Face Applicator (Cheek Right) AG608275ARecalls & reported problems
Inmode M.D Ltd · Cheek Right Applicator · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07290016633542), received between Jul 27, 2021 and Feb 16, 2023 — classified as 2 injuries.
The most frequently reported specific device problem is “Use of Device Problem” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Use of Device Problem2
Patient problems
- Burn(s)1
- Full thickness (Third Degree) Burn1
- Skin Burning Sensation1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryFeb 16, 2023Use of Device Problem; Use of Device Problem
InjuryJul 27, 2021Use of Device Problem
THE PATIENT WAS TREATED WITH EVOKE SYSTEM UTILIZING BOTH THE RIGHT AND LEFT CHEEK APPLICATORS. DURING THE TREATMENT TIME, THE PATIENT CALLED FOR ASSISTANCE AND COMPLAINT OF BURNED SENSATION IN HER RIGHT CHEEK. THE HEALTHCARE TEAM IMMEDIATELY STOPPED THE DEVICE OPERATION AND HAD THE PATIENT HOLD ICE TO THE AREA OF THE INCIDENT. PICTURES OF THE PATIENT TREATMENT AREA REVEALED A ROUND EXPOSED BURN ABOUT 2 CM IN DIAMETER, LOCATED AT THE SKIN PORTION IN CORRESPONDENCE WITH THE LEFT TMJ. MULTI-LAYERED TRANSDERMAL DAMAGE WITH TISSUE NECROSIS WAS EVIDENT. THE PATIENT WAS SENT FOR INJURY MANAGEMENT AND WAS UNDER A CARE OF A DERMATOLOGIST FOR WOUND MANAGEMENT AND HAS BEEN TO A BURN CLINIC. THE DEVICE WAS RETURNED TO THE MANUFACTURER FOR TESTING AND WAS SHOWN TO WORK SAFELY UNDER THE REQUIRED PERFORMANCE SPECIFICATIONS, NO DEVICE DEFECTS OR MALFUNCTIONS WERE DETECTED. FDA SAFETY REPORT ID # (B)(4).
Patient problems: Full thickness (Third Degree) Burn; Skin Burning Sensation
MAUDE report 12228102 ↗
Other Massager, Vacuum, Radio Frequency Induced Heat devices
- NUERA TIGHT APMD145-1ch.USBios Srl · Massager, Vacuum, Radio Frequency Induced Heat6 reports
- NUERA TIGHT APMD140Bios Srl · Massager, Vacuum, Radio Frequency Induced Heat2 reports
- Venus Bliss BL000001Venus Concept Inc. · Massager, Vacuum, Radio Frequency Induced Heat1 report
- EmBodyPlus (Tite) HP AG608083AInmode M.D Ltd · Massager, Vacuum, Radio Frequency Induced Heat1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.