BioProtect Balloon Implant™ System 5135Recalls & reported problems
Bioprotect Ltd · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 7
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 07290014878020), received between Feb 26, 2026 and Sep 13, 2026 — classified as 7 injuries.
The most frequently reported specific device problem is “Difficult or Delayed Positioning” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Difficult or Delayed Positioning2
- Malposition of Device2
- No Apparent Adverse Event1
Patient problems
- Urinary Retention4
- Hemorrhage/Blood Loss/Bleeding2
- Perforation2
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjurySep 13, 2026Difficult or Delayed Positioning
Patient problems: Hemorrhage/Blood Loss/Bleeding; Urinary Retention
MAUDE report 50747514 ↗InjurySep 13, 2026Malposition of Device
Patient problems: Perforation
MAUDE report 50747513 ↗InjuryJul 2, 2026No Apparent Adverse Event
IT WAS REPORTED TO BIOPROTECT LTD. THAT A BIOPROTECT BALLOON IMPLANTATION PROCEDURE WAS PERFORMED ON (B)(6) 2026. DURING PLANNING CT ON (B)(6) 2026 IT WAS FOUND THAT THE BALLOON APPEARANCE ON CT WAS NOT AS EXPECTED. A DISCUSSION WITH THE PHYSICIAN REVEALED THAT THE BALLOON WAS INTENTIONALLY FILLED WITH CONTRAST MEDIA RATHER THAN SALINE AS SPECIFIED IN THE PRODUCT'S INSTRUCTIONS FOR USE (ALSO EMPHASIZED DURING PHYSICIAN'S TRAINING). THE PATIENT UNDERWENT AN ADDITIONAL PROCEDURE WHERE THE CONTRAST MEDIA WAS REMOVED FROM THE BALLOON AND REPLACED WITH SALINE. THE INVESTIGATION FINDINGS SUGGEST THAT THE ROOT CAUSE OF THE EVENT IS RELATED TO ABNORMAL USE RATHER THAN TO A DEVICE-RELATED ISSUE. THE PATIENT DID NOT PRESENT ANY CLINICAL SYMPTOMS, AND HIS RADIATION TREATMENT WAS DELAYED FOR APPROXIMATELY 1 MONTH.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25688861 ↗InjuryMay 19, 2026Malposition of Device
IT WAS REPORTED TO BIOPROTECT LTD. THAT A BIOPROTECT BALLOON IMPLANTATION PROCEDURE WAS PERFORMED ON (B)(6) 2026. ULTRASOUND VISUALIZATION WAS POOR DUE TO PATIENT'S INSUFFICIENT BOWEL PREPARATION AND THIN RECTAL WALL. DIGITAL RECTAL EXAMINATION (DRE) AND CT FOLLOWING THE PROCEDURE CONFIRMED THE BALLOON WAS PLACED IN THE LUMEN OF THE RECTUM. THE PATIENT WAS REFERRED TO A LOCAL ER WHERE THE BALLOON WAS SUBSEQUENTLY ASPIRATED AND REMOVED. A 5 MM PUNCTURE IN THE RECTAL WALL WAS REPAIRED BY A COLORECTAL SURGEON. THE PATIENT IS EXPECTED TO RECEIVE A REPLACEMENT SPACER, AND HIS RADIATION TREATMENT WAS DELAYED.
