERING BED AND PATIENT. COOLING MACHINE STOPPED, COOLING MATTRESS REMOVED AND PATIENT CHANGED. NEW COOLING MATTRESS PUT UNDER PATIENT AND COOLING RESTARTED. NO HARM TO PATIENT. HOLE NOTED AT TOP OF COOLING MATTRESS. CRITICOOL LOT NUMBER M223561 3548 PICU TECHS AND PICU STORES NURSING LEAD AWARE AND MONITORING MACHINE AND COOLING USE FOR ANY FURTHER ISSUES. MANUFACTURER ALREADY INVOLVED AS PREVIOUS ISSUES WITH COOLING MATTRESS LEAKS. MANUFACTURER UPDATED ABOUT RECENT INCIDENT. ON INVESTIGATION -ARTERIAL TRANSDUCER CLIP WAS FOUND TO HAVE BEEN ATTACHED TO THE COOLING MATTRESS AND THE HOLE FOUND WAS NEXT TO WHERE THE TRANSDUCER HAD BEEN ATTACHED TO COOLING MATTRESS. THIS COULD HAVE BEEN THE CAUSE OF THE HOLE IN THE MATTRESS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19624004 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K083662. One submission often covers many models, so check the product and lot details in each record.

  • Z-0153-2023 · Jul 20, 2021 · Open, Classified
    CritiCool, thermoregulation device

Other System, Thermal Regulating devices

All System, Thermal Regulating devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page