Skip to content

Chameleon 9x40Recalls & reported problems

Av Medical Technologies Ltd · Chameleon PTA Balloon Dilatation Catheter 9x40 mm · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07290016745054), received between Jun 3, 2021 and Jun 3, 2021 — classified as 1 malfunction.

The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Burst Container or Vessel1
  • Device Fell1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 3, 2021Burst Container or Vessel; Device Fell; Burst Container or Vessel; Device Fell

    ACCORDING TO THE REPORTER, DURING PROCEDURE, THE PHYSICIAN INFLATED THE BALLOON IN CEPHALIC ARCH USING TWO SYRINGE METHOD AND SALINE AS INFLATION FLUID, 1-2 TIMES OF INFLATION, THE BALLOON BURST LONGITUDINALLY. UPON TRYING TO REMOVE THE BALLOON FROM SHEATH (COMPETITOR¿S BRAND), THE BALLOON GOT CAUGHT IN SHEATH, AND WHEN CAUGHT, IT SNAPPED SPONTANEOUSLY JUST PROXIMAL OF BALLOON INJECTION PORT (3RD PORT). THE BALLOON REMAINED SNAPPED OFF AND STUCK IN SHEATH. IT DID NOT CAUSE ANY VESSEL TRAUMA. UPON REMOVAL OF 7FR SHEATH, THEY WERE ABLE TO PULL THE BALLOON OUT OF SHEATH FOR FURTHER INSPECTION. ALL FRAGMENTS OF THE BALLOON WERE RETRIEVED. THE PROCEDURE WAS COMPLETED WITH NEW SHEATH AND NEW BALLOON (COMPETITOR'S PRODUCT). NO ADDITIONAL MEDICAL INTERVENTION REQUIRED TO RETRIEVE THE DETACHED ITEM. THE FLUSHING WAS PERFORMED PER THE IFU (INSTRUCTIONS FOR USE). NOTHING UNUSUAL OBSERVED ON THE DEVICE PRIOR TO USE. NO OTHER PRODUCTS BEING UTILIZED WITH THE DEVICE. NO OTHER DEFECTS/DAMAGES FOUND ON THE PRODUCT. THEY USED A .035 GUIDE WIRE. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT. NO RESISTANCE ENCOUNTERED WHEN ADVANCING THE DEVICE AND NO EXCESSIVE FORCE USED. THERE WAS NO K

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11930431 ↗

FDA clearance and classification

The FDA regulates product code KRA (catheter, continuous flush) under 21 CFR 870.1210 as a Class II device. 21 CFR 870.1210 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Chameleon 9x40

Has the Chameleon 9x40 been recalled?

No FDA recall record in our data names this exact device identifier (DI 07290016745054), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Chameleon 9x40?

The FDA's MAUDE database lists 1 adverse event report for this device, received on Jun 3, 2021. That report lists device problems “Burst Container or Vessel” (1) and “Device Fell” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Chameleon 9x40?

None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the Chameleon 9x40, and is it still sold?

The FDA's device database lists Av Medical Technologies Ltd as the company responsible for it (brand Chameleon, model 9x40). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Catheter, Continuous Flush devices

All Catheter, Continuous Flush devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page