Patient problems: Perforation
MAUDE report 25216898 ↗InjuryMar 30, 2026Difficult or Delayed Positioning
IT WAS REPORTED TO BIOPROTECT LTD. ON (B)(6) 2026 THAT A BIOPROTECT BALLOON IMPLANTATION PROCEDURE WAS PERFORMED ON (B)(6) 2026. THE PROCEDURE WAS DONE UNDER LOCAL ANESTHESIA FOLLOWING PLACEMENT OF FIDUCIAL MARKERS. DURING THE PROCEDURE, MULTIPLE PLACEMENT ATTEMPTS WERE REQUIRED AFTER AN INITIAL TRACT EXTENDED BEYOND THE PROSTATIC APEX AND ENGAGED THE POSTERIOR CAPSULE. THE BALLOON WAS ULTIMATELY DEPLOYED IN AN ACCEPTABLE POSITION. POST-PROCEDURE, THE PATIENT DEVELOPED URINARY RETENTION REQUIRING FOLEY CATHETER PLACEMENT, WHICH RESOLVED WITHIN 48 HOURS. PLANNING MRI (B)(6) 2026), IMAGING RAISED CONCERN FOR HEMATOMA AND POSSIBLE VASCULAR INJURY. SUBSEQUENT DYNAMIC CONTRAST MRI SUGGESTED A PERIPROSTATIC ARTERIOVENOUS FISTULA, AND THE PATIENT UNDERWENT SUCCESSFUL COIL EMBOLIZATION WITH RESOLUTION OF THE VASCULAR ABNORMALITY. THE REMAINING PELVIC VASCULATURE WAS UNREMARKABLE. THE PATIENT WAS ADMITTED FOR 24-HOUR OBSERVATION AND WAS REPORTED TO BE FINE. HIS RADIATION TREATMENT (SBRT) HAS BEEN DELAYED, AND NEW PLANNING IMAGING WILL BE SCHEDULED.
Patient problems: Hemorrhage/Blood Loss/Bleeding; Urinary Retention
MAUDE report 24730457 ↗InjuryMar 16, 2026Adverse Event Without Identified Device or Use Problem
MFR REPORT #: 3014285231-2026-00002 IT WAS REPORTED TO BIOPROTECT LTD. ON (B)(6) 2026 THAT A BIOPROTECT BALLOON IMPLANTATION PROCEDURE WAS PERFORMED ON (B)(6) 2026. THE PROCEDURE WAS DONE UNDER MONITORED ANESTHESIA CARE (MAC)FOLLOWING PLACEMENT OF FIDUCIAL MARKERS AND COMPLETED AS EXPECTED. ON SIMULATION CT PERFORMED ON (B)(6) 2026, THE PATIENT PRESENTED WITH A DISTENDED BLADDER. A FOLEY CATHETER WAS PLACED AND THE PATIENT WAS PRESCRIBED FLOMAX. ON (B)(6) 2026, THE PHYSICIAN PARTIALLY ASPIRATED THE BALLOON, AND THE CATHETER WAS REMOVED. THE PATIENT'S SYMPTOMS WERE RELIEVED, AND HE STARTED HIS RADIATION TREATMENT AS PLANNED WITH NO DELAY.
Patient problems: Urinary Retention
MAUDE report 24613690 ↗InjuryFeb 26, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO BIOPROTECT LTD. ON (B)(6) 2026 THAT A BIOPROTECT BALLOON IMPLANTATION PROCEDURE WAS PERFORMED ON (B)(6) 2026. THE PROCEDURE WAS DONE UNDER MONITORED ANESTHESIA CARE (MAC)FOLLOWING PLACEMENT OF FIDUCIAL MARKERS AND COMPLETED AS EXPECTED. ON SIMULATION CT PERFORMED ON (B)(6) 2026, THE PATIENT PRESENTED WITH A DISTENDED BLADDER. A FOLEY CATHETER WAS PLACED AND THE PATIENT WAS PRESCRIBED FLOMAX. ON (B)(6) 2026, THE PHYSICIAN PARTIALLY ASPIRATED THE BALLOON, AND THE CATHETER WAS REMOVED. THE PATIENT'S SYMPTOMS WERE RELIEVED, AND HE STARTED HIS RADIATION TREATMENT AS PLANNED WITH NO DELAY.
Patient problems: Urinary Retention
MAUDE report 24454146 ↗
Other Hydrogel Spacer devices
- SpaceOAR VUE™ System - 10mL SV-2101Boston Scientific Corporation · Hydrogel spacer1 recall4 deaths414 reports
- SpaceOAR System SO-2101Boston Scientific Corporation · Hydrogel Spacer1 recall1 death353 reports
- Barrigel 011508Palette Life Sciences, Inc. · Hydrogel Spacer2 deaths6 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page